User`s guide

492 Chapter 15
Measurement Examples
Measuring the SAW Bandpass Filter Using the Segment Sweep
2. Create a Segment Sweep Table
Follow the steps below to make entries in the segment sweep table.
Step 1. Display the segment table.
Step 2. Display the IF bandwidth setting column in the segment table.
NOTE When setup items (power level, delay time, sweep mode, and sweep time in this case) are
not displayed in the segment table, the setting for the channel in use applies to all segments.
Step 3. Enter the setup data in the segment table.
Setup Description Key Operation
Presetting
-
OK
Displaying the segment table
-
Edit Segment Table
Setup Description Key Operation
Moving the focus to the softkey menu
Display of IF bandwidth setting column: ON
List IFBW (turn it ON)
Setup Description Key Operation
Moving the focus to the segment table (select)
Segment 1
Start frequency: 440 MHz
Stop frequency: 915 MHz
Number of points: 47
IF bandwidth: 70 kHz
Segment 2
Start frequency: 915 MHz
Stop frequency: 980 MHz
Number of points: 130
IF bandwidth: 100 kHz
Segment 3
Start frequency: 980 MHz
Stop frequency: 1.035 GHz