User`s guide

488 Chapter 14
Controlling E5091A
Connecting Two E5091As
Connecting Two E5091As
Give consideration to the following items when you make measurements by using two
E5091As.
Set their IDs to different values. The instruments will not work correctly if their IDs are
the same.
Connect calibration standards and the DUT only after confirming the connection and
port assignment of the E5091As.