User`s guide
Chapter 13 465
Setting and Using the Control and Management Functions
System Recovery
13.Setting and Using the
Control and Management
Functions
Procedure to execute the factory recovery function (2)
This procedure is valid for the E5070B : MY42300633, and above, E5071B : Y42301397
and above. If your unit doesn’t follow the procedure written below, please refer to
“Procedure to execute the factory function (1)” on page 456 for the detail.
This section describes how to return the contents of the C drive to the factory state.
NOTE You need the keyboard for this operation.
Step 1. Shut down the E5070B/E5071B.
Step 2. Connect the keyboard to the E5070B/E5071B.
Step 3. Insert the disk for the system recovery into the floppy disk drive of the E5070B/E5071B.
Step 4. Press the standby switch of the E5070B/E5071B to turn it on.
Step 5. When the screen as shown in the figure below appears, press and hold of the
keyboard until this screen disappears.
NOTE After several seconds, the next screen appears automatically even if you do not press any
key, so do not miss it.
If the above message does not appear, the instrument is at fault; contact your local Agilent
customer center listed at the end of this manual or distributor.