User`s guide

408 Chapter 12
Optimizing Measurements
Setting Port Extensions and Loss Values
3. Specifying a loss value as a calculation target
Specify whether you want to include a loss value in the calculation result.
Step 1. Press or to activate the channel for which you want to set auto
port extension.
Step 2. Press .
Step 3. Press
Port Extensions.
Step 4. Press
Auto Port Extension.
Step 5. Press
Include Loss to turn it on.
NOTE If the setting is not made before starting the measurement of the OPEN/SHORT standard, it
does not affect the calculation result.
4. Specifying a DC loss value as a calculation target
Specify whether you want to include a DC loss value in the calculation result.
Step 1. Press or to activate the channel for which you want to set auto
port extension.
Step 2. Press .
Step 3. Press
Port Extensions.
Step 4. Press
Auto Port Extension.
Step 5. Press
Adjust Mismatch to turn it on.
NOTE If the setting is not made before starting the measurement of the OPEN/SHORT standard, it
does not affect the calculation result.
Softkey Function
ON Calculates a loss value.
OFF Does not calculate a loss value.
Softkey Function
ON Calculates a DC loss value.
OFF Does not calculate a DC loss.