User`s guide
Chapter 8 329
Frequency-Offset Measurement (Option 008)
Measurement of Harmonic Distortion
8. Frequency Offset
Measurement
(Option 008)
Figure 8-12 Example of measurement frequency
2. Implementing Receiver Calibration
Receiver calibration is required for the absolute measurement. Receiver calibration of the
port to be used for the measurement increases the accuracy in the absolute measurement.
For detailed information on receiver calibration, see "Receiver Calibration" on page 220.
3. Setting Absolute Measurement Parameters
The frequency-offset function has an added absolute measurement function.
Procedures for setting absolute measurement parameters
Follow the steps below to set the absolute measurement parameters.
Step 1. Press (or ) and (or ) to activate the trace for
which you want to implement absolute measurement.
Step 2. Press to display the “Measurement Menu.”
Step 3. Press
Absolute to select measurement parameters.
Softkey
*1
Function
A (n) Absolute measurement in Port 1, test receiver
B (n) Absolute measurement in Port 2, test receiver
C (n) Absolute measurement in Port 3, test receiver
D (n) Absolute measurement in Port 4, test receiver
R1 (n) Absolute measurement in Port 1, reference receiver
R2 (n) Absolute measurement in Port 2, reference receiver