User`s guide

288 Chapter 7
Fixture Simulator
Converting the Port Impedance of the Measurement Result
Converting the Port Impedance of the Measurement Result
The measured value obtained by using a port impedance of 50 Ω can be converted into a
measured value at an arbitrary port impedance.
Figure 7-3 Port impedance conversion function
Converting the Port Impedance
Step 1. Press - Fixture Simulator - Port Z Conversion.
Step 2. Specify the port reference impedance for the required port.
Select
Port 1 Z0 Real, Port 2 Z0 Real, Port 3 Z0 Real, or Port 4 Z0 Real to set the port
reference impedance in “Real” format.
Select
Port 1 Z0 Real and Port 1 Z0 Imag, Port 2 Z0 Real and Port2 Z0 Imag, Port 3
Z0 Real
and Port 3 Z0 Imag, or Port 4 Z0 Real and Port 4 Z0 Imag to set the port
reference impedance in “Complex” format.
Step 3. Press
Port Z Conversion to change the port impedance conversion function to the ON
state.
Step 4. Press
Return.
Step 5. If
Fixture Simulator is OFF, press the key again to turn it ON.