238 Chapter 4
Calibration
Scalar-Mixer Calibration
Figure 4-77 Connection of power sensor
Step 28. Press
Port x@Freq x.
Step 29. Press
Port x@Freq y.
Step 30. Press
Port y@Freq x.
Step 31. Press
Port y@Freq y.
Step 32. Press
Return.
Step 33. Press
Done to exit the scalar-mixer calibration. This step allows the calibration coefficient
to be calculated, turning on the error correction function automatically.
Operational procedure (when using ECal module)
To start calibration, refer to “Operational Procedure (when using mechanical calibration
kit)” on page 235; the steps up to Step 7 in that procedure are common to this procedure.
Step 1. Select a test port and then press
Power Meter.
NOTE In scalar-mixer calibration, the power meter is used. For more information on how to set
the power meter, see “Preparing to control the power meter” on page 210.
Step 2. Press
Use Sensor. Each time the key is pressed, Channels A and B switch over alternately.
If you are using a one-channel power meter, select Channel A.
Step 3. Connect the power sensor for the selected channel to the selected port, as shown in Figure
4-78.