User`s guide
Chapter 4 203
Calibration
Specifying Different Standard for Each Frequency
4. Calibration
Defining standard for each subclass
This section demonstrates the procedure to specify a different OPEN standard for each
frequency band, using subclasses #1 and #2. In this example, the standards created in
“Defining different standard for each frequency band” on page 202 are used.
Procedure
Step 1. Press .
Step 2. Press
Cal Kit.
Step 3. Select a calibration kit you want to use. (In this example, use the calibration kit defined in
Step 3 of “Defining different standard for each frequency band” on page 202.)
Step 4. Press
Modify Kit.
Step 5. Press
Specify CLSs.
Step 6. Press
Sub Class, and select Sub Class 1.
Step 7. Select the standard type
Open, press Set All, make settings for all ports.
Step 8. Select #1 (
Open 3G).
Step 9. Press
Return.
Step 10. Then, press
Sub Class, and select Sub Class 2.
Step 11. Select the standard type
Open, press Set All, make settings for all ports.
Step 12. Select #2 (
Open 6G).
Step 13. Press
Return.