User`s guide

Chapter 4 197
Calibration
Changing the Calibration Kit Definition
4. Calibration
85036B/E
NOTE The designations (m) and (f) in the name (label) of the standard indicate male (m) and
female (f) analyzer connectors.
1. Short(m) 2. Open(m) 3. Broadband 4. Thru
Label Short(m) Open(m) Broadband Thru
STD Type Short Open Load Delay/Thru
C0 [×10
-15
F]
0 63.5 0 0
C1 [×10
-27
F/Hz]
08400
C2 [×10
-36
F/Hz
2
]
05600
C3 [×10
-45
F/Hz
3
]
0000
L0 [×10
-12
H]
0000
L1 [×10
-24
H/Hz]
0000
L2 [×10
-33
H/Hz
2
]
0000
L3 [×10
-42
H/Hz
3
]
0000
Offset Delay [s] 0 0 0 0
Offset Z0 [Ω]75757575
Offset Loss [Ω/s] 1.13 G 1.13 G 1.13 G 1.13 G
Arb. Impedance [Ω]50507550
7. Short(f) 8. Open(f)
Label Short(f) Open(f)
STD Type Short Open
C0 [×10
-15
F]
041
C1 [×10
-27
F/Hz]
040
C2 [×10
-36
F/Hz
2
]
05
C3 [×10
-45
F/Hz
3
]
00
L0 [×10
-12
H]
00
L1 [×10
-24
H/Hz]
00
L2 [×10
-33
H/Hz
2
]
00
L3 [×10
-42
H/Hz
3
]
00
Offset Delay [s] 17.544 p 17.544 p
Offset Z0 [Ω]75 75
Offset Loss [Ω/s] 1.13 G 1.13 G
Arb. Impedance [Ω]50 50