User`s guide

Chapter 4 119
Calibration
Full 2-Port Calibration
4. Calibration
Figure 4-16 Connecting standards in full 2-port calibration
Procedure
Step 1. Press or to select the channel for which you want to perform the
calibration.
Step 2. Press .
Step 3. Press
Calibrate.
Step 4. Press
2-Port Cal.
Step 5. Press
Select Ports.
Step 6. Select the test ports on which you will perform full 2-port calibration. (In the procedure
below, the selected test ports are denoted as x and y.)
NOTE An asterisk (*) in the upper-right of the softkey indicates that the existing calibration
coefficient will be cleared if you select the test port and execute the acquisition of the
Softkey Function
1-2 Selects test ports 1 and 2
1-3 Selects test ports 1 and 3
1-4 Selects test ports 1 and 4
2-3 Selects test ports 2 and 3
2-4 Selects test ports 2 and 4
3-4 Selects test ports 3 and 4