Caution Do not exceed the operating input power, voltage, and current level and signal type appropriate for the instrument being used, refer to your instrument's Function Reference. Electrostatic discharge(ESD) can damage the highly sensitive microcircuits in your instrument. ESD damage is most likely to occur as the test fixtures are being connected or disconnected. Protect them from ESD damage by wearing a grounding strap that provides a high resistance path to ground.
Safety Summary When you notice any of the unusual conditions listed below, immediately terminate operation and disconnect the power cable. Contact your local Agilent Technologies sales representative or authorized service company for repair of the instrument. If you continue to operate without repairing the instrument, there is a potential fire or shock hazard for the operator. - Instrument operates abnormally. - Instrument emits abnormal noise, smell, smoke or a spark-like light during operation.
Herstellerbescheinigung GERAUSCHEMISSION LpA < 70 dB am Arbeitsplatz normaler Betrieb nach DIN 45635 T.
Regulatory compliance information This product complies with the essential requirements of the following applicable European Directives, and carries the CE marking accordingly: The Low Voltage Directive 73/23/EEC, amended by 93/68/EEC The EMC Directive 89/336/EEC, amended by 93/68/EEC To obtain Declaration of Conformity, please contact your local Agilent Technologies sales office, agent or distributor.
Safety notice supplement ・ This equipment complies with EN/IEC61010-1:2001. ・ This equipment is MEASUREMENT CATEGORY I (CAT I). Do not use for CAT II, III, or IV. ・ Do not connect the measuring terminals to mains. ・ This equipment is POLLUTION DEGREE 2, INDOOR USE product. ・ This equipment is tested with stand-alone condition or with the combination with the accessories supplied by Agilent Technologies against the requirement of the standards described in the Declaration of Conformity.
Agilent E5070B/E5071B ENA Series RF Network Analyzers User’s Guide Eleventh Edition FIRMWARE REVISIONS This manual applies directly to instruments that have the firmware revision A.08.10. For additional information about firmware revisions, see Appendix A. Manufacturing No.
Notices The information contained in this document is subject to change without notice. This document contains proprietary information that is protected by copyright. All rights are reserved. No part of this document may be photocopied, reproduced, or translated to another language without the prior written consent of Agilent Technologies. Microsoft®,MS-DOS®,Windows®,Visual C++®,Visual Basic®,VBA® and Excel® are registered UNIX is a registered trademark in U.S.
Safety Summary The following general safety precautions must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions or with specific WARNINGS elsewhere in this manual may impair the protection provided by the equipment. Such noncompliance would also violate safety standards of design, manufacture, and intended use of the instrument. Agilent Technologies assumes no liability for the customer’s failure to comply with these precautions.
Safety Symbols General definitions of safety symbols used on the instrument or in manuals are listed below. Instruction Manual symbol: the product is marked with this symbol when it is necessary for the user to refer to the instrument manual. Alternating current. Direct current. On (Supply). Off (Supply). In-position of push-button switch. Out-position of push-button switch. A chassis terminal; a connection to the instrument’s chassis, which includes all exposed metal structure. Stand-by.
Certification Agilent Technologies certifies that this product met its published specifications at the time of shipment from the factory. Agilent Technologies further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology, to the extent allowed by the Institution’s calibration facility or by the calibration facilities of other International Standards Organization members.
Typeface Conventions Sample (bold) Boldface type is used when a term is defined or emphasis. Sample (Italic) Italic type is used for emphasis. key Indicates a hardkey (key on the front panel or external keyboard) labeled “Sample.” “key” may be omitted. Sample menu/button/box Indicates a menu/button/box on the screen labeled “Sample” which can be selected/executed by clicking. “menu,” “button,” or “box” may be omitted.
Documentation Map The following manuals are available for the Agilent E5070B/E5071B. • User’s Guide (Part Number E5070-904x0, attached to Option ABA) This manual describes most of the basic information needed to use the E5070B/E5071B. It provides a function overview, detailed operation procedure for each function (from preparation for measurement to analysis of measurement results), measurement examples, specifications, and supplemental information.
VBA Macro The Agilent folder (D:\Agilent) on the hard disk of the E5070B/E5071B contains the VBA macros (VBA Projects) used in this manual. The customer shall have the personal, non-transferable rights to use, copy, or modify the VBA macros for the customer’s internal operations. The customer shall use the VBA macros solely and exclusively for their own purposes and shall not license, lease, market, or distribute the VBA macros or modification of any part thereof.
Contents 1. Precautions Software Installed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 Before contacting us . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27 2. Overview of Functions Front Panel: Names and Functions of Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Contents Setting Channels and Traces. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Setting upper limits of number of channels/traces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Setting channel display (layout of channel windows) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Setting trace display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Contents OPEN/SHORT Response Calibration (reflection test) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107 Procedure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107 THRU Response Calibration (transmission test) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110 Procedure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Contents Storing user characteristics to the ECal module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Backup and recovery of ECal module's built-in flash memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Executing User-characterized ECal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Confidence Check on Calibration Coefficients Using ECal . . . . . . . . . . . . . . . . . . . . . . . . . . .
Contents Setting Up the Trigger and Making Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 243 Setting the Point Trigger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 245 Procedure to Set the Point Trigger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 245 Setting the low-latency external trigger mode . . . . . . . . . . . .
Contents Balance-unbalance conversion (option 313, 314, 413, or 414) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Functions for balanced port (option 313, 314, 413, or 414) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Extending the Calibration Plane Using Network De-embedding. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Using the Network De-embedding Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Contents Time domain function of E5070B/E5071B. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 335 Transformation to time domain. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 336 Measurement flow. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 336 Selecting a type . . . . . . . . . . . . . . . . . . . . . . . . .
Contents Displaying judgment result of limit test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Defining the limit line. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Turning the limit test ON/OFF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Limit line offset. . . . . . . . . . . . . . . . . . . . . . . . . . . .
Contents 428 Setting system controller (USB/GPIB interface) when c drive volume label in hard disk is more than CP810 431 Setting the Internal Clock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 433 Setting the Date and Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 433 Setting the Date/Time Display ON/OFF . . . . . . . . . . . . . . . . . . . . . . . .
Contents Connecting E5070B/E5071B and E5091A. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Powering on . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Setting the E5091A. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Selecting ID for E5091A . . . . . . . . . . . . . . . . . . . . . . .
Contents 2. Setting the Measurement Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 515 3. Performing Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 516 4. Setting a Balance Conversion Topology . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 517 5. Selecting Measurement Parameters . . . . . . . . . . . . . . . . . . .
Contents System capabilities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 565 Automation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 566 17. Measurement Accessories Test Port Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Contents Change 8 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 605 Change 7 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 605 Change 6 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 606 Change 5 . . . . . . . .
Contents Marker Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Marker Function Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Marker Search Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Measurement Menu . . . . . . . . . . . . . . . . .
Contents Sweep Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 743 Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 744 Trigger System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 745 Data Flow . . . . . . . . . . . .
Contents 24
1. Precautions 1 Precautions This chapter describes cautions that must be observed in operating the E5070B/E5071B.
Precautions Software Installed Software Installed The Windows operating system installed in this machine is customized for more effective operation, and has different functions that are not part of the Windows operating system for ordinary PCs (personal computers). Therefore, do not attempt to use the system in ways other than those described in this manual or to install Windows-based software (including anti-virus software) for ordinary PCs as doing so may cause malfunctions. Also note the followings.
Before contacting us If you encounter the following problems during startup or operation of the E5070B/E5071B, in which initial registration of the Windows 2000 Operating System has been properly performed, execute system recovery and update the firmware version. As for the system recovery procedure, refer to “System Recovery” on page 455.
Precautions Before contacting us 28 Chapter 1
2. Overview of Functions 2 Overview of Functions This chapter describes the functions of the E5070B/E5071B that can be accessed from the front panel, LCD screen, and rear panel.
Overview of Functions Front Panel: Names and Functions of Parts Front Panel: Names and Functions of Parts This section describes the names and functions of the parts on the front panel of the E5070B/E5071B. For more details on the functions displayed on the LCD screen, see “Screen Area: Names and Functions of Parts” on page 40. For more about the functions of softkeys, see Appendix D, “Softkey Functions,” on page 649.
Overview of Functions Front Panel: Names and Functions of Parts 1. Standby Switch Used for choosing between power-on ( | ) and standby ( NOTE ) states of the E5070B/E5071B. To turn off the power for the E5070B/E5071B, be sure to follow the steps described below. 2. Next, if necessary, turn off the power supply to the “8. Power Cable Receptacle (to LINE)” on page 56 on the rear panel.
Overview of Functions Front Panel: Names and Functions of Parts 3. ACTIVE CH/TRACE Block A group of keys for selecting active channels and traces. For more on the concepts of channels and traces, see “Setting Channels and Traces” on page 61. 32 Key Selects the next channel as the active channel. (Each time the key is pressed causes the active channel to step up from the channel with the currently designated number to one with a larger channel number.
Overview of Functions Front Panel: Names and Functions of Parts 4. RESPONSE Block A group of keys used mainly for setting up response measurements on the E5070B/E5071B. Key Key Changes between normal and maximum display of the active trace. In normal display, all of the traces defined on the channel (both active and non-active) are displayed on the screen. In maximum display, only the active trace is displayed over the entire area, with non-active traces not displayed.
Overview of Functions Front Panel: Names and Functions of Parts 5. STIMULUS Block A group of keys for defining the stimulus values (signal sources and triggers). Key Displays the data entry bar for specifying the start value of the sweep range in the upper part of the screen. (It also displays the “Stimulus Menu” on page 720 for specifying the sweep range on the right side of the screen.) Key Displays the data entry bar for specifying the stop value of the sweep range in the upper part of the screen.
Overview of Functions Front Panel: Names and Functions of Parts 7. NAVIGATION Block The following descriptions show how the NAVIGATION block keys work both when the focus is on a softkey menu and when the focus is on the data entry area. For more on manipulating tables and dialog boxes, refer to the manipulation procedure for each of these functions.
Overview of Functions Front Panel: Names and Functions of Parts 8. ENTRY Block A group of keys used for entering numeric data. ... Keys (numeric keys) Alternately changes the sign (+, −) of a numeric value in the data entry area. Key Adds a prefix to the numeric data typed by using the numeric key and and then enters that data. One of the two prefixes written on the surface of the key is automatically selected depending on the parameter to be entered. is entered without a prefix.
Overview of Functions Front Panel: Names and Functions of Parts 9. INSTR STATE Block A group of keys related to the macro function, store and call function, control/management function, and the presetting of the E5070B/E5071B (returning it to the preset state). Key Key Key Executes a VBA procedure called “main” that has a VBA module named Module1. Key Stops the VBA procedure being executed. Key Displays the “Save/Recall Menu” on page 715 on the right side of the screen.
Overview of Functions Front Panel: Names and Functions of Parts 10. MKR/ANALYSIS Block A group of keys used for analyzing the measurement results by using the markers, fixture simulator, etc. For more on the functions of the keys in the MKR/ANALYSIS block, see Chapter 2 “Overview of Functions” in the User’s Guide. Displays the “Marker Menu” on page 701 on the right side of the screen. Manipulating the “Marker Menu” enables you to turn the markers on/off and move them by entering stimulus values.
Overview of Functions Front Panel: Names and Functions of Parts 12. Front USB Port NOTE We do not support connections to the USB port of devices other than designated printers, ECal modules, the USB/GPIB interface, and multiport test sets. 13. Ground Terminal Connected to the chassis of the E5070B/E5071B. You can connect a banana-type plug to this terminal for grounding. Chapter 2 39 2.
Overview of Functions Screen Area: Names and Functions of Parts Screen Area: Names and Functions of Parts This section describes the names and functions of parts on the LCD screen of the E5070B/E5071B. Figure 2-2 Screen display 1.
Overview of Functions Screen Area: Names and Functions of Parts Data entry bar NOTE To manipulate the data entry bar by using the front panel keys, the data entry bar must be selected as the object to manipulate (with the focus placed on it). When the focus is placed on the data entry bar, the entire bar is displayed in blue. Pressing or clicking in the “8. ENTRY Block” on page 36 enables you to move the focus to the desired object. 2-1.
Overview of Functions Screen Area: Names and Functions of Parts 3. Softkey Menu Bar A group of keys on the screen called by the softeys and menu bars. You can manipulate these keys by using the NAVIGATION block keys on the front panel, the mouse, or the keyboard. When a touch screen LCD (Option 016) is used, you can perform manipulations by directly touching the screen with your finger instead of using a mouse.
Overview of Functions Screen Area: Names and Functions of Parts 3-4. Highlighted Softkey Pressing and on the front panel or pressing on the keyboard causes the highlighted (selected) softkey to be executed. You can change which softkey in the menu is highlighted by turning or pressing on the keyboard. Pressing the on the front panel or by pressing key on the front panel or the panel or the key on the front 2.
Overview of Functions Screen Area: Names and Functions of Parts 4. Instrument Status Bar The instrument status bar displays the status of the entire instrument. Figure 2-5 Instrument status bar 4-1. Instrument Message/Warning Displays instrument messages and warnings. Instrument messages are displayed in gray and warnings in red. For the meanings of the instrument messages and warnings, see Appendix B, “Troubleshooting,” on page 611. 4-2.
Overview of Functions Screen Area: Names and Functions of Parts 4-6. External Reference Signal Phase Lock When the frequency reference signal is input to the “10. External Reference Signal Input Connector (Ref In)” on page 56 on the rear panel and the measurement signal of the E5070B/E5071B is phase-locked to the reference signal, ExtRef is displayed in blue. Measurement signal is phase-locked to the external reference signal.
Overview of Functions Screen Area: Names and Functions of Parts 5. Channel Window Windows for displaying traces. Because a channel corresponds to a window, it is called a channel window. When the outer frame of a channel window is displayed in light gray, the channel is the active channel (the channel for which setup is being performed). In Figure 2-2 on page 40, channel 1 (the upper window) is the active channel. To make a channel active, use or .
Overview of Functions Screen Area: Names and Functions of Parts 5-1. Channel Title Bar You can assign a title to each channel and have the title displayed on the bar. For more on setting up a channel title bar, see “Labeling a window” on page 89. 5-2. Trace Name/Measurement Parameter When an equation label is input using the equation editor, the measurement parameter changes to the equation label. For details, refer to “Entering the equation label” on page 281. 5-3.
Overview of Functions Screen Area: Names and Functions of Parts 5-5. Trace Status Area The setup for each trace is displayed here.
Overview of Functions Screen Area: Names and Functions of Parts Table 2-1 Trace status display Classification Contents inside [ ] Meaning Equation Editor Equ Equation editor: ON Equ! Equation editor: ON (The equation refers to invalid data.) 2.
Overview of Functions Screen Area: Names and Functions of Parts 5-6. Reference Line Indicators The indicators that indicate the position of the reference line for the Y-axis scale in the rectangular display format. One indicator is to the right and the other is to the left of the scale ( and ). To enter a numeric value for the position of the reference line, open the data entry bar using the keys: - Reference Position.
Overview of Functions Screen Area: Names and Functions of Parts 5-10. Channel Measurement Status Displays the update status of traces on the channel. Measurement in progress. When the sweep time exceeds 1.5 seconds, ↑ is displayed at the point on the trace. # Invalid traces. The measurement conditions have changed, but the traces on the channel currently displayed have not been updated to match the new conditions. (No display) The measurement has not been executed. 5-11.
Overview of Functions Screen Area: Names and Functions of Parts 5-18. Channel Number Indicates the channel number. 5-19. Graticule Labels Y-axis divisions in the rectangular display format. When traces in the rectangular display format are overlaid, the Y-axis divisions for the active trace are displayed. The value of the reference line (the division line between and ) is entered numerically by opening the data entry bar using the keys: + Reference Value.
Overview of Functions Screen Area: Names and Functions of Parts 5-25. Marker Indicators Indicates the positions of markers on the stimulus axis. Active marker indicator Non-active marker indicator 5-26. Statistics Data Turning on the statistics data function displays statistics data here. For more on the statistics data function, see “Determining the Mean, Standard Deviation, and p-p of the Trace” on page 274. Chapter 2 53 2.
Overview of Functions Rear Panel: Names and Functions of Parts Rear Panel: Names and Functions of Parts This section describes the names and functions of the parts on the rear panel of the E5070B/E5071B.
Overview of Functions Rear Panel: Names and Functions of Parts 1. Handler I/O Port The terminal to which an automatic machine (handler) used on a production line is connected. For more on using the handler I/O port, see the Programmer’s Guide. Connector type: 36-pin Ribbon (Centronics) connector 2. Ethernet Port Connector type: 8-pin RJ-45 connector Base standard: 10Base-T/100Base-TX Ethernet (automatic data rate selection) 3.
Overview of Functions Rear Panel: Names and Functions of Parts 8. Power Cable Receptacle (to LINE) The receptacle (outlet) to which the power cable is connected. NOTE To connect the device to a power source (outlet), use the supplied three-prong power cable with a ground conductor. The plug attached to the power cable (on the power outlet side or device side of the cable) serves as the disconnecting device (device that cuts off power supply) of the E5070B/E5071B.
Overview of Functions Rear Panel: Names and Functions of Parts 11. Internal Reference Signal Output Connector (Ref Out) A connector for outputting the internal frequency reference signal from the E5070B/E5071B. By connecting this output connector to the external reference signal input connector of another device, the device can be phase-locked to the internal reference signal of the E5070B/E5071B and used under this condition. 2.
Overview of Functions Rear Panel: Names and Functions of Parts 18. Mini-DIN Mouse Port The port to which a mini-DIN type mouse is connected. Using a mouse enables you to more efficiently perform the operations of menu bars, softkeys, and dialog boxes as well as selecting an active channel or an active trace. The mouse also enables you to move a marker or the scale reference line by using drag-and-drop operations. NOTE Be sure to only use a mouse designated for use with this instrument.
3. Setting Measurement Conditions 3 Setting Measurement Conditions This chapter describes how to set up the measurement conditions for the Agilent E5070B/E5071B Network Analyzer.
Setting Measurement Conditions Initializing Parameters Initializing Parameters The E5070B/E5071B has three different initial settings as shown in Table 3-1 below. Table 3-1 E5070B/E5071B Initial settings and methods for restoring them Initial setting Restore method Preset state • Press - OK on the front panel*1 or • Execute the :SYST:PRES command *RST state Execute the *RST command Factory default setting (how the E5070B/E5071B is set up prior to shipment from the factory) *1.
Setting Measurement Conditions Setting Channels and Traces Setting Channels and Traces The E5070B/E5071B allows you to use up to 16 channels (when the number of traces is up to 4) to perform measurement under 16 different stimulus conditions. For each channel, up to 16 traces (measurement parameters) can be displayed (when the number of channels is up to 4).
Setting Measurement Conditions Setting Channels and Traces Step 5. Press Return. The dialog box that prompts you to restart the firmware appears. Click the Yes button to restart the firmware. Setting channel display (layout of channel windows) The measurement result for each channel is displayed in its dedicated window (channel window). You cannot have a single window display the measurement results from more than one channel.
Setting Measurement Conditions Setting Channels and Traces Figure 3-1 Layout of channel windows 3.
Setting Measurement Conditions Setting Channels and Traces Setting trace display Setting the number of traces Depending on the measurement parameters of the traces displayed for each channel, the sweep necessary for each channel is executed. For more information, refer to “Sweep Order in Each Channel” on page 242. You specify the trace display by setting the number of traces (upper limit of displayed trace numbers). For example, if you set the number of traces to 3, traces 1 through 3 are displayed.
Setting Measurement Conditions Setting Channels and Traces The procedure for setting the graph layout is as follows: Step 1. Press layout. Step 2. Press or to select the channel for which you want to set the graph . Step 3. Press Allocate Traces. Step 4. Press the desired softkey to select the graph layout (refer to Figure 3-2). Figure 3-2 Graph layout 3.
Setting Measurement Conditions Setting Channels and Traces Active channel The active channel is the one whose settings can currently be changed. The window frame of the active channel is displayed brighter than the window frames of the other channels. To change the settings specific to a certain channel, you must first activate the channel. To change the active channel, use the following hardkeys: Hardkey Function Change the active channel to the next channel with the larger channel number.
Setting Measurement Conditions Setting Channels and Traces Parameter setting for each setup item (analyzer, channel, trace) Table 3-2 lists the setting parameters and indicates the setup item (analyzer, channel, or trace) that each parameter controls along with the applicable setup key(s).
Setting Measurement Conditions Setting Channels and Traces Table 3-2 Parameters and setup items they control Parameter Controlled Setup Items Analyzer Channel IF bandwidth √ Calibration √ Marker Setup Key(s) Trace - IF Bandwidth √ (*3) Analysis Fixture simulator √ (*4) - Fixture Simulator √ Time domain - Gating - Transform Parameter conversion √ - Conversion Limit test √ - Limit Test Saving and recalling data √ Macro √ System Printing/Saving display Screen/Beeper/GRIB settings/N
Setting Measurement Conditions Setting the System Z0 Setting the System Z0 NOTE This function is available with the firmware version 3.01 or greater. The procedure for setting the system characteristic impedance (Z0) is as follows: Step 1. Press . Step 2. Press Set Z0. Step 3. Enter the system Z0 using the ENTRY block keys on the front panel. 3.
Setting Measurement Conditions Setting Stimulus Conditions Setting Stimulus Conditions You can set the stimulus condition for each channel independently. Setting sweep type You can select the sweep type from the following four types. Sweep type Description Linear Sweeps frequencies in linear scale. Log Sweeps frequencies in logarithmic scale. Segment Performs a sweep with linear sweep conditions (segments) combined.
Setting Measurement Conditions Setting Stimulus Conditions Setting the Sweep Range with the Lowest and Highest Values Step 1. Press Step 2. Press or to select the channel whose sweep range will be set. . Step 3. Using the ENTRY block keys on the front panel, input the lowest value. Step 4. Press . Step 5. Using the ENTRY block keys on the front panel, input the highest value. Setting the Sweep Range with the Center Value and a Span Step 1. Press Step 2.
Setting Measurement Conditions Setting Stimulus Conditions Figure 3-4 Setting the sweep range using the marker 72 Chapter 3
Setting Measurement Conditions Setting Stimulus Conditions Turning stimulus signal output on/off You can turn on/off the stimulus signal output, but this will prevent you from performing measurement. Therefore, you will not normally use this feature. This is mainly used to turn the output back to on after it has been turned off by the power trip feature. Follow these steps to turn the stimulus signal output on/off: Step 1. Press . Step 2. Press Power. Step 3. Press RF Out.
Setting Measurement Conditions Setting Stimulus Conditions Setting power level with Auto Power Range set function When the Auto Power Range set function is effective, the proper source attenuator and power range are selected automatically, as shown by the figures below, according to the maximum frequency and maximum output power of each channel (except for instruments with option 213, 313, or 413).
Setting Measurement Conditions Setting Stimulus Conditions Figure 3-6 Available power level and selected power range with Auto Power Range set function ON (when maximum frequency is over 3 GHz (for E5071B only))*1 Setting the level or Step 2. Press . to select the channel for which you want to set the 3. Setting Measurement Conditions Step 1. Press power level. Step 3. Press Power. Step 4. Press Port Couple and select the on/off setting of the level coupling for all ports.
Setting Measurement Conditions Setting Stimulus Conditions 3. Enter the power level using the ENTRY block keys on the front panel. Correcting attenuation of power level (using power slope feature) You can use the power slope feature to correct the attenuation of a power level so that it is simply proportional to the frequency (attenuation due to cables and so on), which improves the accuracy of the level actually applied to the DUT. Turning power slope feature on/off Step 1.
Setting Measurement Conditions Setting Stimulus Conditions Setting power range manually By turning off the Auto Power Range set function, you can set the power level at the frequency sweep independently for each test port manually, within the range of −20 dBm to 10 dBm (for instruments with option 214, 314, or 414, the range of −55 dBm to 10 dBm) and at a resolution of 0.05 dB. Selecting range When option 214, 314, or 414 is installed, you can select from the following power ranges.
Setting Measurement Conditions Setting Stimulus Conditions Setting the number of points The number of points is the number of data items collected in one sweep. It can be set to any number from 2 to 1601 for each channel independently. NOTE • To obtain a higher trace resolution against the stimulus value, choose a larger number of points. • To obtain higher throughput, keep the number of points to a smaller value within an allowable trace resolution.
Setting Measurement Conditions Setting Stimulus Conditions Figure 3-7 Sweep Time and Sweep Delay Time Sweep delay is time before starting a sweep for each stimulus (source) port Step 1. Press or Step 2. Press . to select the channel for which sweep time will be set. Step 3. Press Sweep Time. Step 4. Using the ENTRY block keys on the front panel, input the desired sweep time (in seconds).
Setting Measurement Conditions Selecting Measurement Parameters Selecting Measurement Parameters The E5070B/E5071B allows users to evaluate the DUT (device under test) characteristics by using the following measurement parameters. • S-parameters • Mixed mode S-parameters This section gives the definition of S-parameters and explains how to choose their values.
Setting Measurement Conditions Selecting a Data Format Selecting a Data Format The E5070B/E5071B allows you to display measured S-parameters by using the following data formats: R Rectangular display formats • • • • • • • • • Log magnitude format Phase format Expanded phase format Positive phase format Group delay format Linear magnitude format SWR format Real format Imaginary format R Polar format R Smith chart format Rectangular display formats draw traces by assigning stimulus values (linear scale) t
Setting Measurement Conditions Selecting a Data Format Table 3-3 Eight types of rectangular display formats Type Y-axis Data Type Y-axis Unit Application Examples Log magnitude format Magnitude dB •Return loss measurement •Insertion loss measurement (or gain measurement) Phase format Phase (displayed in range from −180° to +180°) Degrees (°) •Measurement of deviation from linear phase Expanded phase format Phase (can be displayed above +180° and below −180°) Degrees (°) •Measurement of dev
Setting Measurement Conditions Selecting a Data Format Smith chart format The Smith chart format is used to display impedances based on reflection measurement data of the DUT. In this format, traces are plotted at the same spots as in the polar format. The Smith chart format allows users to select one of the following five data groups for displaying the marker response values.
Setting Measurement Conditions Selecting a Data Format Selecting a data format Use the following procedure to select a data format. Step 1. Press (or ) and which the data format will be set. Step 2. Press (or ) to select the trace for . Step 3. Press the softkey that corresponds to the desired data format.
Setting Measurement Conditions Setting the Scales Setting the Scales Auto scale The auto scale function is used to tailor each scale (scale/division and the reference line value) automatically in such a way that traces will appear at the proper size on the screen for easy observation. Single Trace Auto Scale Follow the procedure below to perform the auto scale function on a specific trace. Step 1. Press (or ) and which the auto scale function will be performed. Step 2.
Setting Measurement Conditions Setting the Scales Figure 3-11 Manual scale setup on a rectangular display format Manually setting scales on a rectangular display format Step 1. Press (or ) and which scale features will be adjusted. Step 2. Press (or ) to select the trace for . Step 3. Press the softkey that corresponds to the particular feature that needs to be adjusted. NOTE Softkey Function Divisions Defines the number of divisions on the Y-axis.
Setting Measurement Conditions Setting the Scales Manual scale adjustment on the Smith chart/polar format Manual scale adjustment on the Smith chart format or the polar format is done by using the displacement (Scale/Div of the outermost circle, Figure 3-12). Figure 3-12 Manual scale setup on the Smith chart/polar format Step 1. Press (or which the scale will be adjusted. Step 2. Press ) and (or ) to select the trace for . Step 3. Press Scale/Div. Step 4.
Setting Measurement Conditions Setting Window Displays Setting Window Displays Maximizing the specified window/trace display When using multiple channels, it is possible to maximize a specific channel window on the screen. When multiple traces are displayed in a channel window, it is also possible to maximize a specific trace displayed within that channel window. Maximizing a window Step 1. Press maximized. (or Step 2. Press to maximize the channel window.
Setting Measurement Conditions Setting Window Displays Hiding Frequency Information You can hide the frequency information from the screen in order to ensure its confidentiality or for other reasons. Hiding Frequency Information on the Screen Follow the steps below to hide frequency information on the measurement screen. Step 1. Press . Step 2. Press Frequency turn off the frequency display.
Setting Measurement Conditions Setting Window Displays Figure 3-13 Title label input dialog box Step 4. Using the keys in the dialog box, type a label and press Enter. Step 5. Press Title Label to turn on the title display. The title will appear within a frame at the top of the channel window (title bar, Figure 3-14.
Setting Measurement Conditions Setting Window Displays Setting display colors Selecting display mode You can select the display mode of the LCD display from two modes: normal display (background: black) or inverted display (background: white). In normal display, the colors of items are preset so that you can recognize them easily on the display of the instrument.
Setting Measurement Conditions Setting Window Displays Step 5. Press the softkey corresponding to the item for which you want to set the display color.
4. Calibration 4 Calibration This chapter describes the calibration processes that should be used with the Agilent E5070B/E5071B.
Calibration Measurement Errors and their Characteristics Measurement Errors and their Characteristics It is important to understand the factors contributing to measurement errors in order to determine the appropriate measures that should be taken to improve accuracy.
Calibration Measurement Errors and their Characteristics Systematic Errors Systematic errors are caused by imperfections in the measuring instrument and the test setup (cables, connectors, fixtures, etc.). Assuming that these errors are repeatable (i.e., predictable) and their characteristics do not change over time, it is possible to eliminate them mathematically at the time of measurement by determining the characteristics of these errors through calibration.
Calibration Measurement Errors and their Characteristics Directivity error (Ed) Directivity errors are caused by the fact that, in a reflection measurement, signals other than the reflection signal from the DUT are received by receiver T1 (Figure 4-1) through the directivity coupler. When a certain port is a stimulus port, this error can be defined as a constant value for each stimulus port because the state of the termination at the other ports does not change.
Calibration Measurement Errors and their Characteristics Source match error (Es) A source match error is caused when the reflection signal of the DUT reflects at the signal source and enters the DUT again. When a certain port is a stimulus port, this error can be defined as a constant value for each stimulus port because the state of the signal source switch does not change. The number of source match errors in the E5070B/E5071B is equivalent to the number of stimulus ports you use.
Calibration Measurement Errors and their Characteristics Transmission tracking error (Et) A transmission tracking error is caused by the difference in frequency response between the test receiver of a response port and the reference receiver of a stimulus port in transmission measurements. When a certain port is a stimulus port, a transmission tracking error is defined for each of the other ports.
Calibration Calibration Types and Characteristics Calibration Types and Characteristics Table 4-1 shows the different types of calibrations and the features of each method.
Calibration Calibration Types and Characteristics Table 4-1 Calibration Types and Characteristics Calibration Method Standard(s) Used Corrected Error Factor Measurement Parameters Characteristics Simplified full 3-port •Open •Short •Load •Thru Same as full 3-port calibration S11, S21, S31, S12, S22, [S32], S13, [S23], S33 (Part of thru measurement can be skipped) •High-accuracy 3-port calibration*5*1 Full 4-Port ECal module Calibration*8*1 (2-port*6/ 4-port) Following 48 error terms: •Dire
Calibration Calibration Types and Characteristics Table 4-1 Calibration Types and Characteristics Calibration Method Standard(s) Used Corrected Error Factor Measurement Parameters Characteristics Simplified 3-port •Reflection (open or short) •Thru •Line •Match Same as 3-port TRL calibration S11, S21, S31, S12, S22, [S32], S13, [S23], S33 (Part of thru (or line) and line (or match) measurement can be skipped) •High-accuracy 3-port •Reflection (OPEN or SHORT) •THRU •LINE •MATCH Following 48 erro
Calibration Checking Calibration Status Checking Calibration Status Execution status of error correction for each channel You can check the execution status of error correction for each channel with the error correction status. The error correction status is indicated in the channel status bar in the lower part of the window by the symbols in the below table.
Calibration Checking Calibration Status trace. For details on the trace status area, see “5-5. Trace Status Area” on page 48. Acquisition status of calibration coefficient for each channel You can check the acquisition status of the calibration coefficient for each channel with the calibration property. The calibration property displays the acquisition status of the calibration coefficient between test ports for each channel in matrix format.
Calibration Checking Calibration Status Step 1. Press or the calibration property display. Step 2. Press to select the channel for which you want to turn on/off . Step 3. Press Property. Each press toggles the on/off setting.
Calibration Selecting Calibration Kit Selecting Calibration Kit Before executing calibration, you need to select a calibration kit. If you use a calibration kit other than a predefined one, you need to define it. If the connector type of the standard of the calibration kit you use has polarity (the distinction between male and female), you need to change the standard class definition of the calibration kit depending on the standard you actually use.
Calibration Setting the trigger source for calibration Setting the trigger source for calibration You can set the trigger source for calibration before executing calibration. You can select it from “Internal” or “System.” Setting it to “System” allows you to use the same trigger source setting for calibration and measurement.
Calibration OPEN/SHORT Response Calibration (reflection test) OPEN/SHORT Response Calibration (reflection test) In OPEN or SHORT response calibration, calibration data are measured by connecting an OPEN or SHORT standard, respectively, to the desired test port. For frequency response, these calibrations effectively eliminate the reflection tracking error from the test setup in a reflection test using that port (Figure 4-3).
Calibration OPEN/SHORT Response Calibration (reflection test) Step 4. Select OPEN or SHORT response calibration. Softkey Function Response (Open) Displays softkeys for performing an OPEN response calibration (response calibration with an OPEN standard) Response (Short) Displays softkeys for performing a SHORT response calibration (response calibration with a SHORT standard) Step 5. Press Select Port. Step 6. Select the test port upon which you will perform OPEN/SHORT response calibration.
Calibration OPEN/SHORT Response Calibration (reflection test) Step 8. Press Open or Short to start the calibration measurement. Step 9. If an isolation calibration must be performed using a LOAD standard, follow the procedure below. a. Connect a LOAD standard to the test port (connector to which the DUT is to be connected) selected in Step 6. Figure 4-6 Connecting the LOAD standard b. Press Load (Optional) to start the measurement on the LOAD standard. Step 10.
Calibration THRU Response Calibration (transmission test) THRU Response Calibration (transmission test) In THRU response calibration, calibration data are measured by connecting a THRU standard to the desired test port. This calibration effectively eliminates the frequency response transmission tracking error from the test setup in a transmission test using that port (Figure 4-7). It is also possible to carry out an isolation calibration using a LOAD standard in the process of THRU response calibration.
Calibration THRU Response Calibration (transmission test) Softkey Function 3-1 (S31) Selects test port 3 (input) and test port 1 (output); Corresponds to the determination of S31 Selects test port 4 (input) and test port 1 (output); Corresponds to the determination of S41 Selects test port 1 (input) and test port 2 (output); Corresponds to the determination of S12 Selects test port 3 (input) and test port 2 (output); Corresponds to the determination of S32 Selects test port 4 (input) and test port 2 (ou
Calibration THRU Response Calibration (transmission test) Step 8. Press Thru to start the calibration measurement. Step 9. If an isolation calibration must be performed using a LOAD standard, follow the procedure below. a. Connect a LOAD standard to each of the two test ports (connectors to which the DUT is to be connected) selected in Step 6. Figure 4-10 Connecting the LOAD standard b. Press Isolation (Optional) to start the calibration measurement. Step 10.
Calibration 1-Port Calibration (reflection test) 1-Port Calibration (reflection test) In 1-port calibration, calibration data are measured by connecting an OPEN standard, a SHORT standard, and a LOAD standard to the desired test port. This calibration effectively eliminates the frequency response reflection tracking error, directivity error, and source match error from the test setup in a reflection test using that port (Figure 4-11).
Calibration 1-Port Calibration (reflection test) Procedure Step 1. Press calibration. Step 2. Press or to select the channel for which you want to perform the . Step 3. Press Calibrate. Step 4. Press 1-Port Cal. Step 5. Press Select Port. Step 6. Select a test port (and corresponding S parameter) on which 1-port calibration will be performed.
Calibration Enhanced Response Calibration Enhanced Response Calibration In enhanced response calibration, calibration data are measured by connecting an OPEN standard, a SHORT standard, or a LOAD standard to the output port (or a THRU standard between two ports).
Calibration Enhanced Response Calibration Figure 4-14 Connecting the Standard at Enhanced Response Calibration Procedure Step 1. Press calibration. Step 2. Press or to select the channel for which you want to perform the . Step 3. Press Calibrate. Step 4. Press Enhanced Response. Step 5. Press Ports to select the test ports on which an enhanced response calibration will be performed. Softkey display Function 2-1(S21 S11) Measure between test ports 1 and 2. selecting port 1 as the output.
Calibration Enhanced Response Calibration NOTE Softkey display Function 1-4(S14 S44) Measure between test ports 4 and 1. selecting port 4 as the output. 2-4(S24 S44) Measure between test ports 4 and 2. selecting port 4 as the output. 3-4(S34 S44) Measure between test ports 4 and 3. selecting port 4 as the output.
Calibration Full 2-Port Calibration Full 2-Port Calibration In full 2-port calibration, calibration data are measured by connecting an OPEN standard, a SHORT standard, or a LOAD standard to two desired test ports (or a THRU standard between two ports).
Calibration Full 2-Port Calibration Figure 4-16 Connecting standards in full 2-port calibration Procedure Step 1. Press calibration. Step 2. Press or to select the channel for which you want to perform the . Step 3. Press Calibrate. Step 4. Press 2-Port Cal. Step 5. Press Select Ports.
Calibration Full 2-Port Calibration calibration coefficient (pressing Done). Step 7. Press Reflection. Step 8. Connect an OPEN calibration standard to test port x (the connector to which the DUT is to be connected) selected in Step 6. Step 9. Press Port x Open to start the calibration measurement (x denotes the test port to which the standard is connected). Step 10. Disconnect the OPEN calibration standard that was connected in Step 8 and replace it with a SHORT calibration standard. Step 11.
Calibration Full 3-Port Calibration Full 3-Port Calibration In full 3-port calibration, calibration data are measured by connecting an OPEN standard, a SHORT standard, or a LOAD standard to three desired test ports (or a THRU standard between three ports).
Calibration Full 3-Port Calibration Figure 4-18 Connecting standards in full 3-port calibration Procedure Step 1. Press calibration. Step 2. Press or to select the channel for which you want to perform the . Step 3. Press Calibrate. Step 4. Press 3-Port Cal. Step 5. Press Select Ports. Step 6. Select the test ports on which you will perform full 3-port calibration. (In the procedure below, the selected test ports are denoted as x, y, and z.
Calibration Full 3-Port Calibration calibration coefficient (pressing Done). Step 7. Press Reflection. Step 8. Connect an OPEN calibration standard to test port x (the connector to which the DUT is to be connected) selected in Step 6. Step 9. Press Port x Open to start the calibration measurement (x denotes the test port to which the standard is connected). Step 10. Disconnect the OPEN calibration standard that was connected in Step 8 and replace it with a SHORT calibration standard. Step 11.
Calibration Full 4-Port Calibration Full 4-Port Calibration In full 4-port calibration, calibration data are measured by connecting an OPEN standard, a SHORT standard, or a LOAD standard to the four test ports (or a THRU standard between the four ports).
Calibration Full 4-Port Calibration Figure 4-20 Connecting standards in full 4-port calibration Procedure Step 1. Press calibration. Step 2. Press or to select the channel for which you want to perform the . Step 3. Press Calibrate. Step 5. Press Reflection. Step 6. Connect an OPEN calibration standard to test port 1 (the connector to which the DUT is to be connected). Step 7. Press Port 1 Open to start the calibration measurement. Step 8.
Calibration Full 4-Port Calibration Step 13. Repeat Step 6 to Step 11 on test port 3. Step 14. Repeat Step 6 to Step 11 on test port 4. Step 15. Press Return. Step 16. Press Transmission. Step 17. Make a THRU connection between ports 1 and 2 (between the connectors to which the DUT will be connected). Step 18. Press Port 1-2 Thru to start the calibration measurement. Step 19. Repeat Step 17 and Step 18 on ports 1 and 3. Step 20. Repeat Step 17 and Step 18 on ports 1 and 4. Step 21.
Calibration ECal (electronic calibration) ECal (electronic calibration) ECal is a calibration method that uses solid-state circuit technology. ECal offers the following advantages: • Simplified calibration process. • Shorter time required for calibration. • Reduced chance of erroneous operation. • Little degradation of performance due to wear because the ECal module employs PIN diodes and FET switches.
Calibration ECal (electronic calibration) Step 3. Press calibration. Step 4. Press or to select the channel for which you want to perform the . Step 5. Press ECal. Step 6. Press 1 Port ECal. Step 7. Perform a 1-port calibration. Softkey Function Port 1 Performs a 1-port calibration on test port 1 Port 2 Performs a 1-port calibration on test port 2 Port 3*1 Performs a 1-port calibration on test port 3 Port 4*2 Performs a 1-port calibration on test port 4 *1.
Calibration ECal (electronic calibration) Figure 4-22 Connecting ECal module (full 2-port calibration) Step 3. Press calibration. Step 4. Press or to select the channel for which you want to perform the . Step 5. Press ECal. Step 6. To enable isolation calibration, press Isolation and confirm that the display has turned ON. Step 7. Press 2 Port ECal. When using a 2-port E5070B/E5071B (Options 213 or 214), pressing this key performs a 2-port ECal. Step 8.
Calibration ECal (electronic calibration) Unknown Thru Calibration E5070B/E5071B allows you to perform thru calibration of ECal as unknown thru calibration. In this function, the thru calibraion is done with thru standard in ECal, however, the stored thru calibration data in ECal is not used. E5070B/E5071B performs thru calibration as a unknown thru calibration. For more detail on the unknown thru calibration, see “Unknown Thru Calibration” on page 153. Step 1. Press . Step 2. Press ECal. Step 3.
Calibration ECal (electronic calibration) Turning off ECal auto-detect function The ECal module automatically detects the connection between E5070B/E5071B's test ports and ECal module's ports. You can turn off this function to set ports manually. NOTE Even if the connection is wrong with the auto-detect function turned off, no error is displayed. Step 1. Press or auto-detect function. Step 2. Press to select the channel for which you want to turn off the . Step 3. Press ECal. Step 4.
Calibration Full 3-Port and Full 4-Port Calibration using 2-Port ECal Full 3-Port and Full 4-Port Calibration using 2-Port ECal A VBA macro (ECal Assistant) is pre-installed in the E5070B/E5071B to carry out a full 3-port or a full 4-port calibration using the 2-port ECal. NOTE ECal Assistant does not perform isolation calibration. Operational procedure Step 1. Connect the USB port of the ECal module to the USB port on the E5070B/E5071B with a USB cable.
Calibration Full 3-Port and Full 4-Port Calibration using 2-Port ECal Step 8. When a full 3-port calibration is carried out on an E5070B/E5071B with options 413 or 414, select the test ports to be calibrated from the drop-down list box below the 3-Port button (either 1-2-3, 1-2-4, 1-3-4, or 2-3-4). Step 9. In the Select Channel area, select the channel to be calibrated (one of channels 1 ~ 9). Step 10. Press the Next button. The dialog box shown in Figure 4-25 appears.
Calibration Full 3-Port and Full 4-Port Calibration using 2-Port ECal Figure 4-27 ECalAssistant (Connection) dialog box Step 14. Re-connect the ECal module following the instructions given in each dialog box and continue the calibration process. Step 15. When all calibration data have been collected, a dialog box with the Complete! sign appears as shown in Figure 4-28. Press the Done button to finish the calibration. If you wish to cancel the calibration, press the Cancel button.
Calibration Calibration Using 4-port ECal Calibration Using 4-port ECal The E5070B/E5071B allows you to perform calibration using the 4-port ECal module. It provides much simpler operation than when using the 2-port ECal. Especially when using a multi-port test set, calibration time and operator errors can be reduced significantly. Operational procedure To execute full 2-port calibration using the 4-port ECal module, follow these steps. Step 1.
Calibration Calibration Using 4-port ECal Step 6. When you want to turn ON the isolation calibration, press Isolation (set to ON). Step 7. Select the calibration type based on the list below. Softkey Function 1-Port ECal Selects 1-port calibration 2-Port ECal Selects full 2-port calibration 3-Port ECal*1 Selects full 3-port calibration 4-Port ECal*2 Selects full 4-port calibration Thru ECal Selects THRU calibration *1.Options 313, 314, 413, and 414 only *2.Options 413 and 414 only Step 8.
Calibration 2-port TRL calibration 2-port TRL calibration The 2-port TRL calibration function lets you measure calibration data by connecting thru, reflection (open or short), line, or match calibration standards to (between) 2 desired test ports. This calibration provides the most accurate measurement for non-coaxial parts, using 12 error terms in total for calibration in the same way as full 2-port calibration. NOTE This function is available with firmware version A.06.00 or greater.
Calibration 2-port TRL calibration Operational procedure Figure 4-31 Connection of standards in 2-port TRL calibration NOTE The shapes of actual standards differ from those shown in Figure 4-31 which are just symbols for illustration. Step 1. Press or to select the channel for which you want to execute calibration. Step 2. Press . Step 3. Press Calibrate. Step 4. Press 2-Port TRL Cal. Step 5. Press Select Ports. Step 6. Select the test ports for which you want to execute TRL 2-port calibration.
Calibration 2-port TRL calibration that will be cleared if you select the test port and execute acquisition of the calibration coefficient (by pressing Done). Step 7. Thru/Line to start the measurement of the calibration standard. Softkey Function Port x-y Thru Executes thru/line calibration for test ports x and y. Step 8. Reflect to start the measurement of the calibration standard. Softkey Function Port x Reflect Executes reflection calibration for test port x.
Calibration 3-port TRL calibration 3-port TRL calibration The 3-port TRL calibration function lets you measure calibration data by connecting thru, reflection (open or short), line, or match calibration standards to (between) 3 desired test ports. This calibration provides the most accurate measurement for non-coaxial parts, using 27 error terms in total for calibration in the same way as full 3-port calibration. NOTE This function is available with firmware version A.06.00 or later.
Calibration 3-port TRL calibration Figure 4-33 Connection of standards in 3-port TRL calibration Operational procedure NOTE The shapes of actual standards differ from those shown in Figure 4-33 which are just symbols for illustration. Step 1. Press or to select the channel for which you want to execute calibration. Step 2. Press . Step 3. Press Calibrate. Step 4. Press 3-Port TRL Cal. Step 6. Select test ports for which you want to execute TRL 3-port calibration.
Calibration 3-port TRL calibration coefficient (by pressing Done). Step 7. Thru/Line to start the measurement of the calibration standard. Softkey Function Port x-y Thru Executes thru/line calibration for test ports x and y. Port x-z Thru Executes thru/line calibration for test ports x and z. Port y-z Thru Executes thru/line calibration for test ports y and z. Step 8. Reflect to start the measurement of the calibration standard.
Calibration 3-port TRL calibration Step 10. Press Done to finish TRL 3-port calibration. At this point, the calibration coefficient is calculated and saved. The error correction function is automatically turned on. 4.
Calibration 4-port TRL calibration 4-port TRL calibration The 4-port TRL calibration function lets you measure calibration data by connecting thru, reflection (open or short), line, or match calibration standards to (between) 4 test ports. This calibration provides the most accurate measurement for non-coaxial parts, using 48 error terms in total for calibration in the same way as full 4-port calibration. NOTE This function is available with firmware version A.06.00 or later. For firmware version A.06.
Calibration 4-port TRL calibration Figure 4-35 Connection of standards in 4-port TRL calibration Operational procedure NOTE The shapes of actual standards differ from those shown in Figure 4-35 which are just symbols for illustration. Step 1. Press or to select the channel for which you want to execute calibration. Step 2. Press . Step 3. Press Calibrate. 4. Calibration Step 4. Press 4-Port TRL Cal. Step 5. Press Select Ports. Step 6. Thru/Line to start the measurement of the calibration standard.
Calibration 4-port TRL calibration Step 7. Reflect to start the measurement of the calibration standard. Softkey Function Port 1 Reflect Executes reflection calibration for test port 1. Port 2 Reflect Executes reflection calibration for test port 2. Port 3 Reflect Executes reflection calibration for test port 3. Port 4 Reflect Executes reflection calibration for test port 4. Step 8. Line/Match to start the measurement of the calibration standard.
Calibration 4-port TRL calibration Softkey 2-4 Fwd (S42) 2-4 Rvs (S24) 3-4 Line/Match 3-4 Fwd (S43) 3-4 Rvs (S34) Function Executes line/match calibration for test ports 2 and 4.Use this to perform forward-direction measurement only. Executes line/match calibration for test ports 2 and 4.Use this to perform reverse-direction measurement only. Executes line/match calibration for test ports 3 and 4. Use this to perform forward-direction measurement and reverse-direction measurement at the same time.
Calibration Simplified calibration Simplified calibration The simplified calibration calculates the calibration coefficients by skipping part of thru measurement (and line measurement for TRL measurement) that is necessary for the full 3/4-port calibration and the 3/4-port TRL calibration.
Calibration Simplified calibration NOTE For the simplified full 3-port calibration, (Optional) is displayed on the softkey of the type of thru measurement that can be omitted. For example, when the omittable thru measurement is 2-3, 2-3 (Optional) is displayed. The display after the execution of the omittable thru measurement is the same as that for a required thru measurement.
Calibration Simplified calibration is 2-3, 2-3 (Optional) is displayed. The display after the execution of the omittable thru/line measurement is 2-3 Thru/Line. For the softkey of an omittable line/match measurement, (Optional) is displayed as in the case of the omittable thru/line measurement. For example, when the omittable line/match measurement is 2-3, 2-3 (Optional) is displayed.The display after the execution of the omittable line/match measurement is 2-3 Line/Match.
Calibration Partial overwrite Partial overwrite The partial overwrite function is used to perform partial measurement after the execution of calibration, and it overwrites the calibration coefficients. There are three types of calibration coefficients: Er, Es, Ed for reflection, Et for transmission, and Ex for isolation.
Calibration Partial overwrite Step 7. Press Transmission. Step 8. Make a thru connection between test ports x and y (between the connectors to which the DUT is connected) selected in Step 6. Step 9. Press Port x-y Thru (x and y are the thru-connected port numbers) to start the re-measurement of the calibration standard. Step 10. Press Return. Step 11. Press Overwrite to finish the re-calibration for the full 2-port calibration.At this point, the calibration coefficients are re-calculated and saved.
Calibration Unknown Thru Calibration Unknown Thru Calibration Unknown Thru Calibration is the preferred THRU methord of calibrating the E5070B/E5071B to measure a non-insertable device. The major benefits of using a unknown thru calibration are : • It is easy to perform. • Provides better accuracy than defined thru and is usually better than adapter removal. • Dose not relies on existing standard definitions that may no longer be accurate.
Calibration Unknown Thru Calibration Procedure to select the standard type and define the standard factor Step 1. Press . Step 2. Press Modify Calkit. Step 3. Press Define STDs. Step 4. Press {Defined Name}. Step 5. Press STD Type. Step 6. Press Unknown Thru. Step 7. Press Offset Delay, then enter approximate offset delay value of the unknown thru standard. Procedure to define the unknown thru Step 1. Press . Step 2. Press Modify Calkit. Step 3. Press Specify CLSs. Step 4. Press Thru. Step 5.
Calibration Calibration between Ports of Different Connector Types Calibration between Ports of Different Connector Types When you perform calibration between ports of different connector types, you need to use a different calibration kit for each test port. In addition, for transmission measurement between 2 ports, you need to use adapters suitable for the connector types of both ports. For example, in order to perform full 2-port calibration between port 1 of an N-type connector and port 2 of a 3.
Calibration Calibration between Ports of Different Connector Types Operating procedure NOTE This VBA macro changes the definition of the label of calibration kit 10 (calibration kit corresponding to the lowest softkey) temporarily, performs calibration, and restores the previous definition after completing calibration. Therefore, if the VBA macro is aborted for some reason, the definition of calibration kit 10’s label may be lost.
Calibration Calibration between Ports of Different Connector Types 4. Setting Calibration Kit Select a calibration kit for each test port and each pair of test ports and select a standard to use for each reflection/transmission measurement. Step 1. Press Specify Cal Kit (2 in Figure 4-36). The Specify Cal Kit dialog box (Reflection tab) shown in Figure 4-37 appears. Figure 4-37 Specify Cal Kit dialog box (Reflection tab) Step 2.
Calibration Calibration between Ports of Different Connector Types adapter (3 in Figure 4-38). FWD Port 1 of the adapter (port 1 of the 2-port Touchstone file) is connected to the test port of the smaller port number of the E5070B/E5071B. RVS Port 2 of the adapter (port 2 of the 2-port Touchstone file) is connected to the test port of the smaller port number of the E5070B/E5071B.
Calibration Calibration between Ports of Different Connector Types based on the setting in “4. Setting Calibration Kit” on page 157 appears (Figure 4-39). Figure 4-39 Display of calibration kit setting for adapter calibration mode The calibration procedure in the adapter calibration mode is the same as that in the normal calibration except that the standard connected for each calibration data measurement differs.
Calibration Adapter Characterization Adapter Characterization To perform calibration between ports of different connector types, you have to obtain the characteristics of the adapter for use in transmission measurement in advance. The adapter characterization function lets you obtain the adapter’s characteristics (S-parameters) and save them to a 2-port Touchstone file. Use the following VBA macro to execute the adapter characterization.
Calibration Adapter Characterization Concept Adapter characterization is a function that calculates the S-parameters of an adapter based on 3 measurement results obtained by using open/short/load standards connected to the test port, via the adapter, for which 1-port calibration has been performed. This VBA macro uses test port 1 for this measurement. The S-parameters of the adapter can be calculated from the above 3 measurement results.
Calibration Adapter Characterization How to execute adapter characterization 1. Setting Stimulus Conditions Set the stimulus conditions of the channel for which you will execute the adapter characterization. For information on the setting procedure, see Chapter 3, “Setting Measurement Conditions.” 2. Performing Calibration Perform 1-port calibration for test port 1 and test port 2 in the channel for which the stimulus condition has been set.
Calibration Adapter Characterization 5. Setting Adapter Characterization Step 1. Press Setup (3 in Figure 4-41). The Setup dialog box shown in Figure 4-42 appears. Figure 4-42 Setup dialog box Step 2. Make the setting of the calibration kit for the adapter that is connected to test port 1 of the E5070B/E5071B in Cal Kit 1 (1 in Figure 4-42). Step 3. Make the setting of the calibration kit for the adapter that is connected to test port 2 of the E5070B/E5071B in Cal Kit 2 (2 in Figure 4-42).
Calibration Adapter Characterization 6. Measuring Data Measure data when each standard is connected. Step 1. Connect the adapter to test port 1 of the E5070B/E5071B. Step 2. According to the setting of Cal Kit 1, connect each standard to the adapter and then press the corresponding button (4 in Figure 4-41). When the data measurement is complete, the button turns yellow. When you calculate the S-parameters using measurement data in one direction only, the data measurement is complete here. Step 3.
Calibration Adapter Characterization Execution procedure of characterization for test fixture using probe The adapter characterization function also lets you obtain the characteristics of a test fixture that is inserted between the instrument and a DUT that cannot be connected directly to the instrument. In this case, a probe is used, and the obtained characteristics are saved to a 2-port Touchstone file.
Calibration Adapter Characterization 6. Measuring Data Step 1. Connect the probe to the end of the test fixture on the DUT side. Step 2. Depending on the setting of Cal Kit 1, connect each standard to the connector side of the test fixture and then press the corresponding button (4 in Figure 4-41). When the data measurement is complete, the button turns yellow. 7. Saving to file Press the Save button (6 in Figure 4-41) to save the calculated S-parameters to a 2-port Touchstone file. 8.
Calibration Adapter Removal-Insertion Adapter Removal-Insertion About Adapter Removal Adapter removal is a technique used to remove any adapter characteristics form the calibration plane. Usually, 2-port network analyzers removes adapter characteristics by performing two sets of full 2-port calibration as shown below. Figure 4-43 Adapter removal 4.
Calibration Adapter Removal-Insertion However, this method is not suitable for a multi-port network analyzer because it will require as many sets of full 2-port calibration as twice the number of port combinations. Hence, E5070B/E5071B uses the following adapter removal process to remove adapter characteristics. 1. Perform calibration with the adapter in use. 2. Remove the adapter from the port and measure Open, Short and Load values to determine the adapter’s characteristics. 3.
Calibration Adapter Removal-Insertion About Adapter Insertion This above described method also makes it possible to add adapter characteristics to a port with full n-port calibration. This allows you to make a measurement with the adapter. E5070B/E5071B uses the following adapter insertion process to insert adapter characteristics. 1. Perform calibration without the adapter in use. 2. Insert the adapter to the port and measure Open, Short and Load values to determine the adapter’s characteristics. 3.
Calibration Adapter Removal-Insertion Procedure for Adapter Removal / Insertion The s-parameter of a reciprocal adapter can be determined when the following data is available. • Open, Short, and Load measurements. • Actual values derived from the CalKit definitions. • An approximate length of the adapter. • Nature of the intended operation : removal or insertion. Once the s-parameter has been determined, run de-embedding and update the error coefficients.
Calibration User-characterized ECal User-characterized ECal The E5070B/E5071B allows you to execute ECal calibration with user-defined characteristics instead of the ECal characteristics defined as the factory default. This feature is called User-characterized ECal, and it is used to execute ECal calibration when an adapter is connected to the ECal module.
Calibration User-characterized ECal Storing user characteristics to the ECal module Follow these steps to measure characteristics while an adapter is connected to the ECal module and then to store them to the ECal module's built-in flash memory as user characteristics. NOTE With the E5070B/E5071B with 2/3 ports (options 213, 214, 313, or 314), you cannot measure the user characteristics of a 4-port ECal module and store them into the memory by using this VBA macro. 1.
Calibration User-characterized ECal 5. Measuring User Characteristics Step 1. Select Characterize ECAL (1 in Figure 4-47) to display the User Characteristic Measurement screen. Figure 4-47 EcalCharacterization macro (User Characteristic Measurement screen) Step 2. After connecting the adapter to the ECal module as necessary, connect each port of the ECal module and the test port of the E5070B/E5071B.
Calibration User-characterized ECal 6. Storing the User Characteristics to the Memory Step 1. When the measurement is complete, the User Characteristic Store screen shown in Figure 4-49 appears. Specify a user number (a location number in the memory where you want to store the user characteristics) using 1 in Figure 4-49.
Calibration User-characterized ECal Step 4. The dialog box shown in Figure 4-50 is displayed to confirm execution. Press OK (1 in Figure 4-50) to start storing the user characteristics. CAUTION Do not disconnect the USB cable or terminate the VBA macro by force while the VBA macro is storing data to the ECal's built-in flash memory. Doing so may damage the ECal module. Figure 4-50 Execution Confirmation screen The dialog box shown in Figure 4-51 appears while the VBA macro is storing data to memory.
Calibration User-characterized ECal Backup and recovery of ECal module's built-in flash memory Follow these steps to back up the contents of the ECal module's built-in flash memory. Step 1. Connect the USB cable between the USB port of the ECal module and that of the E5070B/E5071B. You can make this connection while the E5070B/E5071B’s power is ON. Step 2. Start the VBA macro according to “4. Starting the VBA MACRO” on page 172. Step 3.
Calibration User-characterized ECal Executing User-characterized ECal The execution procedure for the User-characterized ECal is the same as for normal ECal except that it requires the user characteristics to be selected in advance. Follow these steps to select the user characteristics. Step 1. Press calibration. Step 2. Press or to select the channel for which you want to execute . Step 3. Press ECal. Step 4. Press Characterization. Step 5. Select a user characteristic according to the list below.
Calibration User-characterized ECal Figure 4-53 ECal Characterization Information screen 1: The date when the user characteristics were measured 2: The information you entered in Characterization (1 in Figure 4-49) 3: The stimulus conditions when the user characteristics were measured 4: The information you entered in Adapter Description (3 in Figure 4-49) 5: The information you entered in Connectors (2 in Figure 4-49) 178 Chapter 4
Calibration Confidence Check on Calibration Coefficients Using ECal Confidence Check on Calibration Coefficients Using ECal NOTE This function is available with firmware version 3.50 or higher. Using the ECal module, the E5070B/E5071B lets you verify the obtained calibration coefficients to determine whether correct measurement is possible with them.
Calibration Confidence Check on Calibration Coefficients Using ECal Figure 4-54 Connecting ECal module (for verification of S21) Step 6. Press . Step 7. Press ECal. Step 8. When using an adapter to the ECal, press Characterization and then press the softkey corresponding to the characterization for the adapter you are using. Step 9. Press Confidence Check. Step 10. Compare the data trace and the memory trace and verify whether measurement is correct.
Calibration Changing the Calibration Kit Definition Changing the Calibration Kit Definition In most measurements, the user can use pre-defined calibration kits as they are. However, it may be necessary to change the definition of a calibration kit (or create a new one) when changing the pre-defined connector between male and female (e.g. from OPEN (f) to OPEN (m)*1) or when a special standard is used or a high degree of accuracy is demanded.
Calibration Changing the Calibration Kit Definition Definitions of terms The terms used in this section are defined as follows: Standard An accurate physical device, for which the model is clearly defined, used to determine system errors. With the E5070B/E5071B, the user may define up to 21 standards per calibration kit. Each standard is numbered from 1 through 21. For example, standard 1 for the 85033E 3.5-mm calibration kit is a SHORT standard.
Calibration Changing the Calibration Kit Definition Defining parameters for standards Figure 4-55 and Figure 4-56 show the parameters used in defining standards. Figure 4-55 Reflection Standard Model (SHORT, OPEN, or LOAD) Figure 4-56 Transmission Standard Model (THRU) 4. Calibration Z0 The offset impedance between the standard to be defined and the actual measurement plane. Normally, this is set to the system’s characteristic impedance.
Calibration Changing the Calibration Kit Definition determined through measurement or by dividing the exact physical length of the standard by the velocity coefficient. Loss This is used to determine the energy loss caused by the skin effect along the length (one-way) of the coaxial cable. Loss is defined using the unit of Ω/s at 1 GHz. In many applications, using the value 0 for the loss should not result in significant error. The loss of a standard is determined by measuring the delay (sec.
Calibration Changing the Calibration Kit Definition Redefining a calibration kit To change the definition of a calibration kit, follow the procedure below. Defining a new calibration kit 1. Select the calibration kit to be redefined. 2. Define the type of standard. Select one from among the OPEN, SHORT, LOAD, delay/THRU, and arbitrary impedance standards. 3. Define the standard coefficient. 4. Designate a standard class for the standard. 5. Save the data for the calibration kit that has been redefined.
Calibration Changing the Calibration Kit Definition Step 4. Press Modify Kit. To change the pre-defined connector type (e.g. OPEN(f) to OPEN (m)) skip to Step 14. Step 5. Press Define STDs. Step 6. Select the standard to be redefined from among standards numbered 1 through 21. Step 7. Press STD Type. Step 8. Select the type of standard according to the list below.
Calibration Changing the Calibration Kit Definition Step 10. Press Label and input a new label for the standard using the keypad displayed on the screen. Step 11. Press Return. Step 12. Repeat Step 6 to Step 11 to redefine all standards for which changes are necessary. Step 13. Press Return. Step 14. Press Specify CLSs. Step 15. Select the class to be redefined according to the list below. Softkey Function Sub Class Selects a subclass you want to use.
Calibration Changing the Calibration Kit Definition NOTE Select Set All to use the same standards for all test ports. Step 18. Select the standards to be registered in the class from among standards numbered 1 through 21. To change the connectors between male and female (e.g. OPEN (f) to OPEN (m)), select the appropriately labeled standards here. Step 19. Repeat Step 17 and Step 18 until classes are defined for all test ports that need to be redefined. Step 20. Press Return. Step 21.
Calibration Changing the Calibration Kit Definition Step 7. In the same way, repeat Step 2 to Step 6 to enter the definition of REFLECT to No. 2.Select SHORT for STD Type. Procedure to define Match Step 1. Press 3:No Name - Label. Step 2. Type in MATCH <2G. Step 3. Press STD Type - Load. Step 4. Set Max Frequency to 2GHz. Step 5. Press Return to return to the Define Std menu. Procedure to define Line 1/2 Step 1. Press 4:No Name - Label. Step 2. Type in LINE <7G. Step 3. Press STD Type - Thru. Step 4.
Calibration Changing the Calibration Kit Definition Step 3. Select Sub Class3 . Step 4. Press TRL Line/Match - Set All - LINE >7G - Return. Step 5. Press and check that the name you specified is selected as Cal Kit. By assigning Match and Line 1/2 to subclass 1/2/3 respectively, you can calibrate 3 standards with different frequency bands for TRL line/match calibration.
Calibration Changing the Calibration Kit Definition Setting a media type for the calibration kit You can set a media type for the standard you use. Setting procedure Step 1. Press . Step 2. Press Cal Kit, and select a calibration kit. Step 3. Press Modify Cal Kit. Step 4. Press Define STDs, and select a standard. Step 5. Press Media, and select a media type. Softkey Function Coaxial Selects coaxial as the media type. Waveguide Selects waveguide as the media type.
Calibration Changing the Calibration Kit Definition Saving and loading definition file of calibration kit You can save the definition file of the calibration kit for each standard into a file on a storage device (hard disk drive or floppy disk drive), then recall it later to reproduce the definition. Save procedure Step 1. Press . Step 2. Press Cal Kit, and select a calibration kit you want to save. Step 3. Press Modify Cal Kit. Step 4. Press Export Cal Kit... to open the dialog box. Step 5.
Calibration Changing the Calibration Kit Definition Default settings of pre-defined calibration kits The calibration kits 85033E, 85033D, 85052D, 85032F, 85032B, 85036B/E, 85031B, 85050C/D, and 85052C are pre-defined with their own default settings. 85033E 1. Short 2. Open 3. Broadband 4. Thru Label Short Open Broadband Thru STD Type Short Open Load Delay/Thru C0 [×10-15 F] 0 49.43 0 0 C1 [×10-27 F/Hz] 0 -310.13 0 0 C2 [×10-36 F/Hz2] 0 23.17 0 0 C3 [×10-45 F/Hz3] 0 -0.
Calibration Changing the Calibration Kit Definition 85052D Label STD Type 1. Short 2. Open 5. 3.5/2.92 6. 3.5/SMA 7. 2.92/SMA Short Open 3.5/2.92 3.5/SMA 3.5/SMA Short Open Open Open Open 49.433 6.9558 5.9588 13.4203 C0 [×10 -15 F] 0 C1 [×10 -27 F/Hz] 0 -310.131 -1.0259 -11.195 -1.9452 C2 [×10-36 F/Hz2] 0 23.1682 -0.01435 0.5076 0.5459 C3 [×10-45 F/Hz3] 0 -0.15966 0.0028 -0.00243 0.01594 L0 [×10-12 H] 2.0765 0 0 0 0 L1 [×10-24 H/Hz] -108.54 0 0 0 0 2.
Calibration Changing the Calibration Kit Definition 85032F 1. Short(m) 7. Short(f) 8. Open(f) Label Short(m) Open(m) Short(f) Open(f) STD Type Short Open Short Open C0 [×10-15 F] 0 89.939 0 89.939 -27 F/Hz] 0 2536.8 0 2536.8 [×10-36 2 F/Hz ] 0 -264.99 0 -264.99 -45 F/Hz3] 0 13.4 0 13.4 L0 [×10 -12 H] 3.3998 0 3.3998 0 L1 [×10 -24 H/Hz] -496.4808 0 -496.4808 0 L2 [×10-33 H/Hz2] 34.8314 0 34.8314 0 L3 [×10-42 H/Hz3] -0.7847 0 -0.
Calibration Changing the Calibration Kit Definition 85032B 1. Short(m) 3. Broadband 4. Thru Label Short(m) Open(m) Broadband Thru STD Type Short Open Load Delay/Thru C0 [×10-15 F] 0 119.09 0 0 -27 F/Hz] 0 -36.955 0 0 [×10-36 2 F/Hz ] 0 26.258 0 0 -45 F/Hz3] 0 5.5136 0 0 L0 [×10 -12 H] 0 0 0 0 L1 [×10 -24 H/Hz] 0 0 0 0 L2 [×10-33 H/Hz2] 0 0 0 0 L3 [×10-42 H/Hz3] 0 0 0 0 Offset Delay [s] 93 f 0 0 0 Offset Z0 [Ω] 49.
Calibration Changing the Calibration Kit Definition 85036B/E 1. Short(m) 3. Broadband 4. Thru Label Short(m) Open(m) Broadband Thru STD Type Short Open Load Delay/Thru C0 [×10-15 F] 0 63.
Calibration Changing the Calibration Kit Definition 85031B Label STD Type 1. Short 2. Open 3. Broadband 4. Thru Short Open Broadband Thru Short Open Load Delay/Thru F] 0 92.85 0 0 F/Hz] 0 0 0 0 C2 [×10-36 F/Hz2] 0 7.2 0 0 C3 [×10-45 F/Hz3] 0 4.
Calibration Changing the Calibration Kit Definition 85050C/D 1. Short(m) 3. Broadband 4. Thru Label Short Open Broadband Thru STD Type Short Open Load Delay/Thru C0 [×10-15 F] 0 90.48 m 0 0 -27 F/Hz] 0 763.3 m 0 0 [×10-36 2 F/Hz ] 0 -63.82 m 0 0 -45 F/Hz3] 0 6.434 m 0 0 L0 [×10 -12 H] 356.6 m 0 0 0 L1 [×10 -24 H/Hz] -33.39 0 0 0 L2 [×10-33 H/Hz2] 1.754 0 0 0 L3 [×10-42 H/Hz3] -33.
Calibration Changing the Calibration Kit Definition 85052C Label STD Type 1. Short 2. Open 5. 3.5/2.92 6. 3.5/SMA 7. 2.92/SMA 8. 2.4/1.85 Short Open 3.5/2.92 3.5/SMA 2.92/SMA 2.4/1.85 Short Open Open Open Open Open 49.433 6.9558 5.9588 13.4203 8.9843 -15 F] 0 -27 F/Hz] 0 -310.131 -1.0259 -11.195 -1.9452 -13.9923 C2 [×10-36 F/Hz2] 0 23.1682 -0.01435 0.5076 0.5459 0.3242 C3 [×10-45 F/Hz3] 0 -0.15966 0.0028 -0.00243 0.01594 -0.00112 L0 [×10-12 H] 2.
Calibration Changing the Calibration Kit Definition 85038A/F/M Label STD Type 1. Short 2. Open 3. Broadband 4. Thru Short Open Broadband Thru Short Open Load Delay/Thru F] 0 32.0 0 0 F/Hz] 0 100 0 0 C2 [×10-36 F/Hz2] 0 -50.0 0 0 C3 [×10-45 F/Hz3] 0 100 0 0 L0 [×10-12 H] 0 0 0 0 L1 [×10-24 H/Hz] 0 0 0 0 0 0 0 0 0 0 0 0 C0 [×10 -15 C1 [×10 -27 L2 [×10 L3 -33 [×10-42 H/Hz2] 3 H/Hz ] Offset Delay [s] 66.734 p 66.
Calibration Specifying Different Standard for Each Frequency Specifying Different Standard for Each Frequency Defining different standard for each frequency band This section demonstrates the procedure to define a different open standard for each frequency band, based on the following information. Table 4-6 Example of standard definition Standard Label name Frequencies to be defined Open Open 3G 1 GHz - 3 GHz Open Open 6G 3 GHz - 6 GHz Procedure Step 1. Press . Step 2. Press Cal Kit. Step 3.
Calibration Specifying Different Standard for Each Frequency Defining standard for each subclass This section demonstrates the procedure to specify a different OPEN standard for each frequency band, using subclasses #1 and #2. In this example, the standards created in “Defining different standard for each frequency band” on page 202 are used. Procedure Step 1. Press . Step 2. Press Cal Kit. Step 3. Select a calibration kit you want to use.
Calibration Specifying Different Standard for Each Frequency Disabling standard defined for a subclass The following procedure shows how to disable a standard defined for a subclass. Note that you cannot disable subclass 1 because at least one standard must exist. The following procedure shows how to disable an OPEN standard of subclass 2. Procedure Step 1. Press . Step 2. Press Cal Kit. Step 3. Select a calibration kit you want to use.
Calibration Specifying Different Standard for Each Frequency Notes on how frequency ranges are dealt when using subclasses By using several subclasses, you can set a different standard for each frequency, but note the following on frequency ranges. The following table shows possible cases and whether calibration is possible.
Calibration Specifying Different Standard for Each Frequency Example 4-2 When frequency ranges specified with subclasses do not cover a part of a measurement frequency range For a measurement frequency range (1 GHz - 8 GHz), if you execute calibration with a standard of subclass 1 (1 GHz-5 GHz) and then with a standard of subclass 2 (6 GHz-8 GHz), Done is not available for the undefined portion (5 GHz-6 GHz).
Calibration Specifying Different Standard for Each Frequency Example 4-4 When a part of a frequency range specified with a subclass lies outside the measurement frequency range For a measurement frequency range (1 GHz - 6 GHz), if you define a standard of subclass 1 (1 GHz - 5 GHz) and a standard of subclass 2 (5 GHz-8 GHz), a part of subclass 2 (6 GHz-8 GHz) lies outside the measurement frequency range, but the other part (5 GHz-6 GHz) lies within it, so calibration is possible.
Calibration Power Calibration Power Calibration The E5070B/E5071B has a calibration feature for power level output that uses the a power meter (power calibration). The power calibration function outputs a stimulus signal with a more accurate power level (closer to the set value) by measuring calibration data (power level) in advance with the power meter and sensor. Then this function performs error correction of the power level by using the calibration data.
Calibration Power Calibration Preparing power meter and sensor To execute power calibration, you need to prepare the power meter and power sensor used to acquire power calibration data. Table 4-9 shows available power meters and recommended power sensors for power calibration.
Calibration Power Calibration Preparing to control the power meter When acquiring power calibration data, the power meter is controlled via GPIB from the E5070B/E5071B. To control the power meter from the E5070B/E5071B, connect the USB port of the E5070B/E5071B and the GPIB connector of the power meter through the USB/GPIB interface as shown in Figure 4-61 and set the GPIB address of the connected power meter with the E5070B/E5071B. NOTE The USB/GPIB interface must be ready to use.
Calibration Power Calibration Setting power sensor calibration factor table NOTE Before using the power sensor calibration factor table of the E5070B/E5071B, set the calibration factor to 100% and then calibrate the power sensor. When you use the 437B or 438A as the power meter, you need to set the power sensor calibration factor table with the E5070B/E5071B. If you use a power meter other than the 437B or 438A, refer to the following table.
Calibration Power Calibration frequency is used. Hardkey Function If you select a cell and then press this key, you enter the mode that allows you to edit the cell character by character. If you change a value and then press this key, the value is entered in the cell. Moves up or down in the cell selected in the table. In the character-by-character edit mode, you can select an item or perform a stepped change of data. Moves right or left in the cell selected in the table. ... Enters a value in the cell.
Calibration Power Calibration Figure 4-62 Example of creating power sensor calibration factor table Saving power sensor calibration factor table You can save the power sensor calibration factor table as a CSV (Comma Separated Value) format file. Step 1. Press . Step 2. Press Power Calibration. Step 3. Press Sensor A Settings or Sensor B Settings. Step 4. Press Export to CSV File to open the Save As dialog box.
Calibration Power Calibration Step 5. Select the CSV format file you want to import and press the Open button to recall the power sensor calibration factor table. NOTE This operation is not guaranteed under the following two conditions: 1) you imported a CSV format file created/edited on a spreadsheet program, or 2) you imported a CSV format file that had been exported according to “Saving power sensor calibration factor table” on page 213 but then modified.
Calibration Power Calibration Selecting target port of error correction The power level error correction is executed for each channel/test port. You can set the following items for each channel/test port: • • • • • ON/OFF of error correction Setting of loss compensation Selection of the power sensor Number of power level measurements at one measurement point Calibration data Follow the steps below to select the test port for which you want to set/execute power level error correction. Step 1. Press .
Calibration Power Calibration Creating loss compensation table Follow these steps to set the loss compensation table. Step 1. Press . Step 2. Press Power Calibration. Step 3. Select a port (see “Selecting target port of error correction” on page 215). Step 4. Press Loss Compen. Step 5. Set the frequency (Frequency) and the loss (Loss) of the loss compensation table by using the following hardkeys and softkeys. NOTE Pressing - OK does not affect the current setting of the loss compensation table.
Calibration Power Calibration Saving loss compensation table You can save the loss compensation table as a CSV (Comma Separated Value) format file. Step 1. Press . Step 2. Press Power Calibration. Step 3. Select a port (see “Selecting target port of error correction” on page 215). Step 4. Press Loss Compen. Step 5. Press Export to CSV File to open the Save As dialog box. For information on the Save As dialog box, see the description in Figure 10-9, “Printers window,” on page 372.
Calibration Power Calibration Setting a tolerance for power calibration When a tolerance is set, an error message is displayed and the power calibration is aborted if the averaged measured power value during the power calibration goes outside the specified tolerance. If the power calibration is aborted, the power-level error-correction function is not turned on.
Calibration Power Calibration Figure 4-63 Connection of power sensor Step 8. Press Take Cal Sweep to start the measurement of calibration data. NOTE If the power meter GPIB address is not set correctly or if the power sensor is not connected to the specified channel, an error occurs and calibration data are not measured. You can abort the measurement by pressing Abort during measurement. When the measurement is complete, the power level error correction is automatically turned on.
Calibration Receiver Calibration Receiver Calibration The E5070B/E5071B has a function to calibrate the gain of the individual receivers in absolute value measurement. The receiver calibration function calibrates the gain of the receiver by inputting the output power of a stimulus port that has been assigned a correct value by the power calibration to the port of the receiver you need to calibrate. NOTE Receiver calibration is valid only for the parameters of absolute value measurement.
Calibration Receiver Calibration Step 3. Select a port (see “Selecting target port for error correction”). Step 4. Press Correction. Each time the key is pressed, ON and OFF switches over alternately. Selecting target port for error correction The error correction of receiver ports is performed for each channel/port, and you can set the following items for them: • • Turning on or off error correction Calibration data Step 1. Press . Step 2. Press Receiver Calibration. Step 3.
Calibration Receiver Calibration Figure 4-64 Example of connection between output port and receiver port Step 7. Press Take Cal Sweep to start measurement of calibration data. This step allows the gain of the receiver to be calibrated, turning on the error correction function automatically.
Calibration Vector-Mixer Calibration Vector-Mixer Calibration The E5070B/E5071B has a vector-mixer calibration function for use in measuring frequency conversion devices. The vector-mixer calibration allows you to measure the magnitude, phase and group delay of the mixer’s conversion loss by using in combination calibration standards (OPEN/SHORT/LOAD) and calibration mixer with an IF filter, as well as the network de-embedding function incorporated in the E5070B/E5071B.
Calibration Vector-Mixer Calibration Figure 4-65 Overview of vector-mixer calibration As shown in Figure 4-65, the vector-mixer calibration requires the characteristics data for the calibration mixer with IF filter. Measured mixer A measured mixer (DUT) signifies an unknown target mixer of measurement. However, a measured mixer meeting the requirements for a calibration mixer can be used as a calibration mixer.
Calibration Vector-Mixer Calibration analyzer on which vector calibration has been performed and then connect an OPEN, SHORT or LOAD standard to the end of the IF filter to start reflection measurement. The signals measured at the test port include the reflection signal from the mixer’s RF port, the IF signal (IF+) converted by the mixer and then reflected by the IF filter, and the IF signal (IF-) passing through the IF filer and then reflected by the calibration standard.
Calibration Vector-Mixer Calibration How to execute characterization of calibration mixer 1. Setting Stimulus Conditions Set the stimulus conditions for the channel you want to calibrate. For the necessary steps, refer to “Setting Stimulus Conditions” on page 70. You must also set the external signal source in advance. For the necessary steps, refer to “2. Setting External Signal Source” on page 318. 2.
Calibration Vector-Mixer Calibration 4. Setting IF Frequency Select IF frequency from RF+LO, RF-LO and LO-RF (2 in Figure 4-67), depending on the IF frequency of the calibration mixer. NOTE The number displayed in the Vector Mixer Characterization macro is the frequency set in the E5070B/E5071B and read from it. You must also set the minimum IF frequency at move than 0 kHz. IF BW must be set to much smaller value than IF frequency. 5. Selecting a Calibration Kit Select a calibration kit (3 in Figure 4-67).
Calibration Vector-Mixer Calibration measured mixer. In vector-mixer calibration, where the up/down conversion method is used, the power of the LO signal is distributed to the calibration mixer and the measured mixer through the power splitter. During a characteristics evaluation of the calibration mixer, the LO power level used by the drive of the calibration mixer must be equal to the LO power level with the measured mixer connected.
Calibration Vector-Mixer Calibration Figure 4-70 Saving characteristic data of calibration mixer (with IF filter) NOTE If you check the Setup Option (9 in Figure 4-67), the saved characteristic data will be set for the specified port of the active channel as the characteristic data file of the network de-embedding, and the fixture simulator function will be enabled. If unchecked, only the characteristic data will be saved. Step 3. Press the Close button (10 in Figure 4-67) to exit the macro. 4.
Calibration Vector-Mixer Calibration Characterizing calibration mixer (with IF filter) for balance mixer measurement The VBA macro (Vector Mixer Characterization) provided with the E5070B/E5071B allows you to characterize the calibration mixer (with IF filter) to be used for the balanced mixer measurement.
Calibration Vector-Mixer Calibration Set the stimulus conditions for the channel you want to calibrate. For the necessary steps, refer to “Setting Stimulus Conditions” on page 70. You must also set the external signal source in advance. For the necessary steps, refer to “2. Setting External Signal Source” on page 318. NOTE If you characterize a calibration mixer with an IF filter, we recommend that you perform full 4-port calibration in advance, since it simplifies the evaluation procedures.
Calibration Vector-Mixer Calibration the macro (1 in Figure 4-73). NOTE If failure occurs when reading the data file for the calibration mixer with IF filter, the characterization may have done by using only one port instead of using two ports. Figure 4-73 Balanced Mixer Characterization Macro Step 9.
Calibration Scalar-Mixer Calibration Scalar-Mixer Calibration The E5070B/E5071B has a scalar-mixer calibration function for measuring frequency conversion devices. Scalar-mixer calibration allows you to measure the magnitude value and reflection parameter of the mixer’s conversion loss with very high accuracy by performing calibration using calibration standards (OPEN/SHORT/LOAD) in combination as well as a power meter.
Calibration Scalar-Mixer Calibration Figure 4-74 frequency-offset error model Confirming calibration status Error correction status of each channel You can confirm the progress of error correction for each channel by viewing the error calibration status.
Calibration Scalar-Mixer Calibration Figure 4-75 Status indication for the traces in which scalar-mixer is valid NOTE In scalar-mixer calibration, the normal calibration coefficient is invalid while the frequency-offset sweep is in progress; in this case the scalar-mixer calibration coefficient is used instead. Turning the frequency-offset sweep (Frequency Offset) OFF switches over to the normal calibration coefficient; however, the information on the scalar-mixer calibration coefficient is retained.
Calibration Scalar-Mixer Calibration NOTE Softkey Function 1-4 (rev) Selects reverse direction for test port 1 and 4 1, 4 (both) Selects both directions for test port 1 and 4 3-2 (fwd) Selects forward direction for test port 2 and 3 2-3 (rev) Selects reverse direction for test port 2 and 3 2, 3 (both) Selects both directions for test port 2 and 3 4-2 (fwd) Selects forward direction for test port 2 and 4 2-4 (rev) Selects reverse direction for test port 2 and 4 2, 4 (both) Selects both dir
Calibration Scalar-Mixer Calibration Step 9. Connect the OPEN calibration standard to the test port x (connector for the DUT) you have selected in Step 6. Step 10. Press Port x @Freq x Open to start measurement of the calibration standard. Step 11. Press Port x @Freq y Open to start measurement of the calibration standard. Step 12. Disconnect the OPEN calibration standard you connected in Step 9 and then connect SHORT calibration standard in its place. Step 13.
Calibration Scalar-Mixer Calibration Figure 4-77 Connection of power sensor Step 28. Press Port x@Freq x. Step 29. Press Port x@Freq y. Step 30. Press Port y@Freq x. Step 31. Press Port y@Freq y. Step 32. Press Return. Step 33. Press Done to exit the scalar-mixer calibration. This step allows the calibration coefficient to be calculated, turning on the error correction function automatically.
Calibration Scalar-Mixer Calibration Figure 4-78 Connection of power sensor Step 4. Press Port x@Freq x. Step 5. Press Port x@Freq y. Step 6. Press Port y@Freq x. Step 7. Press Port y@Freq y. Step 8. Press Return. Step 9. Connect the test port you selected in Step 1 to the ECal module, as shown in Figure 4-79, and then press ECal & Done. NOTE Chapter 4 239 4. Calibration If the ECal module is not connected to the E5070B/E5071B, the ECal & Done menu item will not be not available.
Calibration Scalar-Mixer Calibration Figure 4-79 Connection of ECal Module Step 10. The above steps allow the calibration coefficient to be calculated, turning on the error correction function automatically.
5 Making Measurements 241 5. Making Measurement This chapter explains how to carry out measurements with the Agilent E5070B/E5071B by using the trigger function.
Making Measurements Setting Up the Trigger and Making Measurements Setting Up the Trigger and Making Measurements The E5070B/E5071B has one trigger source. When this trigger source detects a trigger signal that has occurred, a sweep is performed for channels in the “Initiate” state in the order of channel 1 to channel 16. You set the “Initiate” or “Idle” status of each channel by changing the trigger mode. For details on the trigger system, refer to Programmer's Guide.
Making Measurements Setting Up the Trigger and Making Measurements Trigger Source The trigger source generates a cue signal that initiates a measurement process. Four types of trigger sources are available as shown in Table 5-2. Table 5-2 Trigger Sources Trigger Sources Function Internal (Internal) Uses a consecutive signal generated by the firmware as a trigger source. Triggers are sent immediately following the completion of each measurement.
Making Measurements Setting Up the Trigger and Making Measurements Step 3. Press the softkey that corresponds to the desired trigger source. Softkey Function Internal Selects internal trigger source External Selects external trigger source Manual Selects manual trigger source Bus Selects bus trigger source 2. Selecting a Trigger Mode Follow the procedure below to select a trigger mode. Step 1. Press be set. Step 2. Press (or ) to select the channel for which the trigger mode will . Step 3.
Making Measurements Setting the Point Trigger Setting the Point Trigger The point trigger provides a point measurement at every trigger, and it can be used to change the trigger event to point trigger mode. Table 5-4 NOTE Trigger mode Trigger event name Function On (On Point) Point measurement is performed when trigger is applied. Off (On Sweep) Measurement is performed for all measurement points when trigger is applied.
Making Measurements Setting the low-latency external trigger mode Setting the low-latency external trigger mode In the low-latency external trigger mode, variations in delay time between the reception of a trigger and the start of a one-point measurement are decreased for point trigger measurement by using trigger pulses supplied to the external trigger input terminal.
Making Measurements Setting the low-latency external trigger mode Step 1. Press . Step 2. Press Ext Trig Delay. Step 3. Enter an external trigger delay time. External trigger delay time and point trigger interval External trigger pulses, supplied until the next measurement becomes ready after the start of a one-point measurement, are ignored, and the next trigger is generated by a pulse supplied after the completion of the one-point measurement.
Making Measurements Setting the low-latency external trigger mode The table below describes signals and times in the figure. Signal, time Description External Trig External trigger signal to be supplied. Sampling Time while the E5070B/E5071B is actually performing measurement. Index /Index signal of the handler I/O port.
Making Measurements Setting the Averaging Trigger Function Setting the Averaging Trigger Function The averaging trigger function is used to execute the sweep the number of times specified by the averaging factor with a single trigger when the sweep averaging function is ON. Averaging trigger Function On Performs the sweep the number of times specified by the averaging factor with a single trigger. Off Performs the sweep once with a single trigger.
Making Measurements Executing a Trigger Only for Active Channel Executing a Trigger Only for Active Channel Follow the procedure below to set the effective range of a trigger. Procedure to set the range Step 1. Press . Step 2. Press Trigger Scope. NOTE Softkey Function All Channel A trigger is executed for all channels. Active Channel A trigger is executed only for the active channel.
6. Data Analysis 6 Data Analysis This chapter explains how to use the analytical functions of the Agilent E5070B/E5071B.
Data Analysis Analyzing Data on the Trace Using the Marker Analyzing Data on the Trace Using the Marker About marker functions The marker can be used in the following ways: • Reading a measured value as numerical data (as an absolute value or a relative value from the reference point) • Moving the marker to a specific point on the trace (marker search) • Analyzing trace data to determine a specific parameter • Using the value of the marker to change the stimulus (sweep range) and scale (value of the
Reading values on the trace You can read the value of a marker displayed on the trace. In rectangular display format, the marker response value is always in the same data format as that of the Y-axis. On the contrary, one format for the marker response values (two values: main and auxiliary) can be selected from among several types. The selection is performed in the data format.
Data Analysis Analyzing Data on the Trace Using the Marker Step 4. Select a marker and turn it on. The softkey used to turn on a marker is also used to activate that marker.
Step 8. To turn off marker(s), press the Clear Marker Menu and then press one of the softkeys as follows: NOTE Softkey Function All OFF Turns off all of the markers on the active trace Marker 1 to Marker 9 Turns off one of markers 1 through 9 on active trace Ref Marker Turns off reference markers on active trace In the preset configuration, the marker settings on traces in a channel are coupled (Marker Couple is turned on).
Data Analysis Analyzing Data on the Trace Using the Marker Reading the Relative Value From the Reference Point on the Trace You can convert the marker reading into a relative value from the reference point. Figure 6-1 Delta marker mode Converting From a Reference Point to a Relative Value Step 1. Following Step 1 to Step 5 in “Reading values on the trace” on page 253, place the reference marker on the point to be used as the reference. Step 2. Press Ref Marker Mode to turn on the reference mode.
Reading only the actual measurement point/Reading the value interpolated between measurement points The point on the trace on which a marker can be placed differs depending on how the discrete marker mode is set up. Figure 6-2 Turning on discrete mode (Discrete ON) A marker moves only between actual measurement points. When a specific marker stimulus value is specified as a numerical value, the marker is placed at the measurement point closest to the specified value.
Data Analysis Analyzing Data on the Trace Using the Marker Setting up markers for each trace/Setting up markers for coupled operations between traces Makers can be set up and moved either in coupled operation for all traces in a channel or independently for each trace. Figure 6-3 Marker coupling Marker Couple is on (Coupling ON) Markers are set up and moved in coupled operation on all traces in a channel. Marker Couple is off (Coupling OFF) Markers are set up and moved independently for each trace.
Listing all marker values in all channels displayed You can list all of the marker values in all of the displayed channels on the screen. Turning On the Marker Table Display Step 1. Press . Step 2. Press Marker Table to turn on the marker table display. The marker table appears in the lower part of the screen (Figure 6-4). Figure 6-4 Marker table ON Chapter 6 259 6.
Data Analysis Analyzing Data on the Trace Using the Marker Specifying the display position of marker values This section describes how to specify the marker value display position as a percentage for each active trace. Figure 6-5 Marker coupling Marker Info X Pos Specifies the horizontal display position by the width of the display area as a percentage. Marker Info Y Pos Specifies the vertical display position by the height of the display area as a percentage. Operational procedure Step 1.
Aligning maker value displays This section describes how to align maker value displays. Figure 6-6 Marker coupling On (Align ON) Displays marker values to align to the display position of trace 1. Off (Align OFF) Displays marker values in the display position defined for each trace. Operational procedure Step 1. Press . Step 2. Press Annotation Options. Step 3. Press Align to toggle on/off. Chapter 6 261 6.
Data Analysis Analyzing Data on the Trace Using the Marker Displaying all marker values for displayed traces This section describes how to display all marker values for displayed traces. Figure 6-7 Marker coupling Display all (Active Only OFF) Displays all marker values for displayed traces. Displays active markers (Active Only ON) Displays markers for the active trace only. Operational procedure Step 1. Press . Step 2. Press Annotation Options. Step 3. Press Active Only to toggle on/off.
Searching for Positions that Match Specified Criteria You can search for a position that matches your specified criteria by using the Marker Search feature. Marker Search allows you to search for a position that matches any of the following criteria.
Data Analysis Searching for Positions that Match Specified Criteria Procedure to Set Search Range Step 1. Press search range. or to activate the channel for which you want to set the Step 2. If the trace coupling within the search range is off, press activate the trace for which you want to set the search range. Step 3. Press or to . Step 4. Press Search Range. Step 5. Press Search Range to turn on the Partial Search feature. Step 6. Press Start. Step 7.
Searching for the maximum and minimum measured values You can search for the maximum or minimum measured value on the trace and move a marker to that point (Figure 6-8). Figure 6-8 Searching for maximum and minimum measured values Search for maximum (Max) Move active marker to point on trace where measured value is greatest Search for minimum (Min) Move active marker to point on trace where measured value is lowest Searching for Maximum and Minimum Values Step 1.
Data Analysis Searching for Positions that Match Specified Criteria Searching for the target value (target search) The target search function enables you to move the marker to the point having the target measured value. Target and Transition Types A target is a point that has a specific measured value on the trace. Targets can be divided into the two groups shown below depending on their transition type.
Figure 6-10 Target search (when transition type is set to “both positive and negative”) Executing a Target Search Step 1. Following Step 1 to Step 4 in “Reading values on the trace” on page 253, activate the marker you are using for the target search. Step 2. Press . Step 3. Press Target. Step 4. Press Target Value and enter the target value in the entry area that appears.
Data Analysis Searching for Positions that Match Specified Criteria Searching for the peak The peak search function enables you to move the marker to the peak on the trace. Definition of the peak A peak is a measurement point whose value is greater or smaller than the adjoining measurement points on its right and left sides. Peaks are classified into the following two types depending on the difference in magnitude from the measurement points on either side of it.
Figure 6-12 Peak search (when peak polarity is positive) Executing a Peak Search Step 1. Following Step 1 to Step 4 in “Reading values on the trace” on page 253, activate the marker you are using for the peak search. Step 2. Press . Step 3. Press Peak. Step 4. Press Peak Excursion and enter the lower limit for the peak excursion value. This sets the peak search to be executed based on the definitions of the newly set lower limit for the peak excursion value and the currently set peak polarity. Step 5.
Data Analysis Determining the Bandwidth of the Trace (Bandwidth Search) Determining the Bandwidth of the Trace (Bandwidth Search) The bandwidth search is a function for determining the bandwidth of the trace, center frequency, cut-off points (on the higher frequency and the lower frequency sides), Q, and insertion loss, based on the position of the active marker. The definitions of the parameters determined through the bandwidth search are shown in Figure 6-13 and Table 6-2.
executed. The response value of this active marker itself is the insertion loss in the bandwidth search (loss). Step 2. Press . Step 3. Press Bandwidth Value and enter the defined bandwidth value in the entry area that appears. Step 4. Press Bandwidth to turn on the bandwidth search. In the upper left of the trace display, six bandwidth parameters are displayed (Figure 6-14). Figure 6-14 Bandwidth search results (defined bandwidth = −3 dB) Chapter 6 271 6.
Data Analysis Obtaining the bandwidth of a trace (notch search) Obtaining the bandwidth of a trace (notch search) The notch search function is used to obtain the bandwidth, center frequency, cutoff points (high-frequency side and low-frequency side), Q, and insertion loss of a trace based on the active marker position. The notch search function starts from the left side of the active marker position, and ends when points that meet the conditions are found.
Executing a notch Search Step 1. Place the active marker on the desired point on the trace on which the notch search is executed. The response value of this active marker itself is the insertion loss in the notch search (loss). Step 2. Press . Step 3. Press Notch Value and enter the notch value in the entry area that appears. Step 4. Press Notch to turn on the notch search. In the upper left of the trace display, six bandwidth parameters are displayed. (See Figure 6-16).
Data Analysis Determining the Mean, Standard Deviation, and p-p of the Trace Determining the Mean, Standard Deviation, and p-p of the Trace You can easily determine the statistics data for a trace (mean, standard deviation, and peak-to-peak). Figure 6-17 and Table 6-4 show the definitions for the statistics data elements.
Comparing Traces/Performing Data Math Each of the traces for which measured data is displayed is provided with an additional trace, called a memory trace, that temporarily stores measured data. You can use the memory trace to compare traces on the screen or to perform complex data math between the memory trace and measured data. The following data math operations are available: Data / Memory Divides the measured data by the data in the memory trace.
Data Analysis Comparing Traces/Performing Data Math Softkey Function Data & Mem Displays data trace and memory trace on the screen. You can now easily compare the data trace and memory trace on the screen. Off Trace is not displayed Step 8. Send the trigger to make measurements.
Performing Parameter Conversion of Measurement Results You can use the parameter conversion function to convert the measurement results of the S-parameter ( S ab ) to the following parameters.
Data Analysis Performing Parameter Conversion of Measurement Results Selecting Conversion Target Parameter Step 1. Press marker. or Step 2. Press Step 3. Press or to activate the channel on which you want to use the to activate the trace on which you want to use the marker. . Step 4. Press Conversion. Step 5. Press Function. Step 6. Press the softkey corresponding to the parameter to which you want to convert the result. When the conversion function is ON, the selected parameter is displayed in “5-5.
Using the Equation Editor Measurement data (corrected data) can be calculated by creating a equation using the equation editor. The equation editor can create a equation label, too. The created equation label is displayed on the measurement parameter area of the screen. For details, refer to “Entering the equation label” on page 281 The following describes how to use the equation editor. Procedure to start the equation editor Step 1. Press equation editor. Step 2. Press editor. Step 3.
Data Analysis Using the Equation Editor 1. A field to enter an equation and equation label An equation and equation label are input here. An equation label can be omitted. 2. Display the history of equations Registered10 equations are displayed in order of newest first. You can reset the equation by selecting an equation displayed and then pressing the OK button. NOTE Turning off the E5070B/E5071B clears the registered equation history, while preset does not clear it. 3.
Enabling/Disabling the equation editor The following procedure describes how to enable/disable the equation entered using the equation editor. Step 1. Press equation editor. Step 2. Press aditor. Step 3. Press or to activate the channel for which you want to use the or to activate the trace for which you want to use the equation . Step 4. Press Equation. Softkey Function ON Enables the equation editor. OFF Disables the equation editor.
Data Analysis Using the Equation Editor 282 Chapter 6
7. Fixture Simulator 7 Fixture Simulator This chapter explains how to use the fixture simulator functions of the Agilent E5070B/E5071B.
Fixture Simulator Overview of Fixture Simulator Overview of Fixture Simulator The fixture simulator is a function that uses software in the E5070B/E5071B to simulate various measurement conditions based on the measurement results. The six functions are as follows.
Fixture Simulator Overview of Fixture Simulator Port extension is an independent function from the fixture simulator, but if the fixture simulator function is on, data processing is automatically executed as a function of the fixture simulator to improve the data processing efficiency. (Measurement result is the same as when the fixture simulator is turned off.) Port extension moves the calibration reference location by setting an electrical delay for a single-ended port.
Fixture Simulator Overview of Fixture Simulator Balance-unbalance conversion (option 313, 314, 413, or 414) A function that uses software to convert the measurement results in an unbalanced DUT state, which are obtained by connecting the DUT to the test port of the E5070B/E5071B, into measurement results in a balanced state. Two test ports of the E5070B/E5071B are connected to one balanced port of the DUT.
Fixture Simulator Extending the Calibration Plane Using Network De-embedding Extending the Calibration Plane Using Network De-embedding Figure 7-2 Port extension and calibration plane extension using network de-embedding Using the Network De-embedding Function Step 1. Prepare a two-port Touchstone data file (.s2p format) corresponding to the network to be removed. Step 2. Press - Fixture Simulator - De-Embedding. Step 3. Press Select Port. Step 4.
Fixture Simulator Converting the Port Impedance of the Measurement Result Converting the Port Impedance of the Measurement Result The measured value obtained by using a port impedance of 50 Ω can be converted into a measured value at an arbitrary port impedance. Figure 7-3 Port impedance conversion function Converting the Port Impedance Step 1. Press - Fixture Simulator - Port Z Conversion. Step 2. Specify the port reference impedance for the required port.
Fixture Simulator Determining Characteristics After Adding a Matching Circuit Determining Characteristics After Adding a Matching Circuit Using the matching circuit embedding function, you can easily obtain the resulting characteristics after adding a matching circuit for each test port (Figure 7-4). Figure 7-4 Matching circuit function 7.
Fixture Simulator Determining Characteristics After Adding a Matching Circuit Figure 7-5 Circuit models for defining matching circuits NOTE In the 2-port matching circuit embedding or the 2-port network de-embedding feature, if the normalized impedance value specified in the user file (2-port Touchstone data file) is different from the port reference impedance setting value of the analyzer, it is automatically converted to adapt to the analyzer setting.
Fixture Simulator Determining Characteristics After Adding a Matching Circuit Step 5. Press Select Circuit. Step 6. Select a matching circuit model (Figure 7-5). Function None No matching circuit is added.
Fixture Simulator Obtaining Characteristics After Embedding/De-embedding 4-port Network Obtaining Characteristics After Embedding/De-embedding 4-port Network The 4-port network embedding/de-embedding feature allows you to add (embed) or remove (de-embed) data of your desired 4-port network defined in a 4-port Touchstone data file to or from measurement values through software-based processing. The embedding/de-embedding supports three types of connection as shown in Figure 7-6.
Fixture Simulator Obtaining Characteristics After Embedding/De-embedding 4-port Network Operational Procedure Step 1. Prepare a 4-port touchstone data file (.s4p format) corresponding to a network you want to embed/de-embed. Step 2. Press - Fixture Simulator - De-Embedding S4P. Step 3. Press Topology - Select Topology and select a network connection type (topology) from A, B, or C (Figure 7-6). connect ports 1 and 2 of the 4-port network.
Fixture Simulator Obtaining Characteristics After Embedding/De-embedding 4-port Network Step 6. Press Type (nwk1) and select a processing type for network 1. NOTE Softkey Function None Specifies no-processing (equivalent to turning off the feature) Embed Selects embedding De-Embed Selects de-embedding If you have not specified an appropriate file in Step 5, you cannot select embedding (Embed) or de-embedding (De-embed).
Fixture Simulator Evaluating Balanced Devices (balance-unbalance conversion function) Evaluating Balanced Devices (balance-unbalance conversion function) The balance-unbalance conversion function simulates a measurement under a balanced state based on measurement results obtained in an unbalanced state (Figure 7-7). This function enables you to evaluate devices with balanced ports. Be sure to set the impedances of the two unbalanced ports equal to each other.
Fixture Simulator Evaluating Balanced Devices (balance-unbalance conversion function) Figure 7-8 Types of balanced devices that can be evaluated with E5070B/E5071B In the terminology of the E5070B/E5071B, ports after the balance conversion are called logical ports (or DUT ports). You can freely assign the test ports of the E5070B/E5071B to logical ports (ports a to d in Figure 7-8).
Fixture Simulator Evaluating Balanced Devices (balance-unbalance conversion function) Measurement parameters of balanced devices Turn on the balance-unbalance conversion function to measure the following parameters. R Mixed mode S-parameter R Imbalance parameter R CMRR (Common Mode Rejection Ratio) Mixed mode S-parameter Figure 7-9 Notation of mixed mode S-parameter Figure 7-10 to Figure 7-12 show the mixed mode S-parameter when measuring each balanced device. Chapter 7 297 7.
Fixture Simulator Evaluating Balanced Devices (balance-unbalance conversion function) Figure 7-10 Mixed mode S-parameter when measuring a single-ended - balanced device Figure 7-11 Mixed mode S-parameter when measuring a balanced - balanced device Figure 7-12 Mixed mode S-parameter when measuring a single-ended - single-ended - balanced device 298 Chapter 7
Fixture Simulator Evaluating Balanced Devices (balance-unbalance conversion function) Imbalance parameter By turning on the balance-unbalance conversion function, you can select the imbalance parameter of the balanced port as the measurement parameter. Figure 7-13 to Figure 7-15 show the imbalance parameter you can select when measuring each balanced device.
Fixture Simulator Evaluating Balanced Devices (balance-unbalance conversion function) Steps for Balance-Unbalance Conversion NOTE When using three test ports, perform a full three-port calibration on the test ports to be used. When using four test ports, perform a full four-port calibration. Extend the calibration plane, if necessary, by using the port extension function or network de-embedding function.
Fixture Simulator Evaluating Balanced Devices (balance-unbalance conversion function) Softkey Function Port 2 (se) Selects a port on the analyzer from among 1, 2, 3, and 4 for connection to logical 2 (Port b in Figure 7-8) Port 3 (bal) Selects two ports on the analyzer from among 1-2, 1-3, 1-4, 2-1, 2-3, 2-4, 3-1, 3-2, 3-4, 4-1, 4-2, and 4-3 for connection to logical port 3 (Port c and Port d in Figure 7-8) Step 7. Press Return. Step 9.
Fixture Simulator Converting Reference Impedance of Balanced Port Converting Reference Impedance of Balanced Port By using the port impedance conversion function, you can specify the impedance of each test port. As a result of this conversion, the impedance of the balanced port in differential mode is set to a value twice as large as the impedance of the two unbalanced ports before conversion, and in common mode to a value one-half as large (Figure 7-17).
Fixture Simulator Converting Reference Impedance of Balanced Port Converting port reference impedance in differential mode If you turn on the differential port impedance conversion function, the port reference impedance in the differential mode is converted to an arbitrary value specified with this function instead of the value in Figure 7-17. Procedure to Turn On/Off Differential Port Reference Impedance Conversion Function Step 2.
Fixture Simulator Converting Reference Impedance of Balanced Port to set the differential port reference impedance in “Real” format. • NOTE Select Port 1 (bal) Real and Port 1 (bal) Imag, Port 2 (bal) Real and Port2 (bal) Imag, or Port 3 (bal) Real and Port 3 (bal) Imag to set the common port reference impedance in “Complex” format. Ports1, 2, and 3 refer to the logical ports 1, 2, and 3.
Fixture Simulator Determining the Characteristics that Result from Adding a Matching Circuit to a Differential Port Determining the Characteristics that Result from Adding a Matching Circuit to a Differential Port You can obtain the characteristics resulting from the pseudo addition of a balance matching circuit to a balanced port created by balance-unbalance conversion.
Fixture Simulator Determining the Characteristics that Result from Adding a Matching Circuit to a Differential Port NOTE For a network defined in a user file, it is assumed that port 1 is connected to the test port and port 2 is connected to the DUT. The setup steps are shown below. Step 1. Press - Fixture Simulator - Diff. Matching. Step 2. Press Select Port. Step 3. Press 1, 2, or 3 to select the port on the DUT to which a differential matching circuit will be added. Step 4.
Fixture Simulator Determining the Characteristics that Result from Adding a Matching Circuit to a Differential Port Step 11. If Fixture Simulator is OFF, press the key again to turn it ON. 7.
Fixture Simulator Example of Using Fixture Simulator Example of Using Fixture Simulator In this section, the fixture simulation function is explained based on an evaluation example for a DUT (balanced SAW filter) with a balanced port. Measurement circuit example for a DUT with balanced port Figure 7-20 shows an example of a measurement circuit used to evaluate a balanced SAW filter. DUT port 1 is an unbalanced port connected to source impedance Rs and input matching circuit L1.
Fixture Simulator Example of Using Fixture Simulator Figure 7-21 DUT evaluation using an actual test fixture 7. Fixture Simulator Problems in measurement with an actual test fixture Evaluating a balanced device with an actual test fixture involves the following problems: • Calibration cannot be performed at the DUT’s terminals. (A DUT’s terminals are in the test fixture and calibration standards cannot be connected to them.
Fixture Simulator Example of Using Fixture Simulator DUT evaluation using the E5070B/E5071B’s fixture simulator The E5070B/E5071B’s fixture simulator function simulates a test fixture by using internal software instead of using an actual test fixture for evaluating DUTs. Figure 7-22 shows an example connection for evaluating a DUT with the E5070B/E5071B’s fixture simulator function.
Fixture Simulator Example of Using Fixture Simulator Figure 7-23 Measurement circuit simulated by fixture simulator 7. Fixture Simulator First, the effect of an undesired network can be eliminated by port extension and/or network de-embedding. In Figure 7-22, since calibration standards cannot be connected to the DUT terminals to perform calibration, calibration should be performed at the connectors to the test fixture.
Fixture Simulator Example of Using Fixture Simulator Advantages of balanced DUT evaluation using fixture simulator Balanced device evaluation using the fixture simulator offers the following advantages: • Calibration reference plane can be easily moved to the DUT’s terminal after calibration is performed at the connectors where calibration standards can be connected. Undesired network can be removed to eliminate measurement errors (port extension, network de-embedding).
8. Frequency Offset Measurement (Option 008) 8 Frequency-Offset Measurement (Option 008) This chapter describes the frequency-translating device measurement and absolute measurement, including harmonic distortion measurement, made by using the frequency-offset mode (option 008).
Frequency-Offset Measurement (Option 008) Overview Overview Overview of Frequency-Offset Measurement The E5070B/E5071B option 008 provides a frequency-offset function and an absolute measurement function. Generally with network analyzers, the frequency must be the same both at the signal source and at the receiver, so frequency-translating devices such as mixers or converters cannot be measured.
Frequency-Offset Measurement (Option 008) Measurement of Mixers Measurement of Mixers The frequency-offset function allows you to measure frequency-translating devices such as mixers. Measurement flow Table 8-1 shows the measurement flow. Table 8-1 Flow of Mixer Measurement Item Description "1. Setting Frequency-Offset" on page 315 Sets frequency-offset "2. Setting External Signal Source" on page 318 Sets external signal source (source of LO signal) "4.
Frequency-Offset Measurement (Option 008) Measurement of Mixers Procedures for setting frequency at each port. The frequencies set at Port-n (n: any port number) should be a multiplier (Mn), divisor (Dn) and offset (On) for the basic frequency setting shown in the following formula (refer to Figure 8-1): = × Mn / Dn + On Figure 8-1 Example of frequency setting at each port NOTE The basic frequency is a frequency range set by the STIMULUS block.
Frequency-Offset Measurement (Option 008) Measurement of Mixers for each port. The use of the Multiplier/Divisor/Offset key is recommended as a rule. In this way, the offset value is automatically retained even if you have changed the setting range of the basic frequency, since the correlation of the frequencies among ports is defined as a formula. Using Start/Stop key will set M and O, which can be determined from the specified frequency and the normal frequency, while maintaining the preset D.
Frequency-Offset Measurement (Option 008) Measurement of Mixers 2. Setting External Signal Source The E5070B/E5071B allows you to control the external signal source that is connected to USB/GPIB interface. Preparations for controlling external signal source To control the external signal source from the E5070B/E5071B, you must connect its USB port and the external signal source’s GPIB connector via the USB/GPIB interface as shown in Figure 8-3.
Frequency-Offset Measurement (Option 008) Measurement of Mixers Procedures for setting external signal source GPIB address Follow the steps below to set GPIB address of the external signal source. Step 1. Press . Step 2. Press Misc Setup. Step 3. Press GPIB Setup. Step 4. Press Signal Generator Address. Step 5. Press Address. Step 6. Type the GPIB address of the external signal source you want to use.
Frequency-Offset Measurement (Option 008) Measurement of Mixers Step 5. Select the external signal source you want to use. NOTE Switching Time allows you to set the wait time after setting the external signal source’s frequency and power in ms.
Frequency-Offset Measurement (Option 008) Measurement of Mixers NOTE The basic frequency is a frequency range set by the STIMULUS block. and keys in the Follow the steps below to set the external signal source frequency. Step 1. Press - Frequency Offset to display the “Frequency Offset Menu.” Step 2. Press External Source. Step 3.
Frequency-Offset Measurement (Option 008) Measurement of Mixers Procedures for controlling external signal source Follow the steps below to control an external signal source. Step 1. Press - Frequency Offset to display the “Frequency Offset Menu.” Step 2. Press External Source to display the “External Source Menu.” Step 3. Press Frequency Format to turn it ON. The frequency setting of the external signal source (LO) is displayed as shown in Figure 8-6.
Frequency-Offset Measurement (Option 008) Measurement of Mixers Avoid Spurious Function NOTE You cannot avoid all spurious signals by using the avoid spurious function. 8. Frequency Offset Measurement (Option 008) Figure 8-8 Procedures for controlling spurious avoidance function Follow the steps below to control the spurious avoidance function. Step 1. Press - Frequency Offset to display the “Frequency Offset Menu.” Step 2. Press Avoid Spurious to turn it ON.
Frequency-Offset Measurement (Option 008) Measurement of Mixers Step 3. Use the X-Axis key to set the frequency data.
Frequency-Offset Measurement (Option 008) Measurement of Mixers 5. Implementing Mixer Calibration The E5070B/E5071B provides a vector-mixer calibration function and a scalar calibration function for measuring frequency conversion devices. For detailed information on mixer calibration, see "Vector-Mixer Calibration" on page 223 and "Scalar-Mixer Calibration" on page 233. 6. Conversion Loss Measurement Conversion loss is a typical measurement parameter of a mixer.
Frequency-Offset Measurement (Option 008) Measurement of Mixers Figure 8-10 Connection of measured mixer NOTE If two or more spectrum components are involved at a similar level, it is necessary to use a filter in order to prevent unnecessary spectrum components from entering the receiver port; in this way, you can prevent the total magnitude from reaching a damage level of 6.3 Vp-p. Procedure for setting measurement parameters Follow the steps below to set the measurement parameters.
Frequency-Offset Measurement (Option 008) Measurement of Mixers Figure 8-11 Step 11. Press to set Group Delay (group delay) to Trace 3. Step 12. Press to use Auto Scale All for trace scale optimization (Figure 8-11). Example of conversion loss measurement NOTE For additional information about the absolute measurements, see Agilent application note 1463-6 “Accurate Frequency-Translating Device Measurements by using the Frequency-Offset Mode, 5989-1420EN”. Chapter 8 327 8.
Frequency-Offset Measurement (Option 008) Measurement of Harmonic Distortion Measurement of Harmonic Distortion Using the frequency-offset function and absolute measurement function in combination allows you to measure harmonic distortion of nonlinear devices such as mixers and amplifiers. Measurement flow Table 8-4 shows the measurement flow. Table 8-4 Measurement Flow of Harmonic Distortion Item Description "1. Setting Frequency-Offset" on page 315 Sets frequency-offset function "2.
Frequency-Offset Measurement (Option 008) Measurement of Harmonic Distortion Figure 8-12 Example of measurement frequency 2. Implementing Receiver Calibration For detailed information on receiver calibration, see "Receiver Calibration" on page 220. 3. Setting Absolute Measurement Parameters The frequency-offset function has an added absolute measurement function. Procedures for setting absolute measurement parameters Follow the steps below to set the absolute measurement parameters. Step 1.
Frequency-Offset Measurement (Option 008) Measurement of Harmonic Distortion Softkey*1 Function R3 (n) Absolute measurement in Port 3, reference receiver R4 (n) Absolute measurement in Port 4, reference receiver *1. n in the parentheses is the stimulus port number. 4. Harmonic Distortion Measurement Harmonic distortion is a typical measurement parameter of nonlinear devices such as mixers and amplifiers.
Frequency-Offset Measurement (Option 008) Measurement of Harmonic Distortion Figure 8-14 Connection of DUT Procedures for setting measurement parameters NOTE Receiver calibration is required before starting absolute measurement. For detailed information on receiver calibration, see "Receiver Calibration" on page 220. Step 1. Press (or ) and (or which you want to implement absolute measurement. Step 2. Press ) to activate the trace for to display the “Measurement Menu.” Step 3.
Frequency-Offset Measurement (Option 008) Measurement of Harmonic Distortion Table 8-5 Segment Table Settings Center Span Point 3 GHz 1 MHz 5 3.5 GHz 0 Hz 1 Step 8. Set Segment Display to Freq Base. Step 9. Press Sweep Type to select Segment. Step 10. Press to use Auto Scale All for trace scale optimization (Figure 8-15).
9. Analysis in Time Domain (Option 010) 9 Analysis in Time Domain (Option 010) This chapter describes how to deal with responses in the time domain by using the time domain function (Option 010).
Analysis in Time Domain (Option 010) Overview Overview Overview of time domain measurement The E5070B/E5071B Option 010 provides the time domain function, which is used to mathematically transform waveforms in the frequency domain that can be measured with a general network analyzer to waveforms in the time domain. Figure 9-1 shows waveforms in the frequency domain and in the time domain when using the same cable.
Analysis in Time Domain (Option 010) Overview Time domain function of E5070B/E5071B The time domain function of the E5070B/E5071B is divided into the following two types of functions. You can use them at the same time. • Transformation function Transforms measurement data in the frequency domain to data in the time domain. For more information, refer to “Transformation to time domain” on page 336. • Gating function Deletes unnecessary data in the time domain from original data in the time domain.
Analysis in Time Domain (Option 010) Transformation to time domain Transformation to time domain The transformation function lets you transform a response in the frequency domain to the corresponding response in the time domain. Measurement flow Table 9-1 shows the measurement flow of this transformation. Table 9-1 Flow of transformation to time domain Item Description “Selecting a type” on page 337 Selects the transformation type from the following.
Analysis in Time Domain (Option 010) Transformation to time domain Selecting a type There are two types of transformation to the time domain: band pass and low pass. The appropriate transformation type depends on the DUT. Comparison between band pass mode and low pass mode Item Appropriate DUT Input signal Band pass DUTs that do not operate with dc current such as band pass filters. You can simulate the response to the impulse signal.
Analysis in Time Domain (Option 010) Transformation to time domain Impulse signal and step signal The E5070B/E5071B lets you simulate the response from the DUT to two types of signals: impulse signal and step signal. The impulse signal is a pulse-shape signal in which the voltage rises from 0 to a certain value and returns to 0 again. The pulse width depends on the frequency sweep range. The step signal is a signal in which the voltage rises from 0 to a certain value.
Analysis in Time Domain (Option 010) Transformation to time domain Setting the window Because the E5070B/E5071B transforms data within a finite frequency domain to data in the time domain, an unnatural change of data at the end points within the frequency domain occurs. For this reason, the following phenomena occur. • Width of impulse signal and rise time of step signal A time width occurs in the impulse signal and a rise time occurs in the step signal.
Analysis in Time Domain (Option 010) Transformation to time domain Operation Step 1. Press (or ) and which you want to set the window. Step 2. Press (or ) to activate a trace for and then press Transform to display the “Transform” menu. Step 3. Press Window and then select a window type. NOTE Softkey Function Maximum Sets the window type to maximum. β of the Kaiser Bessel function is set to 13. Normal Sets the window type to normal. β of the Kaiser Bessel function is set to 6.
Analysis in Time Domain (Option 010) Transformation to time domain Effect of frequency sweep range on response resolution Figure 9-3 shows an example when measuring the same cable while changing the sweep span. When measured in a narrower sweep range, the overlap between two peaks is larger than when measured in a wider sweep range. By performing measurement in a wider sweep range, adjacent peaks can be clearly separated, which means that the response resolution is smaller.
Analysis in Time Domain (Option 010) Transformation to time domain Effect of the window function on the response resolution Lowering the sidelobe level with the window function elongates the width of the impulse signal and the rise time of the step signal.
Analysis in Time Domain (Option 010) Transformation to time domain Measurement range In the time domain function, the measurement range means the range within which the response can be measured without repetition. The repetition of the response occurs because measurement in the frequency domain is performed discretely instead of continuously. The measurement range is inversely proportional to the frequency difference between adjacent measurement points.
Analysis in Time Domain (Option 010) Transformation to time domain resolution and the measurement range.
Analysis in Time Domain (Option 010) Transformation to time domain Setting display range The E5070B/E5071B has the following limitations on the display range you can set. Lower limit − T span *1 Upper limit T span *1 The number of response points displayed on the graph is the same as the number of points regardless of the response resolution. Operation Step 1. Press (or ) and which you want to set the display range. (or ) to activate the trace for - Press Transform to display the “Transform” menu.
Analysis in Time Domain (Option 010) Transformation to time domain NOTE The following requirements must be met to enable the transformation function. • • The sweep mode is the linear sweep. There are three or more measurement points.
Analysis in Time Domain (Option 010) Deleting Unnecessary Data in Time Domain (gating) Deleting Unnecessary Data in Time Domain (gating) Measurement Flow Table 9-1 shows the measurement flow, and Figure 9-8 shows the change in the waveform at each step of the flow. Table 9-5 Measurement flow Item Description 1. Measurement in frequency domain Executes measurement in frequency domain 2.
Analysis in Time Domain (Option 010) Deleting Unnecessary Data in Time Domain (gating) Operational procedure Step 1. Press (or ) and which you want to set the gate type. (or ) to activate the trace for - Press Gating to display the “Gating” menu. Step 2. Step 3. Press Type to toggle between band pass (Bandpass) and notch (Notch). Setting gate shape The gate is a filter whose shape looks like a band pass filter. There are several parameters that indicate the gate shape.
Analysis in Time Domain (Option 010) Deleting Unnecessary Data in Time Domain (gating) Operational procedure Step 1. Press (or ) and which you want to set the gate shape. (or ) to activate a trace for - Press Gating to display the “Gating” menu. Step 2. Step 3. Press Shape and then select the gate shape from the following. Softkey Function Maximum Sets the gate shape to “maximum.” Normal Sets the gate shape to “normal.” Wide Sets the gate shape to “wide.
Analysis in Time Domain (Option 010) Deleting Unnecessary Data in Time Domain (gating) range. You can set the center and span by dragging and dropping flags indicating the gate range (Figure 9-10). NOTE Enabling gating function When you enable the gating function, data within the specified range is deleted. When the transformation function is enabled, the flags indicating the gate range is displayed as shown in Figure 9-10. NOTE In Figure 9-10, the gate type is set to band pass.
Analysis in Time Domain (Option 010) Characteristics of Response in Time Domain Characteristics of Response in Time Domain This section describes masking and the identification of the mismatch type, which are important for analyzing the response in the time domain. Masking Masking is a phenomenon in which a mismatch at a location near the calibration surface affects the response at the next mismatch location.
Analysis in Time Domain (Option 010) Characteristics of Response in Time Domain Identifying mismatch type The transformation in the low pass mode simulates the response in the TDR measurement. In addition to mismatch locations, the response includes information on the mismatch type. Figure 9-12 shows each mismatch type and the response waveform corresponding to it. In the low pass mode, you can simulate the response of the step signal and the impulse signal.
10 Data Output 353 10. Data Output This chapter explains the concepts behind saving/recalling internal data and printing the information that is displayed on the screen. Procedures for performing these tasks with the Agilent E5070B/E5071B are also given.
Data Output Saving and Recalling Instrument State Saving and Recalling Instrument State You can save the instrument state of the E5070B/E5071B into a file on mass storage (hard disk drive or floppy disk drive) and then recall it later to reproduce that state. You can select the stored data from the following four types.
Data Output Saving and Recalling Instrument State Y: Recall is possible. N: Recall is impossible. *1.Includes E5070A equipped with power sweep feature. *2.Includes E5071A equipped with power sweep feature.
Data Output Saving and Recalling Instrument State • NOTE When Files that are saved with different system spec. versions include calibration data, only states and trace data are recalled. If you recall an incompatible file, an error occurs and the device recovers to the presetting. Save procedure Selecting Content to be Saved NOTE This setting takes effect both when saving the entire instrument state into a file and when saving the instrument state for each channel into memory. Step 1. Press . Step 2.
Data Output Saving and Recalling Instrument State Saving Instrument State Follow the procedure below to save internal data from the E5070B/E5071B. Step 1. Press . Step 2. Press Save State. Step 3. When you want to use a pre-defined file name (State01.sta - State08.sta, Autorec.sta) Press State01 - State08 or Autorec. NOTE If “A:\Autorec.sta” or “D:\Autorec.sta” is found on the system at startup, the E5070B/E5071B is automatically configured using the saved settings.
Data Output Saving and Recalling Instrument State On the E5070B/E5071B, the following drives are available for saving/recalling files. Select the appropriate drive from the Save In pull-down menu shown in Figure 10-1 Drive Description 3.5” Floppy [A:] Select this drive when saving or recalling a file to/from a floppy disk*1. [D:] Select this drive when saving or recalling a file to/from the hard disk drive (D drive). *1.When using the built-in floppy disk drive on the E5070B/E5071B, insert a 1.
Data Output Saving and Recalling Instrument State Recall Procedure Follow the procedure below to recall internal data from the E5070B/E5071B. NOTE If you recall a file that includes traces (its content was set to State &Trace or All when it was saved), the trigger source is automatically set to Manual. Step 1. Press . Step 2. Press Recall State. Step 3. When you want to recall State01.sta - State08.sta, Autorec.sta Press State01 - State08 or Autorec. When you want to recall other files 1.
Data Output Saving and Recalling Instrument State Recall Procedure Using “Recall by File Name” Feature You can use the recall feature with the Recall by File Name softkey for files you have named freely and saved in the D:\State folder. This function lets you recall a file you have named freely and saved by simple softkey operation, eliminating annoying operation using the Open dialog box.
Data Output Saving/Recalling Instrument State for Each Channel into/from Memory Saving/Recalling Instrument State for Each Channel into/from Memory The E5070B/E5071B allows you to save/recall the instrument state for each channel independently. This function allows you to save the instrument state of the active channel independently into one of four registers (A to D, volatile memory) and to recall the instrument state from the register to restore it as the state of the currently active channel.
Data Output Saving Trace Data to a File Saving Trace Data to a File Saving data as a CSV file The E5070B/E5071B allows the user to save data for the active trace on the active channel to a CSV file (file extension *.csv) and to load the data into PC application software for further processing. Trace data are saved in the format shown below. Example 10-1 Example of saved trace data "# Channel 1" "# Trace 1" Frequency, Formatted Data, Formatted Data +3.00000000000E+005, +1.41837599227E-002, +1.
Data Output Saving Trace Data to a File Saving data in Touchstone format You can also save trace data of a E5070B/E5071B active channel to a Touchstone format file, based on 1- to 4-port models. Touchstone file data format You can save data in “log magnitude - angle”, “linear magnitude - angle”, or “real number - imaginary number.” When AUTO is selected, the data format is automatically set according to the display format of the active trace.
Data Output Saving Trace Data to a File Table 10-3 Header of s4p !Agilent Technologies,E5071B,, !Date !Data & Calibration Information !Freq Sww:Method(Stat) Sxw:Method(Stat) Syw:Method(Stat) Szw:Method(Stat) • • Swz:Method(Stat) Sxz:Method(Stat) Syz:Method(Stat) Szz:Method(Stat) # Hz S FMT R Z0 Each item has the following meaning. Sww to Szz S parameters of the selected test port; corresponds in ascending order, beginning with 1 to w.
Data Output Saving Trace Data to a File Figure 10-5 2-port Touchstone file Figure 10-6 3-port Touchstone file 10.
Data Output Saving Trace Data to a File Figure 10-7 4-port Touchstone file Restrictions when saving data in Touchstone format The following restrictions apply when saving measurement data into Touchstone format. • When both fixture simulation and port impedance conversion are on, all Z0 of the ports to be saved must be set to the same value. If Z0 is different among the ports, no error occurs, but only the Z0 of the smallest port number is output to the header.
Data Output Saving Trace Data to a File Softkey Function LogMag/Angle Select “log magnitude - angle” data format LinMag/Angle Select “linear magnitude - angle” data format Real/Imaginary Select “real - imaginary number” data format *1. When the display format of the active trace is set to one other than log magni- tude format (LogMag), linear magnitude format (LinMag), or real-imaginary number format (Real/Imag), the data format is automatically set to real-imaginary number. Step 5.
Data Output Saving the Screen Image to a File Saving the Screen Image to a File Along with printing, the E5070B/E5071B allows the user to save screen images as bitmap (.bmp) or portable network graphics (.png) files. Saved files can be loaded into PC application software for further processing. Saving the Screen Image to a File Follow the procedure below to save a screen image to a file. Step 1. Display the screen to be saved as a file.
Data Output Organizing Files and Folders Organizing Files and Folders You can organize files and folders (copy, move, delete, rename, or format a floppy disk) with Windows© Explorer©. Figure 10-8 Windows Explorer NOTE Do not modify any files and folders in drives other than drive A: and drive D:. Doing so will cause malfunctions. To Open Windows Explorer Step 1. Press . Step 2. Press Explorer.... To Copy a File or Folder Step 1. Select a source file or folder in Windows Explorer. Step 2.
Data Output Organizing Files and Folders To Move a File or Folder Step 1. Select a source file or folder in Windows Explorer. Step 2. Select Edit - Cut from the menu bar. Step 3. Open the destination folder. Step 4. Select Edit - Paste from the menu bar. To Delete a File or Folder Step 1. Select a file or folder you want to delete in Windows Explorer. Step 2. Select Edit - Delete from the menu bar. To Rename a File or Folder Step 1. Select a file or folder you want to rename in Windows Explorer. Step 2.
Data Output Printing Displayed Screen Printing Displayed Screen By connecting a printer to the parallel port or USB port of the E5070B/E5071B, you can print the displayed screen of the E5070B/E5071B. Supported printers For the latest information of the supported printers for the E5070B/E5071B, refer to “Printer Compatibility“ of http://www.agilent.com/find/ena_support/. NOTE The drivers for all supported printers at the time of shipment are installed in the E5070B/E5071B.
Data Output Printing Displayed Screen Printed/saved images The display image saved in the volatile memory (clipboard) is printed/saved. If no image is saved in the clipboard, the image displayed at the time of print execution is printed/saved. Saving image to clipboard also has a screen capture feature. When you press , the image displayed on the screen immediately before pressing is saved in the clipboard. NOTE The image in the clipboard is cleared when you execute print/save.
Data Output Printing Displayed Screen Step 6. The printer with the check mark ( ) on its icon is selected as the default printer for printing. If you want to change it, select (highlight) the icon of your preferred printer in the Printers window and then click Set as Default Printer (1 in Figure 10-10) in the File menu. Figure 10-10 File menu in Printers window Step 7. Click Printing Preferences... (2 in Figure 10-10) in the File menu. The Printing Preferences dialog box for the selected printer appears.
Data Output Printing Displayed Screen Executing print Follow these steps to print the screen information: Step 1. Display the screen you want to print. Step 2. Press to save the currently displayed screen onto the clipboard. Step 3. As necessary, press Invert Image to toggle between [OFF] for printing in colors close to the actually displayed screen and [OFF] for printing in inverse colors. Step 4. Press Print to start printing. To cancel the printing in progress, press Abort Printing.
11. Limit Test 11 Limit Test This chapter describes the concepts behind the limit test , the ripple test, and the bandwidth test. It also explains how to perform it using the Agilent E5070B/E5071B.
Limit Test Limit Test Limit Test The limit test feature allows you to set the limit line for each trace and then perform the pass/fail judgment for the measurement result. Concept of limit test The limit test is a function to perform pass/fail judgment based on the limit line you set with the limit table.
Displaying judgment result of limit test Judgment result of measurement points and trace Measurement points that fail are displayed in red on the screen. The judgment result of the trace is indicated by Pass or Fail displayed in the upper right part of the graph.
Limit Test Limit Test Defining the limit line To use the limit test, you must first define the limit line. You can define a limit table for each trace, and you can define up to 100 limit lines (segments) in a limit table. Defining a segment The following steps describe how to define a segment. Step 1. Press will be used. Step 2. Press used. Step 3.
Figure 11-3 Limit table Step 6. Press Add to add a segment to the limit table and then specify the segment parameter values shown below.
Limit Test Limit Test Figure 11-4 Example of limit lines (when limit table is set as Figure 11-3) NOTE You can define a limit line that is able to freely overlap the stimulus range of another limit line. Defining one limit line having the same type as a second limit line whose stimulus range overlaps with the first one results in two or more limit values at the same measurement point.
Saving/calling the limit table You can save the limit table to a file that you can then freely bring up on the screen later and use. You can import a file saved in CSV format (extension: *.csv) into spreadsheet software on a PC for later use (a numerical value will be saved as strings that include its unit). The limit table is saved in the following format. Example 11-1 Limit Table Saved in CSV Format "# Channel 1" "# Trace 1" Type, Begin Stimulus, End Stimulus, Begin Response, End Response MAX, 200.
Limit Test Limit Test Turning the limit test ON/OFF You can set the limit test ON/OFF for each trace individually. Setting the limit test ON/OFF The following steps explain how to set the limit test ON/OFF. Step 1. Press will be used. or Step 2. Press used. Step 3. Press or to activate the channel on which the limit test function to activate the trace on which the limit test function will be to display the Analysis menu. Step 4. Press Limit Test to display the Limit Test menu.
Limit line offset By adding a certain offset to the limit value, you can adjust the limit line so that it conforms to the device output. Step 1. Press or limit test function. Step 2. Press function. Step 3. Press or to activate the channel to which you want to apply the to activate the trace to which you want to apply the limit test to display the Analysis menu. Step 4. Press Limit Test to display the softkeys for the limit test. Step 5.
Limit Test Limit Test Figure 11-6 Amplitude offset Figure 11-7 Marker amplitude offset Initializing the limit table The following operations initialize the limit table.
Ripple Test Independently of the limit test, you can evaluate the measurement results on a pass/fail basis by setting a limit for the ripple. This function is called the ripple test. Concept of ripple test The ripple test is a function for evaluating the results on a pass/fail basis based on the ripple limit, which is set using the ripple limit table. You can specify up to 12 frequency bands, which permits a test for each frequency band.
Limit Test Ripple Test Displaying ripple test results Measurement point and test results Failed measurement points will be displayed in red on the screen. The test result for the trace will be indicated as Pass or Fail in the upper-right area of the graph. You can also display the ripple value at the selected frequency band. The result will be displayed as Ripln:Pass (or Fail) for each trace. n denotes the trace number.
Step 1. Press . Step 2. Press Ripple Limit. Step 3. Press Fail Sign. This menu toggles between on and off. In addition to the screen, the following features also let you confirm the test results: • Beep notifying that the result was “fail” • Status register (for further information, see the Programmer's Guide) Configuring ripple limit You must configure the ripple limit before you can use the ripple test function.
Limit Test Ripple Test Figure 11-9 Ripple limit table Step 6. Press Add to add a frequency band to the ripple limit table and then specify the following parameters for the frequency band: Parameter Description Type Selects a frequency type, either ON or OFF. ON Band used for the ripple test. OFF Band not used for the ripple test. Begin Stimulus*1*2 Specifies the start point for the stimulus value in the ripple test.
Figure 11-10 Example of ripple limit configuration (for limit table in Figure 11-9) NOTE The individual frequency bands for the ripple test can overlap each other; in this case, the ripple limit test is performed for each frequency band. NOTE Even if the E5070B/E5071B's span value is set to zero, you must enter a parameter for both Begin Stimulus and End Stimulus.
Limit Test Ripple Test In the first line, the channel number for the active channel at the time of file saving will be output. In the second line, the trace number for the active trace at the time of file saving will be output. The third line is a header indicating the segment items that are output from the fourth line onward. From the fourth line onward, the segment data are output. Operational procedure Follow the steps below to save/recall the ripple limit table.
Step 4. Press Ripple Limit to display the Ripple Limit menu. Softkey Function Ripple Test Sets the ripple test ON/OFF. Ripple Limit Sets the ripple limit line display ON/OFF. Ripple Value Sets how the ripple values are displayed. Available settings are off, absolute value (difference between maximum and minimum values within the band) display, and margin (difference between absolute value of ripple and ripple limit) display.
Limit Test Bandwidth Test Bandwidth Test The bandwidth test function can be used for testing bandwidth for the band-pass filters. The bandwidth test finds the peak of a signal in the passband and locates a point on each side of the passband at an amplitude below the peak specified in test setup. The frequency between these two points is the bandwidth of the filter. Then the obtained bandwidth is compared to miminum and maximum allowable bandwidth that you specify beforehand.
Displaying Bandwidth Test Results Test Result for Trace The test result of the trace will be indicated in the upper-right area of the graph for each trace, following BWn:. "n" denotes the trace number (see Figure 11-12). The results are shown as Pass, Wide, Narrow, or >Span (Fail). You can also display the bandwidth value. For information on how to display the results, see “Turning On/Off Bandwidth Test and Displaying Results” on page 394.
Limit Test Bandwidth Test • Status register (for further information, see the Programmer's Guide) Set up bandwidth test You must set up the bandwidth threshold and the upper and lower limits before you can use the bandwidth test function. You can specify the threshold, upper limit, and lower limit for each trace. Operational procedure Follow the steps below to set up the bandwidth test. Step 1. Press or bandwidth test function. Step 2. Press or bandwidth test function. Step 3.
12. Optimizing Measurements 12 Optimizing Measurements This chapter describes how to optimize your measurements when using the Agilent E5070B/E5071B.
Optimizing Measurements Expanding the Dynamic Range Expanding the Dynamic Range The dynamic range is the finite difference between the maximum input power level and the minimum measurement power level (noise floor) of the analyzer. In evaluating a characteristic accompanied by a large change in the amplitude (the pass band and stop band of a filter, for example), it is important to increase the dynamic range.
Optimizing Measurements Expanding the Dynamic Range Turning on Sweep Averaging Using sweep averaging also enables you to reduce the effects of random noise on measurements. Sweep averaging averages data from each point (vector quantity) based on the exponential average of a continuous sweep weighted by the averaging factor specified by the user. Sweep averaging is expressed in Equation 12-1.
Optimizing Measurements Reducing Trace Noise Reducing Trace Noise Any of the following methods can be used to lower the trace noise. • Turning on sweep averaging • Turning on smoothing • Narrowing the IF bandwidth For more about sweep averaging and the IF bandwidth, see “Turning on Sweep Averaging” on page 397 and “Narrowing the IF bandwidth” on page 396. Turning on Smoothing Smoothing can be used to reduce noise that has relatively small peaks.
Optimizing Measurements Reducing Trace Noise Figure 12-4 Effects of smoothing (group delay format) 12. Optimizing Measurements Setting up smoothing Set up the smoothing operation by following the steps below. Step 1. Press (or ) and which smoothing will be defined. Step 2. Press (or ) to activate the trace on . Step 3. Press Smo Aperture. Step 4. Change the smoothing aperture (%) in the data entry area. Step 5. Press Smoothing to turn ON smoothing.
Optimizing Measurements Improving Phase Measurement Accuracy Improving Phase Measurement Accuracy This section describes the following functions that can be used to improve phase measurement accuracy. • • • “Electrical Delay” on page 400 “Setting port extensions” on page 403 “Phase offset” on page 402 Electrical Delay Electrical Delay is a function that adds or removes a pseudo-lossless transmission line with a variable length corresponding to the receiver input.
Optimizing Measurements Improving Phase Measurement Accuracy Step 3. Press . Step 4. Press Marker → Delay to set the electrical delay to the group delay value at the position of the active marker (a value smoothed with the aperture of 20% regardless of the smoothing setting). NOTE An absolute value is used regardless of the reference marker mode’s on/off status. 12.
Optimizing Measurements Improving Phase Measurement Accuracy Phase offset Phase offset is a function used to add or subtract a predetermined value relative to the frequency to and from the trace. Using this function enables you to simulate the phase offset occurring as a result of, say, adding a cable. The phase offset can be specified from −360° to +360°. Using the Phase Offset Function Step 1. Press (or ) and which you want to specify the phase offset. Step 2. Press (or ) to activate the trace for .
Optimizing Measurements Setting Port Extensions and Loss Values Setting Port Extensions and Loss Values Setting port extensions Port extension corrects the electrical delay of each test port (phase shift) only. It cannot remove errors caused by the loss in and incorrect matching of cables, adapters, or test fixtures. NOTE You can define port extension channel by channel. Setting port extension for one particular channel does not affect other channels. Operational procedure Step 1. Press extension.
Optimizing Measurements Setting Port Extensions and Loss Values Setting loss values In addition to port extension, you can set loss values for each port. By correcting loss due to port extension, more accurate measurement results are obtained. There are two types of loss value settings: loss values at two frequency points for a specified port, and a DC loss value. You can make these settings at the same time for each port. NOTE You can set loss values channel by channel.
Optimizing Measurements Setting Port Extensions and Loss Values 2. Setting a DC loss value Step 1. Press loss value. Step 2. Press or to activate the channel for which you want to set a DC . Step 3. Press Port Extensions. Step 5. Press Select Port to select the port for which you want to set a DC loss value. Step 6. Press Loss at DC, and enter a DC loss value. Step 7. If you want to set a DC loss value for other ports, repeat Step 5 to Step 6. Enabling port extensions and loss values Step 1.
Optimizing Measurements Setting Port Extensions and Loss Values Using the auto port extension function The auto port extension function measures port extension and loss values for each port using the OPEN/SHORT standard connected to the port, automatically calculates them, and set them. When the auto port extension function is completed, the port extensions and loss values are updated to the calculated values. NOTE You can use both open and short measurement values in the auto port extension function.
Optimizing Measurements Setting Port Extensions and Loss Values Step 5. Method to set the frequencies used for calculation. Function Current Span Executed using the frequency range set currently. Active Marker Executed using the frequency at the active marker.*1 In this case, the result is applied to Loss1. Loss2 is ignored. User Span Executed using a start value and a stop value you set. *1.Even if the active marker has been disabled, it is automatically enabled. Step 6.
Optimizing Measurements Setting Port Extensions and Loss Values 3. Specifying a loss value as a calculation target Specify whether you want to include a loss value in the calculation result. Step 1. Press port extension. Step 2. Press or to activate the channel for which you want to set auto . Step 3. Press Port Extensions. Step 4. Press Auto Port Extension. Step 5. Press Include Loss to turn it on. NOTE Softkey Function ON Calculates a loss value. OFF Does not calculate a loss value.
Optimizing Measurements Setting Port Extensions and Loss Values 5. Measuring the OPEN/SHORT standard and executing calculation Calculate port extensions and loss values based on the calculation results using the OPEN/SHORT standard. Step 2. Press or to activate the channel for which you want to set auto . Step 3. Press Port Extensions. Step 4. Press Auto Port Extension. Step 5. If you use the OPEN standard, press Measure OPEN, and select the port(s) for which you want to execute measurement.
Optimizing Measurements Setting Port Extensions and Loss Values NOTE If you execute both open measurement and short measurement, the average of the calculation results is reflected to the port extension and loss value. 6. Deleting the result of open/short measurement When you exit from the softkey menu in the same level after open/short measurement, the measurement results are deleted. Note that you can use a GPIB command. NOTE Port extension and loss values that have been calculated are not cleared.
Optimizing Measurements Reducing Measurement Error in High Temperature Environments Reducing Measurement Error in High Temperature Environments NOTE The high temperature measurement mode must be turned off when the analyzer is used at an ambient temperature below 28°C. Otherwise, the measurement accuracy may be degraded. Procedure Step 1. Press . Step 2. Press Service Menu. Step 3. Press High Temperature to turn ON/OFF the high temperature measurement mode. Chapter 12 411 12.
Optimizing Measurements Improving Measurement Throughput Improving Measurement Throughput This section explains the following three methods to improve the measurement throughput.
Optimizing Measurements Improving Measurement Throughput Swept mode As shown in Figure 12-6, in the stepped mode, the frequency is changed stepwise and sampling is performed at a fixed frequency for each measurement point. On the other hand, in the swept mode, sampling is performed with the frequency always swept for each measurement point.
Optimizing Measurements Improving Measurement Throughput problem before performing the actual measurement. Notes for measuring DUTs with long electrical delay time When sweeping the frequency of a signal applied to the DUT ( F ), there is a frequency difference between the input side and output side of the DUT due to the delay time that occurs in the DUT ( ΔT ).
Optimizing Measurements Improving Measurement Throughput Fast mode The fast mode is a sweep mode in which the sweep time is shortened by decreasing the waiting time before sampling (stepped mode) or speeding up the sweep (swept mode) to the limit of the analog performance.
Optimizing Measurements Improving Measurement Throughput Turning off the updating of information displayed on the LCD screen Turning off the updating of information displayed on the LCD screen eliminates the processing time required to update displays within the analyzer, improving measurement throughput. If it is not necessary to check displayed information during measurements, turning off real-time updating is an effective means of improving throughput.
Optimizing Measurements Performing a Segment-by-Segment Sweep (segment sweep) Performing a Segment-by-Segment Sweep (segment sweep) This section describes the concept of the segment sweep and how to perform it. Concept of Segment Sweep • By skipping the frequency range, which does not need to be measured, you can sweep and measure only the portions you need. • You can define the optimum measurement conditions for each of the segments you designate.
Optimizing Measurements Performing a Segment-by-Segment Sweep (segment sweep) Conditions for setting up a segment sweep The following conditions apply when setting up a segment sweep. • The frequency range of a segment must not overlap with that of another segment. (The start frequency of a segment must be higher than the stop frequency of the immediately preceding segment.
Optimizing Measurements Performing a Segment-by-Segment Sweep (segment sweep) Sweep delay time and sweep time in a segment sweep The definitions for sweep delay time and sweep time, which you can specify in the segment sweep, are shown in Figure 12-9. Figure 12-9 Sweep delay time and sweep time in segment sweep 12.
Optimizing Measurements Performing a Segment-by-Segment Sweep (segment sweep) Figure 12-11 Comparison of methods used to displaying segments 420 Chapter 12
Optimizing Measurements Performing a Segment-by-Segment Sweep (segment sweep) Procedure Creating a segment table or Step 2. Press . to select the channel for which you want to create the Step 3. Press Edit Segment Table. The segment table appears in the lower part of the screen. Step 4. To change the frequency range setting mode or to set the IF bandwidth, power level, sweep delay time, sweep mode, and sweep time for each segment, use the following softkeys.
Optimizing Measurements Performing a Segment-by-Segment Sweep (segment sweep) Step 5. Enter each item in the following table for each added segment (line) to create the segment table.
Optimizing Measurements Performing a Segment-by-Segment Sweep (segment sweep) Figure 12-12 Example of creating segment table 12. Optimizing Measurements Useful functions when using a mouse By right-clicking on the selected cell, you can use the following shortcut menu.
Optimizing Measurements Performing a Segment-by-Segment Sweep (segment sweep) Executing segment sweep To execute a segment sweep by using the segment table you have created, you must specify the sweep type for that sweep operation by following the steps below. Step 1. Press (or segment sweep operation. Step 2. Press ) to select the channel on which you will execute the . Step 3. Press Sweep Type. Step 4. Press Segment.
Optimizing Measurements Performing a Segment-by-Segment Sweep (segment sweep) Saving a newly created segment table in CSV format As discussed in “Creating a segment table” on page 421, you can export the newly created segment table as a CSV (comma-separated value) formatted file (so it can be used easily in software that requires a different format). Step 1. Press . Step 3. Press Export to CSV File to open the Save As dialog box.
Optimizing Measurements Performing a Segment-by-Segment Sweep (segment sweep) 426 Chapter 12
13.Setting and Using the Control and Management Functions 13 Setting and Using the Control and Management Functions This chapter describes how to set and use the control and management functions not directly linked with measurement or analysis.
Setting and Using the Control and Management Functions Setting the GPIB Setting the GPIB This section describes how to set the interface necessary to use the GPIB (General Purpose Interface Bus) of the E5070B/E5071B. For information on the concept and concrete implementation of the auto measurement using GPIB, refer to “Programmers Guide.
Setting and Using the Control and Management Functions Setting the GPIB Figure 13-1 USB/GPIB Interface Detected dialog box Step 2. Confirm that VISA Interface Name is set to GPIB0 (1 in Figure 13-1) and SICL Interface Name is set to hpib7 (2 in Figure 13-1) and then click the Accept button (3 in Figure 13-1). If the setting is correct, the procedure is complete. If the setting is different, click the Edit button (4 in Figure 13-1).
Setting and Using the Control and Management Functions Setting the GPIB Step 5. The IO Config dialog box (Figure 13-3) appears. Select (highlight) GPIB0 hpib7 (1 in Figure 13-3) and then click the Edit button (2 in Figure 13-3). NOTE In the IO Config dialog box, do not click buttons other than specified here or do not change other settings because doing so may cause serious damage to the functions of the E5070B/E5071B. Figure 13-3 IO Config dialog box Step 6.
Setting and Using the Control and Management Functions Setting the GPIB Setting system controller (USB/GPIB interface) when c drive volume label in hard disk is more than CP810 When controlling an external device from the E5070B/E5071B, connect the USB port of the E5070B/E5071B and the GPIB port of the external device through the USB/GPIB interface.
Setting and Using the Control and Management Functions Setting the GPIB If you need to check/change the setting of the USB/GPIB interface after connecting the USB/GPIB interface, follow these steps: Step 1. Press . Step 2. Press Misc Setup. Step 3. Press GPIB Setup. Step 4. Press System Controller Configuration. Step 5. Agilent Connection Expert (Figure 13-5) appears. After selecting the USB/GPIB(GPIB0) (1 in Figure 13-5), click Change Properties... button (2 in Figure 13-5). Step 6.
Setting and Using the Control and Management Functions Setting the Internal Clock Setting the Internal Clock The E5070B/E5071B has the built-in clock that keeps track of the date and time. This clock is used for the following functions. • To display the current date and time in the instrument status bar at the lower part of the screen • To write date and time information when saving internal data or a VBA program Setting the Date and Time Step 1. Press . Step 2. Press Misc Setup. Step 3.
Setting and Using the Control and Management Functions Setting the Internal Clock Figure 13-7 Date/Time Properties Dialog Box (“Time Zone” Tab) Step 7. In the drop-down list box select a time zone. Step 8. To make the summertime setting automatically, check Automatically adjust clock for daylight saving changes to assign the check mark (√) to it. Step 9. Press the OK button.
Setting and Using the Control and Management Functions Setting the Mouse Setting the Mouse The user can change the setup for the mouse connected to the E5070B/E5071B and the movement of the pointer. Setup Step NOTE Be sure to use a mouse and a keyboard for mouse setup operations. Step 1. Press . Step 2. Press Misc Setup. Step 3. Press Control Panel to open the Control Panel window. Figure 13-8 Control Panel Window 13.Setting and Using the Control and Management Functions Step 4.
Setting and Using the Control and Management Functions Setting the Mouse Step 5. The Mouse Properties dialog box (Figure 13-9) is displayed. Define the setup for a right-handed/left-handed person in the Buttons configuration area. Define also the setup for double-click speed in the Double-click speed area. Figure 13-9 Mouse Properties Dialog Box (Buttons tab) Step 6. Click the Pointers tab (Figure 13-10).
Setting and Using the Control and Management Functions Setting the Mouse Step 7. Enter a registration name into the Scheme box and specify the shapes of pointers for the registration name in the box below. To create a registration name, click the Save As... button. Enter the registration name into the Save Scheme dialog box that appears, and click the OK button. Step 8. Click the Motion tab (Figure 13-11). Figure 13-11 Mouse Properties Dialog Box (Motion tab) Step 10. Click the OK button. Step 11.
Setting and Using the Control and Management Functions Configuring the Network Configuring the Network NOTE When you use the E5070B/E5071B by connecting it to your LAN, consult your network administrator and make the setting of the LAN correctly. This section describes how to set the following basic items necessary to connect the E5070B/E5071B to the LAN (Local Area Network).
Setting and Using the Control and Management Functions Configuring the Network Step 6. When switching from disable to enable: Double-click the Local Area Connection icon (1 in Figure 13-12) in the Network and Dial-up connections window to enable the network connection function. When switching from enable to disable: Double-click the Local Area Connection icon (1 in Figure 13-12) in the Network and Dial-up Connections window. The Local Area Connection Status dialog box (Figure 13-13) appears.
Setting and Using the Control and Management Functions Configuring the Network Step 6. The Local Area Connection Properties dialog box (Figure 13-14) appears. Select (highlight) Internet Protocol (TCP/IP) (1 in Figure 13-14) and then click the Properties button (2 in Figure 13-14). Figure 13-14 Local Area Connection Properties dialog box Step 7. The Internet Protocol (TCP/IP) Properties dialog box (Figure 13-15) appears.
Setting and Using the Control and Management Functions Configuring the Network Step 8. In the Internet Protocol (TCP/IP) Properties dialog box, click the OK button (6 in Figure 13-15). Step 9. In the Local Area Connection Properties dialog box, click the OK button (3 in Figure 13-14). Step 10. In the Local Area Connection Status dialog box, click the Close button (3 in Figure 13-13). Step 11. Click the × button (2 in Figure 13-12) in the upper right of the Network and Dial-up Connections window.
Setting and Using the Control and Management Functions Configuring the Network Step 6. The Identification Changes dialog box (Figure 13-17) appears. Enter the computer name in the Computer Name box (1 in Figure 13-17). Figure 13-17 Identification Changes dialog box Step 7. The Network Identification dialog box (Figure 13-18) appears. Click the OK button. Figure 13-18 Network Identification dialog box Step 8. In the Identification Changes dialog box, click the OK button (2 in Figure 13-17). Step 9.
Setting and Using the Control and Management Functions Remote Control Using HTTP Remote Control Using HTTP You can access the web page installed in the E5070B/E5071B by using the hypertext transfer protocol (http) and the E5070B/E5071B's IP address from the external PC's web browser. Through the built-in web page, you can control the E5070B/E5071B remotely and display the measurement screen on external PCs.
Setting and Using the Control and Management Functions Accessing Hard Disk of E5070B/E5071B from External PC Accessing Hard Disk of E5070B/E5071B from External PC If you connect the E5070B/E5071B to LAN, you can access the hard disk (D drive) in the E5070B/E5071B as a network drive from an external PC connected to the same LAN.
Setting and Using the Control and Management Functions Accessing Hard Disk of E5070B/E5071B from External PC Step 4. The USERS(D:) Properties dialog box (Figure 13-21) appears. Select the Sharing tab. Figure 13-21 USERS(D:) Properties dialog box (General tab) 13.Setting and Using the Control and Management Functions Step 5. Click the New Share button (1 in Figure 13-22).
Setting and Using the Control and Management Functions Accessing Hard Disk of E5070B/E5071B from External PC Step 6. The New Share dialog box (Figure 13-23) appears. Enter the share name (name used when accessed from the external PC) in the Share Name box (1 in Figure 13-23) and click the OK button (2 in Figure 13-23). Figure 13-23 New Share dialog box Step 7. In the USERS(D:) Properties dialog box, click the OK button (2 in Figure 13-22).
Setting and Using the Control and Management Functions Disabling USB Mass Storage Device Disabling USB Mass Storage Device You can disable any USB-compatible external mass storage devices in order to ensure confidentiality or for other reasons. Steps for Setting Modification The following procedure shows how to disable a USB Mass Storage Device. Step 1. Press . Step 2. Press Explorer.... Step 3. Double-click DisableUsbStorage.exe from D:\Agilent\Service. Step 4.
Setting and Using the Control and Management Functions Locking the Front Keys, Keyboard, and/or Mouse (Touch Screen) Locking the Front Keys, Keyboard, and/or Mouse (Touch Screen) You can lock (disable) the front keys, keyboard, and/or mouse (touch screen). This feature prevents erroneous operation caused by inadvertently touching any of these devices. Locking the Front Keys, Keyboard, and/or Mouse Step 1. Press . Step 2. Press Misc Setup. Step 3. Press Key Lock. Step 4.
Setting and Using the Control and Management Functions Setting the Beeper (Built-in Speaker) Setting the Beeper (Built-in Speaker) The E5070B/E5071B has a built-in speaker that sounds a beep tone. The beeper allows you to make two types of settings shown in Table 13-2. Table 13-2 Type Function Operation complete beeper Warning beeper Sounds a beep tone to inform the user that operations have completed.
Setting and Using the Control and Management Functions Turning off the LCD Screen Backlight Turning off the LCD Screen Backlight You can switch off the backlight (illumination) of the LCD screen of the E5070B/E5071B. This extends the life of the backlight when using it continuously over a long period. Turning off the LCD Screen Backlight Step 1. Press . Step 2. Press Backlight to switch the backlight on/off. Switching off the backlight causes indications on the LCD screen to be almost invisible.
Setting and Using the Control and Management Functions Checking the product information Checking the product information Checking the serial number The serial number of the E5070B/E5071B can be checked using the following procedure. Procedure to check the serial number Step 1. Press . Step 2. Press Service Functions. Step 3. Press Enable Options. The serial number is displayed in the softkey menu bar.
Setting and Using the Control and Management Functions Checking the product information Figure 13-24 Firmware Revision dialog box Step 3. Press OK to close the dialog box.
Setting and Using the Control and Management Functions Setting the preset function Setting the preset function Showing/hiding the confirmation buttons when presetting The preset function can be executed without displaying the OK and Cancel softkey buttons when pressing the preset button of the E5070B/E5071B. Operational procedure Step 1. Press . Step 2. Press Misc Setup. Step 3. Press Preset Setup. Step 4. Confirm to toggle on (show)/off (hide) the confirmation buttons.
Setting and Using the Control and Management Functions Setting the preset function Saving a user-preset instrument state To execute the user-preset function, you must have a preset setting file that has been saved. Follow these steps to save a preset instrument state of the E5070B/E5071B. Operational procedure Step 1. Sets up a preset instrument state you want to save. Step 2. Press . Step 3. Press Save State. Step 4. Pressing User Pres saves the instrument state.
Setting and Using the Control and Management Functions System Recovery System Recovery By executing system recovery, you can return the system of the E5070B/E5071B (the Windows operating system and the firmware) to the factory state (at the time of purchase*1). The system recovery procedures vary depending on the mother board the E5070B/E5071B is equipped with.
Setting and Using the Control and Management Functions System Recovery Procedure to execute the factory function (1) This procedure is valid for the E5070B : JP1KKxxxxx, MY42300632 and below, E5071B : JP1KKxxxxx, MY42301396 and below. If your unit doesn’t follow the procedure written below, please refer to “Procedure to execute the factory recovery function (2)” on page 465 for the detail. This section describes how to return the contents of the C drive to the factory state.
Setting and Using the Control and Management Functions System Recovery Step 6. The following screen appears. Check that Floppy is selected (highlighted) (if Floppy is not selected, select it with of the keyboard), and press of the keyboard. If you want to cancel the Agilent Technologies System Utilities Recovery & Backup Options for the E5070/71B Choose One of the following: _________________________________________________________________ 1. Recover Factory Backup Image 2. Create User Backup Image 3.
Setting and Using the Control and Management Functions System Recovery Step 8. The message as shown below appears. Press factory recovery, press here. of the keyboard. If you want to cancel the You chose to Restore your system by installing the original factory installed OS and system software. WARNING: Press C to Continue only if you are sure that you want to proceed. The C: Drive will be completely overwritten with no chance of recovering any data.
Setting and Using the Control and Management Functions System Recovery Procedure to create the user backup image (1) This procedure is valid for the E5070B : JP1KKxxxxx, MY42300632 and below, E5071B : JP1KKxxxxx, MY42301396 and below. If your unit doesn’t follow the procedure written below, please refer to “Procedure to create the user backup image (2)” on page 468 for the detail. This section describes how to create the user backup image.
Setting and Using the Control and Management Functions System Recovery Step 6. The following screen appears. Check that Floppy is selected (highlighted) (if Floppy is not selected, select it with of the keyboard), and press Step 7. The message as shown below appears. Press create user backup image, press here. of the keyboard. of the keyboard.
Setting and Using the Control and Management Functions System Recovery Step 8. The message as shown below appears. Press create user backup image, press here. of the keyboard. If you want to cancel the You chose to create a backup image file of your system. The system will perform a quick integrity check of the file structure on the C: Drive. It will then copy the C: partition to an image file and store it on the System Recovery partition. Press C to Continue or E to Exit: _ Step 9.
Setting and Using the Control and Management Functions System Recovery Procedure to execute the user recovery function (1) This procedure is valid for the E5070B : JP1KKxxxxx, MY42300632 and below, E5071B : JP1KKxxxxx, MY42301396 and below. If your unit doesn’t follow the procedure written below, please refer to “Procedure to execute the user recovery function (2)” on page 471 for the detail. Returns the contents of the C drive to a user-specified state.
Setting and Using the Control and Management Functions System Recovery Step 6. The following screen appears. Check that Floppy is selected (highlighted) (if Floppy is not selected, select it with of the keyboard), and press of the keyboard. If you want to cancel the Agilent Technologies System Utilities Recovery & Backup Options for the E5070/71B Choose One of the following: _________________________________________________________________ 1. Recover Factory Backup Image 2. Create User Backup Image 3.
Setting and Using the Control and Management Functions System Recovery Step 8. The message as shown below appears. Press user recovery, press here. of the keyboard. If you want to cancel the You chose to recover your own system backup image file. WARNING: Press C to Continue only if you are sure that you want to proceed. The C: partition will be completely overwritten with no chance of recovering any data.
Setting and Using the Control and Management Functions System Recovery Procedure to execute the factory recovery function (2) This procedure is valid for the E5070B : MY42300633, and above, E5071B : Y42301397 and above. If your unit doesn’t follow the procedure written below, please refer to “Procedure to execute the factory function (1)” on page 456 for the detail. This section describes how to return the contents of the C drive to the factory state. NOTE You need the keyboard for this operation. Step 1.
Setting and Using the Control and Management Functions System Recovery Step 6. The following screen appears. Select “+Removable Devices” with keyboard, and press of the of the keyboard. Step 7. The message as shown below appears. Press user recovery, press here. of the keyboard. If you want to cancel the Agilent Technologies System Utilities Recovery & Backup Options for the E5070/71B Choose One of the following: _________________________________________________________________ 1.
Setting and Using the Control and Management Functions System Recovery Step 8. The message as shown below appears. Press factory recovery, press here. of the keyboard. If you want to cancel the You chose to Restore your system by installing the original factory installed OS and system software. WARNING: Press C to Continue only if you are sure that you want to proceed. The C: Drive will be completely overwritten with no chance of recovering any data.
Setting and Using the Control and Management Functions System Recovery Procedure to create the user backup image (2) This procedure is valid for the E5070B : MY42300633, and above, E5071B : Y42301397 and above. If your unit doesn’t follow the procedure written below, please refer to “Procedure to create the user backup image (1)” on page 459 for the detail. This section describes how to create the user backup image.
Setting and Using the Control and Management Functions System Recovery Step 6. The following screen appears. Select “+Removable Devices” with keyboard, and press of the of the keyboard. of the keyboard. If you want to cancel the Agilent Technologies System Utilities Recovery & Backup Options for the E5070/71B Choose One of the following: _________________________________________________________________ 1. Recover Factory Backup Image 2. Create User Backup Image 3. Recover User Backup Image 4.
Setting and Using the Control and Management Functions System Recovery Step 8. The message as shown below appears. Press create user backup image, press here. of the keyboard. If you want to cancel the You chose to create a backup image file of your system. The system will perform a quick integrity check of the file structure on the C: Drive. It will then copy the C: partition to an image file and store it on the System Recovery partition. Press C to Continue or E to Exit: _ Step 9.
Setting and Using the Control and Management Functions System Recovery Procedure to execute the user recovery function (2) This procedure is valid for the E5070B : MY42300633, and above, E5071B : Y42301397 and above. If your unit doesn’t follow the procedure written below, please refer to “Procedure to execute the user recovery function (1)” on page 462 for the detail. Returns the contents of the C drive to a user-specified state. To use this function, you must create the user backup image in advance.
Setting and Using the Control and Management Functions System Recovery Step 6. The following screen appears. Select “+Removable Devices” with keyboard, and press of the of the keyboard. Step 7. The message as shown below appears. Press user recovery, press here. of the keyboard. If you want to cancel the Agilent Technologies System Utilities Recovery & Backup Options for the E5070/71B Choose One of the following: _________________________________________________________________ 1.
Setting and Using the Control and Management Functions System Recovery Step 8. The message as shown below appears. Press user recovery, press here. of the keyboard. If you want to cancel the You chose to recover your own system backup image file. WARNING: Press C to Continue only if you are sure that you want to proceed. The C: partition will be completely overwritten with no chance of recovering any data.
Setting and Using the Control and Management Functions Calibration of the Touch Screen Calibration of the Touch Screen When you have executed system recovery on the E5070B/E5071B equipped with an Option 016 touch screen, you have to calibrate the touch screen. Follow the procedure described below to calibrate the touch screen. Step 1. Press . Step 2. Press Service Menu. Step 3. Press Test Menu. Step 4. Press Adjust Touch Screen. The touch screen calibration screen (Figure 13-25) appears.
Setting and Using the Control and Management Functions Initial Source Port Control function Initial Source Port Control function Firmware Rev. A.03.54 and later provides an additional feature, "Initial Source Port Control." It protects the output amplifier inside the instrument against any potential damage due to transient voltage that may be externally applied. After the firmware is installed, this feature is activated at power-on.
Setting and Using the Control and Management Functions Initial Source Port Control function To disable the Initial Source Port Control feature (applicable to Rev.A.03.54 and later): When using the front panel menu - Service-Init Src Ctrl [OFF] When using the SCPI command - SYSTem:ISPControl[:STATe] OFF When using the VBA command - SCPI.SYSTem.ISPControl.
14. Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set.
Controlling E5091A Connecting E5070B/E5071B and E5091A Connecting E5070B/E5071B and E5091A Required devices The devices required to connect the E5070B/E5071B to the E5091A are listed below.
Controlling E5091A Connecting E5070B/E5071B and E5091A As shown in Figure 14-2, connect the N-type cable between the front panel of the E5070B/E5071B and that of the E5091A. Make the connection so that the numbers of the test ports of the E5070B/E5071B and those of the interconnection ports of the E5091A match. Figure 14-2 Connection between E5070B/E5071B and E5091A (front view) Powering on After connecting the E5070B/E5071B and the E5091A, follow these steps to power both devices on. Step 2.
Controlling E5091A Setting the E5091A Setting the E5091A This section describes the settings of the E5091A. Table 14-1 shows the flow used for item setting.
Controlling E5091A Setting the E5091A Selecting the E5091A Model Select the E5091A model you want to set. For the E5091A-016, you can select the 13-port device or 16-port device function. Softkey Model E5091_9 Select the E5091A option 009. E5091_13 Select the E5091A option 016 for the 13-port device function. E5091_16 Select the E5091A option 016 for the 16-port device function. Step 1. Press - E5091A Setup to display the E5091A setup menu. Step 2.
Controlling E5091A Setting the E5091A Table 14-3 When the E5091A-009 is Connected Softkey Function Port2 Selects a test port of the E5091A to which you want to connect port 2 of the E5091A. You can select the port from T1*1 or T2. Port3 Selects a test port of the E5091A to which you want to connect port 3 of the E5091A. You can select the port from R1+, R2+, or R3+. Port4 Selects a test port of the E5091A to which you want to connect port 4 of the E5091A.
Controlling E5091A Setting the E5091A NOTE The same test ports cannot be connected to each port. In such a case, the other test ports’settings will be automatically changed. Displaying the E5091A properties By displaying the E5091A properties shown in Figure 14-4, you can obtain the assignment information of the test ports for each channel. This is useful when you need to check the test port assignment, for example, when you perform calibration.
Controlling E5091A Setting the E5091A Setting control line The E5070B/E5071B can control the output from the control line (Figure 14-5) of the E5091A and control the DUT (for example, switching the frequency band of the front end module). The procedure is shown below. For the specifications of the DUT control line, refer to E5091A Users & Service Guide. Figure 14-5 Control line Operational procedure Step 1. Press Step 2. Press control line. - E5091A Setup to display the E5091A setup menu.
Controlling E5091A Setting the E5091A Enabling control of E5091A If you enable control of the E5091A, it becomes possible to switch the measurement path and the output function of the control line of the E5091A during measurement. NOTE The E5070B/E5071B needs the additional time (about 3 ms) to control the E5091A when you have enabled control of the E5091A. Operational procedure Step 1. Press - Multiport Test Set Setup to display the E5091A setup menu. Step 2.
Controlling E5091A Calibration Calibration Follow these steps to perform calibration with the E5091A connected: Step 1. Press active channel. (or ) to set the channel that you want to calibrate to the Step 2. Follow “Displaying the E5091A properties” on page 483 to display the E5091A properties. Step 3. According to the procedures in Chapter 4, “Calibration,” on page 93, perform the calibration.
Controlling E5091A Performing Measurement Performing Measurement Trigger state and switching the setting of the E5091A The following table shows how the setting in the E5091A is switched from when the trigger state is the stop state. For more information on the trigger state, refer to E5070B/E5071BProgrammers Guide. Trigger state Switching the setting of E5091A Stop The setting is not switched.
Controlling E5091A Connecting Two E5091As Connecting Two E5091As Give consideration to the following items when you make measurements by using two E5091As. • Set their IDs to different values. The instruments will not work correctly if their IDs are the same. • Connect calibration standards and the DUT only after confirming the connection and port assignment of the E5091As.
15 Measurement Examples 489 15. Measurement Examples This chapter introduces examples of actual device measurements made by using the Agilent E5070B/E5071B.
Measurement Examples Measuring the SAW Bandpass Filter Using the Segment Sweep Measuring the SAW Bandpass Filter Using the Segment Sweep This section illustrates how to use the segment sweep function to evaluate a SAW bandpass filter with a center frequency of 947.5 MHz. Evaluation Steps Here, the DUT is evaluated by following the steps described in Table 15-1. Table 15-1 Evaluating the DUT Using the Segment Sweep Step Description “1.
Measurement Examples Measuring the SAW Bandpass Filter Using the Segment Sweep 1. Determine the Segment Sweep Conditions Figure 15-1 shows the results of evaluating the transmission characteristics of the SAW bandpass filter in the range of 440 MHz to 3 GHz by using the linear sweep. Figure 15-1 Transmission characteristics of SAW bandpass filter (440 MHz to 3 GHz, linear sweep) The measurement conditions are determined for each frequency range.
Measurement Examples Measuring the SAW Bandpass Filter Using the Segment Sweep 2. Create a Segment Sweep Table Follow the steps below to make entries in the segment sweep table. Step 1. Display the segment table. Setup Description Presetting Displaying the segment table Key Operation - OK - Edit Segment Table Step 2. Display the IF bandwidth setting column in the segment table.
Measurement Examples Measuring the SAW Bandpass Filter Using the Segment Sweep Setup Description Key Operation Number of points: 55 IF bandwidth: 70 kHz Segment 4 Start frequency: 1.07 GHz Stop frequency: 2 GHz Number of points: 93 IF bandwidth: 70 kHz Segment 5 Start frequency: 2.6 GHz Stop frequency: 3 GHz Number of points: 41 IF bandwidth: 70 kHz Figure 15-2 Completed segment table 493 15.
Measurement Examples Measuring the SAW Bandpass Filter Using the Segment Sweep 3. Select the Segment Sweep as the Sweep Type The segment sweep is selected as the sweep type. Setup Description Key Operation Sweep type: Segment sweep - Sweep Type - Segment 4. Execute the Calibration In this step, a 2-port ECal is executed on the two ports to be used. Step 1. Connect the ECal module across test ports 1 and 2. Figure 15-3 Connecting the ECal module Step 2. Execute the 2-port ECal.
Measurement Examples Measuring the SAW Bandpass Filter Using the Segment Sweep 5. Connect the DUT The DUT is connected across test ports 1 and 2. Figure 15-4 Connecting the DUT 6. Execute the Measurement A trigger is applied to perform the measurement. Setup Description Trigger mode: Single Key Operation - Single (Or Continuous) 7. Define the Setup for Display The choice is made between frequency base and order base as the segment display mode.
Measurement Examples Measuring the SAW Bandpass Filter Using the Segment Sweep Figure 15-5 Segment display: frequency base Figure 15-6 Segment display: order base 496 Chapter 15
Measurement Examples Evaluating a Duplexer Evaluating a Duplexer This section illustrates how to evaluate a duplexer (Tx center frequency: 1.88 GHz, Rx center frequency: 1.96 GHz). Evaluation Steps Here, the DUT is evaluated by following the steps described in Table 15-3. Table 15-3 Evaluating the DUT Using the Segment Sweep Step Description “1. Determine the Segment Sweep Conditions” on page 497 Segment sweep conditions are determined according to the characteristics of the DUT. “2.
Measurement Examples Evaluating a Duplexer 2. Create a Segment Sweep Table Entries are made in the segment sweep table following the steps described below. Step 1. Display the segment table. Setup Description Presetting Key Operation - OK Displaying the segment table - Edit Segment Table Step 2. Enter the setup data in the segment table (Figure 15-7). NOTE In this step, the IF Bandwidth, power level, delay time, and sweep time are not entered segment by segment.
Measurement Examples Evaluating a Duplexer 3. Select the Segment Sweep as the Sweep Type The segment sweep is selected as the sweep type. Setup Description Key Operation Sweep type: Segment sweep (Moves the focus to the softkey menu) - Return - Sweep Type - Segment (or - Sweep Type - Segment) 4. Execute the Calibration In this step, a 2-port ECal module and 3-/4-port module installed in the E5070B/E5071B are used to execute calibration on the three ports used in the measurement. Step 1.
Measurement Examples Evaluating a Duplexer Figure 15-9 EcalAssistant (port/channel selection) dialog box Step 5. Following the instructions in the dialog box, select the type of ECal, test ports, and the channel. Setup Description Key Operation Select Ports ECal type: Full 3-port calibration 3 Port Test ports to be used for ECal: 1, 2, and 3 1, 2, 3 Select Channel Channel on which ECal is to be executed: Channel 1 Channel: 1 Step 6. Press the Next button.
Measurement Examples Evaluating a Duplexer Figure 15-11 Connecting the ECal module Step 8. The EcalAssistant (complete) dialog box (Figure 15-12) appears. Press the Done button to terminate the calibration. Figure 15-12 ECalAssistant (complete) dialog box 5. Connect the DUT The DUT is connected to test ports 1, 2, and 3. Figure 15-13 Connecting the DUT 501 15.
Measurement Examples Evaluating a Duplexer 6. Define the Setup for Display The setup for display is defined. Setup Description Key Operation Number of traces to be displayed: 5 Trace placement: Trisected - Number of Traces - 5 Allocate Traces - Measurement Parameter Trace 1: S13 - S13 Trace 2: S21 - - S21 Trace 3: S23 - - S23 Trace 4: S33 - - S33 Trace 5: S11 - - S11 7. Execute the Measurement A trigger is applied to execute the measurement.
Measurement Examples Evaluating a Duplexer Figure 15-14 Measurement result (segment display: order base) 9. Analyze the Parameters The parameters for the duplexer are determined. Step 1. Determine the insertion loss and 3-dB bandwidth for Tx. Setup Description Key Operation Marker coupling: OFF - Couple (Turn it OFF) Activating Trace 1 Marker 1: ON Search/Tracking: ON - Tracking (Turn it ON) Moving the marker to the trace maximum Max Bandwidth search: ON Bandwidth (Turn it ON.
Measurement Examples Evaluating a Duplexer Setup Description Key Operation Moving Marker 1 to the trace maximum Max Bandwidth search: ON Bandwidth (Turn it ON) In the example shown in Figure 15-15, the insertion loss (loss) is 1.627 dB and the 3-dB bandwidth (BW) is 71.04 MHz. Step 3. Determine the isolation between Tx and Rx. Setup Description Key Operation Activating Trace 3 Marker 1: ON Search/Tracking: ON - Tracking (Turn it ON) Moving Marker 1 to the peak near 1.
Measurement Examples Evaluating a Duplexer Setup Description Key Operation Moving Marker 1 to the peak in the pass band Peak - Search Left|Search Right (press as many times as necessary)*1 *1. If you cannot move the marker to the desired peak, use Peak Excursion to change the peak deviation and then execute the search again. (Example: Peak Excursion ). In the example of Figure 15-15, return loss (response value of Marker 1) is 13.80 dB. Step 6. Turn on the marker table display.
Measurement Examples Evaluating a Duplexer Step 2. Enter the setup data in the limit table for trace 1 (Figure 15-16). Figure 15-16 Completed limit table for trace 1 Step 3. Display the limit table for Trace 2 (S21). Setup Description Key Operation Activating Trace 2 Step 4. Enter the setup data in the limit table for trace 2 (Figure 15-17). Figure 15-17 Completed limit table for trace 2 11. Execute the Limit Test The limit test is executed. Step 1. Turn on the limit line and limit test for Trace 1.
Measurement Examples Evaluating a Duplexer Step 2. Turn on the limit line and limit test for Trace 2. Setup Description Key Operation Activating Trace 2 Limit Line: ON Limit Line (Turn it ON) Limit Test: ON Limit Test (Turn it ON) Step 3. Apply a trigger to execute the measurement. Setup Description Trigger Mode: Single Figure 15-18 Key Operation - Single (or Continuous) Limit test results Step 4. Maximize the screen display of Trace 1 to examine its details.
Measurement Examples Evaluating a Duplexer Figure 15-19 Enlarged display of trace 1 Step 5. Maximize the screen display of Trace 2 to examine its details. Figure 15-20 Setup Description Key Operation Activating Trace 2 (The display of Trace 2 is maximized.
Measurement Examples Measuring the Deviation from a Linear Phase Measuring the Deviation from a Linear Phase This section illustrates how to determine the deviation from a linear phase in the pass band of a 1.09-GHz bandpass filter. Evaluation Steps Here, the DUT is evaluated by following the steps described in Table 15-5. Table 15-5 Evaluating the Deviation from a Linear Phase Step Description “1. Connect the DUT” on page 509 The DUT is connected. “2.
Measurement Examples Measuring the Deviation from a Linear Phase 2. Define the Measurement Conditions The measurement conditions are defined by following the steps described below. Setup Description Key Operation Presetting - OK Center frequency: 1.09 GHz Frequency span: 20 MHz Measurement parameter: S21 - S21 Data format: Expand Phase - Expand Phase Executing the Auto Scale - Auto Scale 3. Execute the Calibration The THRU response calibration is executed.
Measurement Examples Measuring the Deviation from a Linear Phase Figure 15-22 Phase characteristics of the DUT 5. Specify the Electrical Delay The electrical delay is entered to flatten the phase trace. Setup Description Entering the electrical delay Figure 15-23 Key Operation - Electrical Delay (Flattening a trace) Entering the electrical delay 511 15.
Measurement Examples Measuring the Deviation from a Linear Phase 6. Measure the Deviation from a Linear Phase The statistics data is used to read the deviation from a linear phase (peak-to-peak) (Figure 15-24). Setup Description Key Operation Executing the auto scale Displaying the statistics data Figure 15-24 - Auto Scale - Statistics (Turn it ON.
Measurement Examples Measuring an Unbalanced and Balanced Bandpass Filter Measuring an Unbalanced and Balanced Bandpass Filter This section introduces an example of actually evaluating the unbalanced and balanced SAW bandpass filter with a center frequency of 942.5 MHz. Figure 15-25 shows the measurement circuit in the condition for evaluating a DUT. Figure 15-25 Measurement circuit Evaluation Steps Here, the DUT is evaluated by following the steps described in Table 15-6.
Measurement Examples Measuring an Unbalanced and Balanced Bandpass Filter 1. Connecting the DUT Connect the DUT to the E5070B/E5071B by using the instrument’s three test ports (Figure 15-26).
Measurement Examples Measuring an Unbalanced and Balanced Bandpass Filter 2. Setting the Measurement Conditions Follow the procedure below to set the measurement conditions. The measurement parameters for balanced measurements should be set after unbalanced-balanced conversion. Here, set the measurement parameters for observing the characteristics achieved during unbalanced measurements.
Measurement Examples Measuring an Unbalanced and Balanced Bandpass Filter 3. Performing Calibration Perform a full three-port calibration for the three ports to be used. Step 1. Set the type and conditions of calibration. Setting Description Calibration kit to use: 85033D Key Operation - Cal Kit - 85033D Type of calibration: Full three-port calibration Calibrate - 3-Port Cal Test ports to calibrate: 1, 2, 3 Select Ports - 1-2-3 (check only) Step 2. Perform a reflection calibration.
Measurement Examples Measuring an Unbalanced and Balanced Bandpass Filter 4. Setting a Balance Conversion Topology Follow the procedure below to set the balanced conversion topology. Table 15-7 Setting Description Key Operation Set port 1 on the DUT to unbalanced and port 2 on the DUT to balanced. - Fixture Simulator - Topology Device - SE-Bal (check only) Set the connecting destination of port 1 on the DUT (unbalanced) to test port 1 of the analyzer.
Measurement Examples Measuring an Unbalanced and Balanced Bandpass Filter Step 4. Set the measurement parameter (mixed mode S-parameter) and data format for trace 3. Setting Description Key Operation Unbalanced-balanced conversion of trace 3: ON BalUn (turns it ON) Measurement parameter: Sss11 Data format: Smith chart (marker display: R+jX) - - Fixture Simulator - - Sss11 - Smith - R + jX Step 5. Set the measurement parameter (mixed mode S-parameter) and data format for trace 4.
Measurement Examples Measuring an Unbalanced and Balanced Bandpass Filter 6. Extending the Calibration Plane (removing the cause of error) In this section you will use the port extension function to remove an electrical delay caused by cables or fixtures located between the calibration reference plane and the DUT to be evaluated.
Measurement Examples Measuring an Unbalanced and Balanced Bandpass Filter Step 1. Set the port reference impedance of port 1 on the DUT (unbalanced) to 50 Ω. Setting Description Key Operation Reference impedance of test port 1: 50 Ω - Fixture Simulator Port Z conversion - Port 1 Z0 - Step 2. In order to set the impedance of the differential mode of port 2 on the DUT (balanced) to 200 Ω, set the impedances of two unbalanced ports before conversion each to 100 Ω.
Measurement Examples Measuring an Unbalanced and Balanced Bandpass Filter 8. Adding a Matching Circuit Here, add an inductance of 47 nH in parallel to port 2 on the DUT (balanced). It is also possible to add a matching circuit to the port before unbalanced-balanced conversion. For more information, see “Determining the Characteristics that Result from Adding a Matching Circuit to a Differential Port” on page 305.
Measurement Examples Measuring Parameters with Cable Measuring Parameters with Cable This section introduces an example of how to detect the location of a mismatch that occurs in a cable by using the time domain function. Overview of evaluation procedure In this example, a DUT is evaluated according to the steps shown in Table 15-8. Table 15-8 Evaluation Procedure for Deviation from Linear Phase Step Description “1. Setting the Measurement Conditions” on page 522 Set the measurement conditions. “2.
Measurement Examples Measuring Parameters with Cable 3. Connecting the DUT Connect the DUT as shown in Figure 15-32. Figure 15-32 Connecting the DUT 4. Auto Scale Execute the auto scale function. Setting Description Executing auto scale Figure 15-33 Key Operation - Auto Scale Response in frequency domain 523 15.
Measurement Examples Measuring Parameters with Cable 5. Setting the Time Domain Function Set the conversion function to display the response in time the domain. If you enable this setting, the response in time domain is displayed as shown in Figure 15-34. A peak indicating a small mismatch appears at the location of the connector. Setting Description Key Operation Data format: real - Real Setting the transformation type to low-pass impulse - Transform - Lowpass Imp.
Measurement Examples Evaluating Transmission Characteristics of a Front End Module Evaluating Transmission Characteristics of a Front End Module This example shows how to measure the transmission characteristics of a 6-port front end module, as shown in Figure 15-35, by using the E5070B/E5071B and the E5091A. Figure 15-35 Front end module Overview of evaluation procedure In this example, a DUT is evaluated according to the procedure shown in Table 15-9.
Measurement Examples Evaluating Transmission Characteristics of a Front End Module Table 15-9 Evaluation Procedure for 6-port Front End Module Procedure Description “10. Executing Measurement” on page 531 Execute the measurement and perform auto scale. 1. Determining Measurement Conditions In this example, perform measurement under the measurement conditions in Table 15-9. Table 15-10 Channel Sweep conditions Start frequency Stop frequency NOP 1 400 MHz 1.
Measurement Examples Evaluating Transmission Characteristics of a Front End Module Step 3. Select the test ports assigned to ports 1 to 4 for channel 1. Setting Description Key Operation Assign test port A to port 1. Port1 - A Assign test port T1 to port 2. Port2 - T1 Assign test port R1+ to port 3. Port3 - R1+ Assign test port R1- to port 4. Port4 - R1- Step 4. Assign test ports for channels 2, 3, and 4. Press the channel and then make the setting in the same way as Step 3.
Measurement Examples Evaluating Transmission Characteristics of a Front End Module includes the balanced port. Setting Description Key Operation Set channel 2 to the active channel. activated) (press it until channel 2 is Set DUT port 1 to unbalance and DUT port 2 to balance. - Fixture Simulator - Topology Device - SE-Bal Set the destination to which DUT port 1 (unbalance) is connected to test port 1 of the analyzer.
Measurement Examples Evaluating Transmission Characteristics of a Front End Module Step 2. Set channel 1 to the active channel. Setting Description Key Operation Switch the active channel. Step 3. Check the test ports assigned to ports 1 to 4 in the E5091A properties and connect the 4-port ECal module to those ports. Figure 15-36 Connecting the 4-port ECal Step 4. Execute the calibration. Key Operation Select full 2-port calibration. 2-Port ECal Select the port and execute the calibration.
Measurement Examples Evaluating Transmission Characteristics of a Front End Module Step 5. Perform calibration for channels 2, 3, and 4 according to the same procedure as in Step 2 to Step 4. NOTE Because the test port assignment setting for channels 1 and 2 and that for channels 3 and 4 are the same, you need not change the ECal connection. 9. Connecting DUT Connect the DUT as shown in Figure 15-37.
Measurement Examples Evaluating Transmission Characteristics of a Front End Module 10. Executing Measurement Step 1. Display the trigger menu. Description Key Operation Display the trigger menu. Step 2. Set the trigger source to “manual.” Setting Description Key Operation Set the trigger source to “manual.” Trigger Source - Manual Step 3. Set the trigger mode for channel 1 to “continuous.” Setting Description Key Operation Set channel 1 to the active channel. Set the trigger mode to “continuous.
Measurement Examples Evaluating Transmission Characteristics of a Front End Module Figure 15-38 Example of measuring a front end module 532 Chapter 15
Measurement Examples Executing Power Calibration Executing Power Calibration This section shows an example of executing power calibration using the E4418B power meter and the E4412A power sensor. Overview of execution procedure In this example, power calibration is executed according to the steps shown in Table 15-11. Table 15-11 Execution Procedure of Power Calibration Step Description “1. Connecting Power Meter” on page 533 Connects the power meter to the E5070B/E5071B. “2.
Measurement Examples Executing Power Calibration 2. Setting Address of Power Meter Follow these steps to configure the power meter's GPIB address. Setting Description GPIB address of the power meter: 14 Key Operation - Misc Setup - GPIB Setup - Power Meter Address - 3.
Measurement Examples Executing Power Calibration 6. Connecting Power Sensor Connect the power sensor as shown in Figure 15-40. Figure 15-40 Connecting power sensor 7. Measuring Calibration Data Follow this step to measure the calibration data: Setting Description Measuring calibration data - Power Calibration - Take Cal Sweep 535 15.
Measurement Examples Executing Power Calibration 536 Chapter 15
16. Specifications and Supplemental Information 16 Specifications and Supplemental Information This chapter provides specifications and supplemental information for the Agilent E5070B/E5071B Network Analyzer.
Specifications and Supplemental Information Definitions Definitions All specifications apply over a 5°C to 40°C range (unless otherwise stated) and 90 minutes after the instrument has been turned on. Specification (spec.): Warranted performance. Specifications include guardbands to account for the expected statistical performance distribution, measurement uncertainties, and changes in performance due to environmental conditions.
Corrected System Performance The specifications in this section apply for measurements made with the Agilent E5070B/E5071B Network Analyzer under the following conditions: • • • Table 16-1 No averaging applied to data Environmental temperature of 23°C ±5°C, with less than 1°C deviation from the calibration temperature Response and isolation calibration not omitted System Dynamic Range Description Specification Supplemental Information System Dynamic Range*1*2 300 kHz to 3 MHz IF bandwidth = 3 kHz 8
Specifications and Supplemental Information Corrected System Performance Table 16-2 Corrected System Performance With Type-N Device Connectors, 85032F Calibration Kit Network analyzer: E5070B/E5071B, Calibration kit: 85032F (Type-N, 50 Ω), Calibration: full 2-port IF bandwidth = 10 Hz, No averaging applied to data, Environmental temperature = 23°C±5°C with <1°C deviation from calibration temperature, Isolation calibration not omitted Specification (dB) Description 3 MHz to 3 GHz 3 GHz to 6 GHz 6 GHz to
Table 16-3 Corrected System Performance With Type-N Device Connectors, 85092C Electronic Calibration Module Network analyzer: E5070B/E5071B, Calibration module: 85092C (Type-N, 50 Ω) electronic calibration (ECal) module, Calibration: full 2-port IF bandwidth = 10 Hz, No averaging applied to data, Environmental temperature = 23°C±5°C with <1°C deviation from calibration temperature, Isolation calibration not omitted Specification (dB) Description 3 MHz to 3 GHz 3 GHz to 6 GHz 6 GHz to 8.
Specifications and Supplemental Information Corrected System Performance Table 16-4 Corrected System Performance With 3.5 mm Device Connector Type, 85033E Calibration Kit Network analyzer: E5070B/E5071B, Calibration kit: 85033E (3.
Table 16-5 Corrected System Performance With 3.5 mm Device Connector Type, 85093C Electronic Calibration Module Network analyzer: E5070B/E5071B, Calibration module: 85093C (3.5 mm, 50 Ω) electronic calibration (ECal) module, Calibration: full 2-port IF bandwidth = 10 Hz, No averaging applied to data, Environmental temperature = 23°C±5°C with <1°C deviation from calibration temperature, Isolation calibration not omitted Specification (dB) Description 3 MHz to 3 GHz 3 GHz to 6 GHz 6 GHz to 8.
Specifications and Supplemental Information Uncorrected System Performance Uncorrected System Performance Table 16-6 Uncorrected System Performance (Correction: Off, System Correction: On) Description Specification 3 MHz to 3 GHz 3 GHz to 6 GHz 6 GHz to 8.5 GHz Directivity 25 dB 20 dB 15 dB Source Match 25 dB 20 dB 15 dB Load Match 17 dB 12 dB 10 dB Transmission Tracking ± 1.0 dB ± 1.0 dB ± 1.0 dB Reflection Tracking ± 1.0 dB ± 1.0 dB ± 1.
Test Port Output (Source) Table 16-7 Test Port Output Frequency Description Specification Range E5070B E5071B 300 kHz to 3 GHz 300 kHz to 8.5 GHz Resolution 1 Hz Source Stability Standard Option 1E5 ±5 ppm (5°C to 40°C, typical) ±0.05 ppm (23°C±5°C, typical) ±0.5 ppm/year (typical) CW Accuracy Standard Option 1E5 Table 16-8 Supplemental Information ±5 ppm, 23°C±5°C ±1 ppm, 23°C±5°C Test Port Output Power*1 Description Specification Level Accuracy (at 23°C±5°C) 300 kHz to 10 MHz 10 MHz to 8.
Specifications and Supplemental Information Test Port Output (Source) Table 16-8 Test Port Output Power*1 Description Specification Supplemental Information Level Linearity (high temperature mode: ON) 300 kHz to 3 GHz 3 GHz to 4.25 GHz 4.25 GHz to 6 GHz 6 GHz to 8.5 GHz ±1.5 dB (at −15 dBm to 10 dBm) ±1.5 dB (at −15 dBm to 9 dBm) ±2.0 dB (at −15 dBm to 7 dBm) ±2.0 dB (at −15 dBm to 5 dBm) Level Linearity (swept mode) 300 kHz to 3 GHz 3 GHz to 4.25 GHz 4.25 GHz to 6 GHz 6 GHz to 8.5 GHz ±1.
Test Port Input Table 16-10 Test Port Input Levels Description Specification Supplemental Information Maximum Test Port Input Level 300 kHz to 3 GHz 3 GHz to 4.25 GHz 4.25 GHz to 6 GHz 6 GHz to 8.5 GHz +10 dBm +9 dBm +7 dBm +5 dBm Damage Level 300 kHz to 8.5 GHz RF +20 dBm*1, ±10 VDC (Source attenuator = 0 dB), ±25 VDC (Source attenuator = 5 dB or more), typical Crosstalk*2 3 MHz to 3 GHz 3 GHz to 6 GHz 6 GHz to 7.5 GHz 7.5 GHz to 8.5 GHz −120 dB −109 dB −99 dB −89 dB *1.
Specifications and Supplemental Information Test Port Input Table 16-11 Test Port Input (Trace Noise) Description Specification Supplemental Information Trace Noise*1 Magnitude 300 kHz to 3 MHz (source power level = +10 dBm) 5 mdB rms (typical) 8 mdB rms (high temperature mode: ON, typical) 3 MHz to 3 GHz (source power level = +10 dBm) 1 mdB rms (23°C±5°C) 4 mdB rms (high temperature mode: ON, typical) 3 GHz to 4.25 GHz (source power level = +9 dBm) 1.2 mdB rms (23°C±5°C) 4.
Table 16-12 Test Port Input (Stability) Description Specification Supplemental Information Stability Magnitude*1 3 MHz to 3 GHz 0.005 dB/°C (at 23 °C±5°C, typical) 3 GHz to 6 GHz 0.01 dB/°C (at 23 °C±5°C, typical) 6 GHz to 8.5 GHz 0.04 dB/°C (at 23 °C±5°C, typical) Stability Phase*1 3 MHz to 3 GHz 0.1 °/°C (at 23 °C±5°C, typical) 3 GHz to 6 GHz 0.2 °/°C (at 23 °C±5°C, typical) 6 GHz to 8.5 GHz 0.8 °/°C (at 23 °C±5°C, typical) *1.Stability is defined as a ratio measurement at the test port.
Specifications and Supplemental Information Test Port Input Table 16-14 Test Port Input (Group Delay)*1 Description Specification Aperture (selectable) (frequency span)/(number of points − 1) Maximum Aperture 25% of frequency span Supplemental Information Minimum Delay Limited to measuring no more than 180° of phase change within the minimum aperture. Accuracy See graph below, typical The following graph shows group delay accuracy with type-N full 2-port calibration and a 10 Hz IF bandwidth.
General Information Table 16-15 System Bandwidths Description Supplemental Information IF Bandwidth Settings Range Table 16-16 10 Hz to 100 kHz Nominal settings are: 10, 15, 20, 30, 40, 50, 70, 100, 150, 200, 300, 400, 500, 700, 1k, 1.5k, 2k, 3k, 4k, 5k, 7k, 10k, 15k, 20k, 30k, 40k, 50k, 70k, 100kHz Front Panel Information Description Supplemental Information RF Connectors Type Type-N, female, 50 Ω (nominal) Display Size 10.4 in TFT color LCD Resolution VGA (640 × 480)*1 *1.
Specifications and Supplemental Information General Information Table 16-17 Rear Panel Information Description Supplemental Information External Trigger Connector Type BNC, female Input level LOW threshold voltage: 0.5 V HIGH threshold voltage: 2.
Table 16-17 Rear Panel Information Description Supplemental Information Handler I/O Port 36-pin centronics, female; provides connection to handler system Line Power*2 Frequency 47 Hz to 63 Hz Voltage 90 to 132 VAC, or 198 to 264 VAC (automatically switched) VA Max 350 VA max. *1.USB Test and Measurement Class (TMC) interface that communicates over USB, complying with the IEEE 488.1 and IEEE 488.2 standards. *2.A third-wire ground is required.
Specifications and Supplemental Information General Information Table 16-18 EMC, Safety and Environment Description Supplemental Information European Council Directive 73/23/EEC IEC 61010-1:1990+A1+A2 / EN 61010-1:1993+A2 INSTALLATION CATEGORY II, POLLUTION DEGREE 2 INDOOR USE IEC60825-1:1994 CLASS 1 LED PRODUCT CAN/CSA C22.2 No. 1010.1-92 Environment This product complies with the WEEE Directive (2002/96/EC) marking requirements.
Table 16-19 Analyzer Environment and Dimensions Description Supplemental Information Operating Environment Temperature +5 °C to +40 °C Error-Corrected Temperature Range 23 °C ± 5 °C with < 1°C deviation from calibration temperature Humidity 20% to 80% at wet bulb temperature < +29 °C (non-condensing) Altitude 0 to 2,000 m (0 to 6,561 feet) Vibration 0.
Specifications and Supplemental Information General Information Figure 16-2 Dimensions (front view, E5071B with Option 313, in millimeters, nominal) Figure 16-3 Dimensions (front view, E5071B with Option 213, in millimeters, nominal) 556 Chapter 16
Figure 16-4 Dimensions (rear view, with Option 1E5, in millimeters, nominal) Figure 16-5 Dimensions (side view, in millimeters, nominal) Chapter 16 557 16.
Specifications and Supplemental Information General Information Figure 16-6 Dimensions (top view, in millimeters, nominal) 558 Chapter 16
Measurement Throughput Summary Table 16-20 Typical Cycle Time for Measurement Completion*1*2 (ms) Number of Points 51 201 401 1601 Start 1 GHz, Stop 1.2 GHz, 100 kHz IF bandwidth Uncorrected 4 5 7 18 2-port cal 5 8 13 42 Start 300 kHz, Stop 3 GHz, 100 kHz IF bandwidth Uncorrected 11 12 13 23 2-port cal 20 23 25 46 Start 300 kHz, Stop 8.5 GHz, 100 kHz IF bandwidth Uncorrected 19 24 24 24 2-port cal 37 46 48 50 *1.Typical performance. *2.Sweep mode: Fast swept.
Specifications and Supplemental Information Measurement Throughput Summary Typical Cycle Time for Measurement Completion*1*2 (ms) Table 16-22 Number of Points 51 201 401 1601 Start 1 GHz, Stop 1.2 GHz, 100 kHz IF bandwidth Uncorrected 7 17 29 90 2-port cal 12 32 55 178 Start 300 kHz, Stop 3 GHz, 100 kHz IF bandwidth Uncorrected 14 27 43 130 2-port cal 26 50 84 258 Start 300 kHz, Stop 8.5 GHz, 100 kHz IF bandwidth Uncorrected 16 30 49 146 2-port cal 30 57 96 291 *1.
Table 16-24 Data Transfer Time*1 (ms) Number of Points 51 201 401 1601 64-bit floating point 5 16 29 109 ASCII 21 79 156 617 REAL 64 2 2 3 5 ASCII 34 128 254 995 SCPI over GPIB*2 SCPI over 100 Mbps LAN (Telnet)*2 SCPI over 100 Mbps LAN (SICL-LAN)*2 REAL 64 4 4 5 8 ASCII 6 14 26 95 REAL 64 4 4 5 8 ASCII 6 14 26 95 1 1 1 1 SCPI over USB (SICL-USB)*3 COM*4 Variant type *1.Typical performance. *2.Measured using a VEE 6.
Specifications and Supplemental Information Measurement capabilities Measurement capabilities Number of measurement channels Up to 16 independent measurement channels. A measurement channel is coupled to stimulus response settings including frequency, IF bandwidth, power level, and number of points. Number of display windows Each measurement channel has a display window. Up to 16 display windows (channels) can be displayed.
Source control Measured number of points per sweep User definable from 2 to 1601. Sweep mode Normal stepped, normal swept, fast stepped and fast swept. Sweep type Linear sweep, segment sweep, log sweep and power sweep. Segment sweep Define independent sweep segments. Set number of points, test port power levels, IF bandwidth, delay time, sweep time and sweep mode independently for each segment. Sweep trigger Set to continuous, hold, or single, sweep with internal, external, manual, or bus trigger.
Specifications and Supplemental Information Data accuracy enhancement Data accuracy enhancement Measurement calibration significantly reduces measurement uncertainty due to errors caused by system directivity, source and load match, tracking and crosstalk. Full 2-port, 3-port, or 4-port calibration removes all the systematic errors for the related test ports to obtain the most accurate measurements.
Storage Internal hard disk drive Store and recall instrument states, calibration data, and trace data on 3 GB, minimum, internal hard drive. Trace data can be saved in CSV (comma separated value) format. All files are MS-DOS® -compatible. Instrument states include control settings, limit lines, segment sweep tables, and memory trace data. File sharing Internal hard disk drive (D:) can be accessed from an external Windows® PC through LAN.
Specifications and Supplemental Information Automation Automation Methods Internal analyzer execution Controlling via GPIB Applications can be developed in a built-in VBA® (Visual Basic for Applications) language. Applications can be executed from within the analyzer via COM (component object model) or using SCPI. The GPIB interface operates to IEEE 488.2 and SCPI protocols. The analyzer can be controlled by a GPIB external controller. The analyzer can control external devices using a USB/GPIB interface.
17. Measurement Accessories 17 Measurement Accessories This chapter introduces the accessories that can be used with the Agilent E5070B/E5071B for various measurements.
Measurement Accessories Test Port Cables Test Port Cables The following cables are used to connect the DUT and the network analyzer. N6314A 50 Ω N Type RF Cable (300 kHz ~ 9 GHz) An RF cable 610 mm in length with male N type connectors on both ends. N6315A 50 Ω N Type RF Cable (300 kHz ~ 9 GHz) An RF cable 610 mm in length with a male and a female N type connector on each end.
Measurement Accessories Calibration Kits Calibration Kits Calibration kits are used to improve the accuracy of the analyzer in various measurements. The electronic type reduces the time required for calibration, mis-connections, and wear on connectors since it requires fewer changes of connection than the mechanical type. Specifications for calibration kits and the availability of particular calibration kits are subject to change without prior notice.
Measurement Accessories Calibration Kits 85054D Economy Mechanical Calibration Kit N Type (dc ~ 18 GHz) The 85054D includes the followings. Agilent Cat. No.
Measurement Accessories Calibration Kits Coaxial Electronic Calibration Kits 85092C RF Two-Port ECal Module (300 kHz ~ 9 GHz) Description Option M0F Module with 50 Ω N type (m)/N type (f) connectors Option 00M Module with 50 Ω N type (m)/N type (m) connectors Option 00F Module with 50 Ω N type (f)/N type (f) connectors 17.
Measurement Accessories Calibration Kits For Devices with 3.5 mm (SMA) Connectors Coaxial Mechanical Calibration Kits 85033E Mechanical Calibration Kit 3.5 mm 50 Ω (30 kHz ~ 9 GHz) The 85033E kit includes the following items. Agilent Cat. No. Description 85033-60016 3.5 mm (m), terminated 85033-60017 3.5 mm (f), terminated 85033-60018 3.5 mm (m), open 85033-60019 3.5 mm (f), open 85033-60020 3.5 mm (m), short 85033-60021 3.
Measurement Accessories Calibration Kits 85052C Mechanical Calibration Kit 3.5 mm (45 MHz ~ 26.5 GHz) The 85052C kit includes the following items. Description 00902-60003 3.5 mm (m), terminated 00902-60004 3.5 mm (f), terminated 85052-60006 3.5 mm (m), short 85052-60007 3.5 mm (f), short 85052-60008 3.5 mm (m), open 85052-60009 3.5 mm (f), open 85052-60032 3.5 mm (f)-(f) adaptor 85052-60033 3.5 mm (m)-(m) adaptor 85052-60034 3.5 mm (f)-(m) adaptor 85052-60035 3.
Measurement Accessories Calibration Kits Coaxial Electronic Calibration Kits 85093C RF Two-Port ECal Module (300 kHz ~ 9 GHz) Option Description Option M0F Module with 3.5 mm (m)/3.5 mm (f) connectors Option 00M Module with 3.5 mm (m)/3.5 mm (m) connectors Option 00F Module with 3.5 mm (f)/3.5 mm (f) connectors N4431A RF Four-Port ECal Module (300 kHz ~ 9 GHz) Option Description Option 010 Module with four 3.
Measurement Accessories Calibration Kits For Devices with 7 mm Connectors Coaxial Mechanical Calibration Kits 85031B Mechanical Calibration Kit 7 mm 50 Ω (300 kHz ~ 6 GHz) The 85031B kit includes the following items. Description 00909-60008 7 mm , broadband load 85031-60001 7 mm , combination open/short 17. Measurement Accessories Agilent Cat. No. 85050C Mechanical Calibration Kit 7 mm 50 Ω (45 MHz ~ 18 GHz) The 85050C kit includes the following items. Agilent Cat. No.
Measurement Accessories Calibration Kits For Devices with 7-16 Connectors Coaxial Mechanical Calibration Kits 85038A Mechanical Calibration Kit 7-16 50 Ω (DC ~ 7.5 GHz) The 85038A kit includes the following items. Agilent Cat. No. Description 85038-80007 7-16 (m) broadband load 85038-80006 7-16 (f) broadband load 85038-80005 7-16 (m) short 85038-80004 7-16 (f) short 85038-80003 7-16 (m) open 85038-80002 7-16 (f) open 85038F Mechanical Calibration Kit 7-16 50 Ω (DC ~ 7.
Measurement Accessories Adaptors Adaptors 11853A 50 Ω N Type Accessory Kit The 11853A kit includes the following items. Description 1250-1472 N type (f)-(f) adaptor kit (two adaptors) 1250-1475 N type (m)-(m) adaptor kit (two adaptors) 11511A N type (f), short 11512A N type (m), short 17. Measurement Accessories Agilent Cat. No. 11878A N type to 3.5 mm Adaptor Kit The 11878A kit includes the following items. Agilent Cat. No. Description 1250-1744 3.
Measurement Accessories System Accessories System Accessories System Racks and Cases CAUTION Option Agilent Cat. No.
18. Information on Maintenance 18 Information on Maintenance This chapter explains the measures you should take to maintain the Agilent E5070B/E5071B.
Information on Maintenance Backing Up the Data Backing Up the Data Be sure to back up regularly your important data (including program) files in this instrument to a CD-R or other backup medium. Agilent Technologies shall not be liable for any data damages caused by troubles of this instrument. Making Backup Files Making backup files on a floppy disk You can make backup files on a floppy disk using the copy function. See “Organizing Files and Folders” on page 369 for making a copy.
Information on Maintenance Removing/Mounting Removable Hard Disk Removing/Mounting Removable Hard Disk This section provides information on how to mount and remove a removable hard disk as well as write the system correction data. Whether a hard disk is removable can be ascertained by referring to the serial number. For further information, see “Change 9” on page 604 in the “Manual Changes.” CAUTION For calibration or repair of the E5070B/E5071B, send it with the removable hard disk mounted.
Information on Maintenance Removing/Mounting Removable Hard Disk Figure 18-1 Mounting/Removing Removable Hard Disk -1 Figure 18-2 Mounting/Removing Removable Hard Disk -2 582 Chapter 18
Information on Maintenance Removing/Mounting Removable Hard Disk Writing System Correction Data from Backup Memory to Removable Hard Disk The following procedure shows how to write the system correction data stored in the backup memory to the newly mounted removable hard disk. Step 1. Turn on the E5070B/E5071B. Step 2. Press . Step 3. Press Load Project... in the softkey menu. Step 4. A dialog box appears for you to select the program to be loaded. Select RestoreSysCorFile.
Information on Maintenance Cleaning this Instrument Cleaning this Instrument This section describes how to clean the instrument. WARNING To protect yourself from electrical shock, be sure to unplug the power cable from the outlet before cleaning the instrument. Never clean the internal components of the instrument. Cleaning an LCD Use one of the following methods to clean the display surface regularly. NOTE • For normal cleaning, rub the surface gently with a dry, soft cloth.
Information on Maintenance Replacement of Parts with Limited Service Life Replacement of Parts with Limited Service Life This instrument incorporates parts with limited service life as shown in Table 18-1. Using the recommended replacement time shown in Table 18-1 as a guide, request the Company’s Service Center to replace these parts. However, a part may need to be replaced at an earlier time than that listed in the table, depending on such conditions as location, frequency of use, and where it is stored.
Information on Maintenance Cautions Applicable to Requesting Repair, Replacement, Regular Calibration, etc. Cautions Applicable to Requesting Repair, Replacement, Regular Calibration, etc. Backing Up Data in the Hard Disk The user is requested to back up the stored programs and data into external media by using the instrument’s storing function before requesting the Company’s Service Center to repair the instrument or replace hard disks. See “Making Backup Files” on page 580 for how to make backup files.
A. Manual Changes A Manual Changes This appendix contains the information required to adapt this manual to versions or configurations of the E5070B/E5071B manufactured earlier than the current printing date of this manual. The information in this manual applies directly to E5070B/E5071B units having the serial number printed on the title page of this manual.
Manual Changes Manual Changes Manual Changes To adapt this manual to your E5070B/E5071B, refer to Table A-1, and Table A-2. Table A-1 Table A-2 Manual Changes by Serial Number Serial Prefix or Number Make Manual Changes MY42100211 (E5070B), MY42100407 (E5071B) Change 2 MY421, JP1KK Change 5 MY422 Change 6 MY423 Change 9 Manual Changes by Firmware Version Version Make Manual Changes A.03.0x Change 1 A.03.53 Change 3 A.03.54 Change 4 A.03.62 Change 7 A.04.00 Change 8 A.05.
Manual Changes Manual Changes Change 15 The firmware rivision A.08.01 or lower does not support the following functions. Please delete the descriptions about these functions from this manual. o Notch search function Change 14 The firmware rivision A.06.50 or lower does not support the following functions. Please delete the descriptions about these functions from this manual.
Manual Changes Manual Changes This VBA macro lets you perform 2-/3-/4-port TRL/LRM calibration for any selected port. TRL/LRM calibration provides a level of accuracy equivalent to full 2-/3-/4-port calibration using short, open, thru (SLOT) standards. To perform TRL calibration, you need to prepare thru, reflection (open or short), and line standards. The thru and line must have the same reference impedance Z0 and transfer constant (i.e., be the same material).
Manual Changes Manual Changes Figure A-2 Turning off system error correction NOTE The System Correction Off button does not appear when system error correction has already been turned off. 2-port TRL/LRM calibration operating procedure 1. Selecting Test Port Select test ports (1 in Figure A-3). NOTE For more information on starting VBA macro, see “Operating Procedure” on page 590 . A.
Manual Changes Manual Changes Figure A-4 Calibration kit definition screen Step 2. Define each standard as follows Reference impedance (Z0) Enter a value of the reference impedance of thru, line and match (1 in Figure A-4). Reflection Select a standard type (2 in Figure A-4) and enter a delay value (3 in Figure A-4) in ps. Thru Enter a delay value (4 in Figure A-4) in ps. If necessary, also enter its offset loss value (5 in Figure A-4) in Gohm/sec. Match Enter a frequency range (6 in Figure A-4) in GHz.
Manual Changes Manual Changes NOTE When you define each calibration standard, the following points should be considered. • Reference impedance Z0 must be set to the same value as the ENA’s system impedance Z0 value. • When you use Line 1, Line 2, and Line 3, their defined frequency ranges must overlap by at least 10 kHz. Saving and loading calibration kit definitions Press the Save button (11 in Figure A-4) to save the definition of the current calibration kit to a file.
Manual Changes Manual Changes Figure A-5 Performing 2-port TRL/LRM calibration Step 2. Press the Thru button (1 in Figure A-6). Step 3. Make the through condition for test ports 1 and 2 and press the [ ]Thru 1-2 button (2 in Figure A-6). The through measurement is performed, and an asterisk appears at the [ ]Thru 1-2 button and then a check mark [v] also appears on the Thru button. Figure A-6 Example of performing 2-port TRL/LRM calibration Step 4. Press the Reflection button (3 in Figure A-6). Step 5.
Manual Changes Manual Changes The line measurement is performed, and an asterisk appears at the [ ]Line1 1-2 button and then a check mark [v] also appears on the Line1 button. Step 12. Press the Line2 button (6 in Figure A-6). Step 13. Connect the Line 2 standard between test ports 1 and 2 and press the [ ]Line2 1-2 button. The line measurement is performed, and an asterisk appears at the [ ]Line2 1-2 button and then a check mark [v] also appears on the Line2 button.
Manual Changes Manual Changes For more information on defining a calibration kit, see ??. 3. Performing Calibration Measure the necessary calibration data and enable error correction. NOTE The definition of the frequency range of the line standard and match standard used for measurement must cover the sweep range of the channel for which you perform calibration.
Manual Changes Manual Changes Calibration Type Measurement ports of Reflection and Match cal. Measurement paths of Thru and Line cal. 3-port TRL/LRM cal. for Ports: 1-2-3 All ports (1, 2, and 3) 1-2 and 1-3 3-port TRL/LRM cal. for Ports: 1-2-4 All ports (1, 2, and 4) 1-2 and 1-4 3-port TRL/LRM cal. for Ports: 1-3-4 All ports (1, 3, and 4) 1-3 and 3-4 3-port TRL/LRM cal. for Ports: 2-3-4 All ports (2, 3, and 4) 2-3 and 3-4 4-port TRL/LRM cal.
Manual Changes Manual Changes Figure A-9 Example of performing multiport TRL/LRM calibration Step 5. Press the Reflection button (4 in Figure A-9). Step 6. Connect the Reflection standard to test port 1 and press the [ ]Reflection1 button. The reflection measurement is performed, and an asterisk appears at the [ ]Reflection1 button. Step 7. Connect the Reflection standard to test port 2 and press the [ ]Reflection2 button.
Manual Changes Manual Changes The line measurement is performed, and an asterisk appears at the [ ]Line2 1-2 button. Step 18. Connect the Line 2 standard between test ports 1 and 3 and press the [ ]Line2 1-3 button. The line measurement is performed, and an asterisk appears at the [ ]Line2 1-3 button and then a check mark [v] also appears on the Line2 button. NOTE When the frequency ranges of the line and the match overlap, the data of the line is used in the overlapped frequency area.
Manual Changes Manual Changes Note on use When the fixture simulator is ON and the port impedance conversion is ON, Z0 of all ports to be saved must be set to the same value. Operating Procedure 1. Starting VBA macro Step 1. Press [Macro Setup]. Step 2. Press Load Project. Step 3. The Open dialog box appears. Specify the file name “D:\Agilent\SaveToTouchstone.vba” and press the Open button. Step 4. Press [Macro Run] to start the macro. (Refer to Figure A-10.) Figure A-10 SaveToTouchstone dialog box 2.
Manual Changes Manual Changes once. NOTE Regardless of the on/off state of the balance-unbalance conversion, measurement is performed without the balance-unbalance conversion. Step 5. When the measurement is complete, the Save As dialog box appears. Specify a file name and press the Save button. Step 6. When saving to the file is complete, the start screen appears again. 3. Closing VBA macro Step 1. Press the Close button (6 in Figure A-10) to exit from the macro.
Manual Changes Manual Changes Figure A-12 Two-port Touchstone file Figure A-13 Three-port Touchstone file 602 Appendix A
Manual Changes Manual Changes Figure A-14 Four-port Touchstone file Change 10 The firmware revision A.05.00 or lower does not support the following functions. Please delete the descriptions about these functions from this manual. User preset function o Function to display the marker values for non-active traces. o Function to change the display position where the marker values are displayed. o Function to align the marker values.
Manual Changes Manual Changes o Function to have the multiport test set E5091A-016 correspond to this instrument. o Function to select the 85052C as the calibration kit. o Function to specify up to eight calibration standards for each calibration class. o Reading/Writing files of the calibration standard. o Function to set the number of measurement points to a maximum of 20001.
Manual Changes Manual Changes Figure A-16 Dimensions (side view, in millimeters, nominal) Change 8 The firmware revision A.04.00 or below does not support the following functions. Please disregard the descriptions of these functions in this manual. Offset limit line function o Ripple test function o Bandwidth test function o Enhanced response calibration o Frequency information appearing as asterisks o Disable USB mass storage devices Change 7 The firmware revision A.03.
Manual Changes Manual Changes o Assignable x-axis such as RF+LO, RF-LO, and LO-RF frequencies for each active trace o Conjugation for converting vector mixer measurement parameters o 7 mm calibration kits such as 85031B and 85050C/D o Calibration data and calibration coefficient clear functions Change 6 Use Following specifications for the serial prefix MY422. Table A-3 Description Specification Supplemental Information Damage Level 300 kHz to 8.
Manual Changes Manual Changes *1.The test port dynamic range is calculated as the difference between the test port rms noise floor and the source maximum output power. The effective dynamic range must take measurement uncertainty and interfering signals into account. *2.May be limited to 90 dB at particular frequencies below 350MHz or above 4.25GHz due to spurious receiver residuals.
Manual Changes Manual Changes Table A-5 Test Port Output Power*1 Description Sweep Range (without extended power range) 300 kHz to 4.25 GHz 4.25 GHz to 6 GHz 6 GHz to 8.5 GHz Level Resolution Specification Supplemental Information −15 dBm to 10 dBm −15 dBm to 8 dBm −15 dBm to 6 dBm −20 dBm to 10 dBm −20 dBm to 8 dBm −20 dBm to 6 dBm 0.05 dB *1.Source output performance on port 1 only. Other port output performance is typical.
Manual Changes Manual Changes Table A-7 Test Port Input (Trace Noise) Description Specification Supplemental Information 300 kHz to 3 MHz (source power level = +10 dBm) 0.035 ° rms (23°C±5°C, typical) 0.05 ° rms (high temperature mode: ON, typical) 3 MHz to 4.25 GHz (source power level = +10 dBm) 0.007 ° rms (23°C±5°C, typical) 0.02 ° rms (high temperature mode: ON, typical) 4.25 GHz to 6 GHz (source power level = +8 dBm) 0.021 ° rms (23°C±5°C, typical) 0.
Manual Changes Manual Changes o Simplified full 3/4 port calibration o Loading and executing program in batch process 610 Appendix A
B Troubleshooting 611 B. Troubleshooting This Chapter describes the steps to take in troubleshooting when your Agilent E5070B/E5071B appears to be operating improperly. Explanations are also given for the error warning messages displayed on the screen.
Troubleshooting Troubleshooting Troubleshooting This section describes the steps you should take when you believe the Agilent E5070B/E5071B is operating improperly. The results of these simple investigative procedures may help you avoid the down-time and inconvenience of repair service.
Troubleshooting Troubleshooting Troubleshooting during Operation The Sweep Action Stops during Measurement or Is Not Executed. • The sweep action stops during measurement or is not executed, but the front keys and softkeys are operational. There is a possibility of a failure. Contact Agilent Technology’s Customer Contact listed at the end of this guide or the company from which you bought the device. The Error Message “Port N receiver overload” (N denotes a port number) is Displayed.
Troubleshooting Troubleshooting • The touch screen becomes inoperable. ³ Using the front panel keys, turn System – Key Lock – Touch Screen & Mouse Lock OFF. ³ Execute the calibration of the touch screen. For information on the execution procedure, see “Calibration of the Touch Screen” on page 474. • The mouse becomes inoperable. ³ Using the front panel keys, turn System – Key Lock – Touch Screen & Mouse Lock OFF. • All of the front panel keys, keyboard, and mouse become inoperable.
Troubleshooting Troubleshooting possibility of a failure. Contact Agilent Technology’s Customer Contact listed at the end of this guide or the company from which you bought the device. B.
Troubleshooting Troubleshooting You Cannot Read a File from a Floppy Disk. • You cannot read a file from a floppy disk. ³ Confirm that the floppy disk is inserted correctly. Insert the floppy disk until the eject button pops up fully. ³ A stored file may be damaged by a magnetic field. Confirm that the file can be read on a PC. When taking all these measures does not make it possible to read the file, there is a possibility of a failure.
Troubleshooting Error Messages Error Messages An error message is displayed against a red background in the instrument message/warning area in the lower left part of the screen. Pushing a front panel key or executing :DISP:CCL command clears the error message. Errors caused by the operation of a front panel key simply appear on the display. They are not stored in the error queue with some exceptions. An error with a positive error number is one uniquely defined for this instrument.
Troubleshooting Error Messages -100 Command error A comprehensive syntax error has occurred showing that this instrument cannot detect a more detailed error. This code simply shows that a command error defined in 11.5.1.1.4, IEEE488.2 has occurred. 60 Continuous switching may damage source attenuator This error occurs when different source attenuator (power range) settings are present during measurement on two or more channels.
Troubleshooting Error Messages -178 Expression data not allowed An expression-data element has been received at a position where this instrument does not accept one. -170 Expression error When the expression data is put to syntactic analysis, an error not corresponding to one of Error Numbers -171 through -179 occurs.
Troubleshooting Error Messages -256 File name not found The file name specified is not found and hence the command is not executed correctly. This error occurs when you try to read a file that does not exist in a disk or a disk is not correctly inserted into the drive to read or write a file, for example. 107 File transfer failed This error occurs when writing data into or reading data from a file (MMEM:TRAN command) fails.
Troubleshooting Error Messages -213 Init ignored Because another measurement is in progress, the request for initiating a measurement (“INIT” command) is ignored. 35 Insufficient ECal module memory This error occurs when the embedded memory is insufficient to save the user property in ECal module during the user definition ECal processing. -161 Invalid block data Block data has been expected, but the block data that appears is invalid for some reason (see 7.7.6.2, IEEE488.2).
Troubleshooting Error Messages The suffix does not comply with the syntax defined in 7.7.3.2, IEEE488.2. Or it does not suit E5070B/E5071B. L 53 Log sweep requires 2 octave minimum span The span of sweep range is not satisfied the requirement for logarithmic sweep. The sweep type is automatically changed to linear sweep when this error occurs. For example, this error occurs when, with the start and stop frequency are set 1 MHz and 2 MHz respectively, the sweep type is changed to logarithmic sweep.
Troubleshooting Error Messages where this instrument does not accept one. O 200 Option not installed The command received has been ignored because of the mismatch between the contents of an option for this instrument and the command. For example, this error occurs when the source attenuator (power range) is set at a value other than zero (SOUR:POW:ATT command) in a model not having the extended power output option.
Troubleshooting Error Messages 222 Port 2 receiver overload The input to Test Port 2 exceeds the maximum input level. The measurement value is not correct. When a DUT is an amplifier or the like, this error may occur, damaging the receiver in the worst case. Should this error occur with a passive part used as the DUT or with nothing connected to the test port, this instrument is faulty. Contact an Agilent Technology sales office or the company from which you bought the instrument.
Troubleshooting Error Messages This error occurs when the previous printing is still in progress or the printer fails (offline, short of paper, etc.) at time of outputting the display image on the LCD screen to the printer (HCOP:IMM command). 121 Print failed This error occurs when printing fails for reasons other than Error 120, Printer error. -284 Program currently running This error occurs when the PROG:SEL:STAT RUN command is executed with the VBA program in the Run state.
Troubleshooting Error Messages This error occurs when reading an instrument status file (State01.sta, etc.) (MMEM:LOAD:STAT command) fails. S 106 Save failed This error occurs when writing an instrument status file (State01.sta, etc.) (MMEM:STOR:STAT command) fails. 33 Selected parameter not valid for confidence check This error occurs when the mix mode S parameter has been selected for the S parameter you want to check, while using the confidence check function for calibration coefficient.
Troubleshooting Error Messages -138 Suffix not allowed A suffix is attached to a numeric value element to which a suffix is not allowed to be attached. -134 Suffix too long The unit is too long. The unit is expressed in 12 or more characters (see 7.7.3.4, IEEE488.2). 55 Sweep mode changed to stepped sweep You cannot change the sweep mode to the swept mode.
Troubleshooting Error Messages This instrument receives and detects a trigger command (“TRIG”) or an external trigger signal. But it is ignored due to the timing condition (This instrument is not in the wait-fortrigger state, for example). Change the setup so that a trigger command or an external trigger signal can be sent after the instrument has entered the wait-for- trigger state. U 300 Unable to estimate adapter length Adapter length can not be estimated at zero span.
Troubleshooting Error Messages Ensure the connection of the power sensor. 75 Valid signal generator not found No valid external signal source is connected. This error occurs when the specified type of external signal source is not connected to the specified GPIB address, though the control of the external signal source is enabled (SENS:OFFS:LOC:STAT ON and SENS:OFFS:LOC:CONT ON). It also occurs when USB/GPIB interface is set but not connected.
Troubleshooting Warning Message Warning Message A warning message is displayed in the instrument message/Warning area in the lower left part of the display against a gray background. Pushing a front panel key or executing :DISP:CCL command clears the message. This message simply appears on the display, being not known to a remote environment such as a GPIB. This message is not displayed when another error (against a red background) has already been displayed in the instrument message/Warning area.
Troubleshooting Warning Message This message appears when you set the sweep type to power sweep, power slope function to enabled (ON, correction coefficient is other than zero), power calibration to ON, frequency offset to enabled, spurious avoidance of frequency offset to enabled, or external signal source control to enabled, while the sweep mode is set to swept mode or high speed swept mode.
Troubleshooting Warning Message 632 Appendix B
C. List of Default Values C List of Default Values This appendix gives the default values, settings for Save/Recall of an object, and settings for backing up an object when using the Agilent E5070B/E5071B.
List of Default Values List of Default Values, Save/Recall Settings, and Backup Settings List of Default Values, Save/Recall Settings, and Backup Settings The table below shows the following settings for the Agilent E5070B/E5071B.
Key Operation Factory-shipped Setting Default Value Save/ Recall *RST Backup Available Means of Defining a Setting S11 ← ← √ K/C Sss11 (When Fixture Simulator is set on, set Topology at SE-Bal, and, BalUn on in Analysis mode) ← ← √ K/C Sdd11 (When Fixture Simulator is set on, set Topology at Bal-Bal, and BalUn on in Analysis mode) ← ← √ K/C Sss11 (When Fixture Simulator is set on, set Topology at SE-SE-Bal, and BalUn on in Analysis mode) ← ← √ K/C Log Mag ← ← √ K/C Divisions
List of Default Values List of Default Values, Save/Recall Settings, and Backup Settings Key Operation Factory-shipped Setting Default Value Save/ Recall *RST Correction Backup Available Means of Defining a Setting OFF ← ← 1 ← ← K/C 1 ← ← K/C 2-1 (S21) ← ← K/C 2-1 (S21 S11) ← ← K/C 1 ← ← K/C 1-2 ← ← K/C 1-2-3 ← ← K/C 1-2 ← ← K/C 1-2-3 ← ← K/C √ K/C Calibrate Response (Open) Select Port Response (Short) Select Port Response (Thru) Select Ports Enhanced Respons
Key Operation Factory-shipped Setting Default Value Save/ Recall *RST Backup Available Means of Defining a Setting Port Extensions Extensions OFF ← ← √ K/C Auto Port Extension 1,2,3,4 ← ← √ K/C Current Span ← ← √ K/C User Span Start 300.00 kHz ← ← √ K/C User Span Stop 3.0000 GHz (E5070B) 8.
List of Default Values List of Default Values, Save/Recall Settings, and Backup Settings Key Operation Factory-shipped Setting Default Value Save/ Recall *RST Backup Available Means of Defining a Setting Receiver Calibration Select Port 1 ← ← Correction OFF ← ← 1 ← ← K 2-1 (fwd) ← ← K A ← ← K 2-1 (fwd) ← ← K Source Port Mixer/Converter Calibration K √ K/C When Frequency Offset is set on, this softkey is enabled.
Key Operation Factory-shipped Setting Default Value Save/ Recall *RST Backup Available Means of Defining a Setting Sweep Time AUTO ← ← √ K/C Sweep Delay 0.
List of Default Values List of Default Values, Save/Recall Settings, and Backup Settings Key Operation Factory-shipped Setting Default Value Save/ Recall *RST Backup Available Means of Defining a Setting Frequency Offset External Source LO Frequency Multiplier OFF ← ← √ K/C 0.0000 ← ← √ K/C Divisor 1.0000 ← ← √ K/C Offset 0.0000 Hz ← ← √ K/C Start 0.0000 Hz ← ← √ K/C Stop 0.
Key Operation Factory-shipped Setting Default Value Save/ Recall *RST Backup Available Means of Defining a Setting Search Range OFF ← ← √ K/C Start 0.0000 Hz ← ← √ K/C Stop 0.0000 Hz ← ← √ K/C Search Range Couple ON ← ← √ K/C Bandwidth OFF ← ← √ K/C -3.0000 dB (When one of the marker is on) ← ← √ K/C Bandwidth Value OFF ← ← √ K/C -3.
List of Default Values List of Default Values, Save/Recall Settings, and Backup Settings Key Operation Factory-shipped Setting Default Value Save/ Recall *RST Backup Available Means of Defining a Setting Fixture Simulator BalUn OFF ← ← √ K/C OFF ← ← √ K/C Port Matching Port Matching 1 ← ← √ K/C None ← ← √ K/C C 0.000e-12 F ← ← √ K/C G 0.0000 S ← ← √ K/C L 0.000e-9 H ← ← √ K/C R 0.0000 Ω ← ← √ K/C Port ZConversion OFF ← ← √ K/C Port1 Z0 Real 50.
Key Operation Factory-shipped Setting Default Value Save/ Recall *RST Backup Available Means of Defining a Setting Fixture Simulator Diff ZConversion Port3 (bal) Real 100.00 Ω ← ← √ K/C Port3 (bal) Imag 0.00 Ω ← ← √ K/C OFF ← ← √ K/C 25.000 Ω ← ← √ K/C Port1 (bal) Imag 0.000 Ω ← ← √ K/C Port2 (bal) Real 25.000 Ω ← ← √ K/C Port2 (bal) Imag 0.000 Ω ← ← √ K/C Port3 (bal) Real 25.000 Ω ← ← √ K/C Port3 (bal) Imag 0.
List of Default Values List of Default Values, Save/Recall Settings, and Backup Settings Key Operation Factory-shipped Setting Default Value Save/ Recall *RST Backup Available Means of Defining a Setting Limit Test Limit Line Offsets Stimulus Offset 0Hz ← ← √ K/C Amplitude Offset 0Hz ← ← √ K/C ON ← ← √ K/C Ripple Limit Test OFF ← ← √ K/C Ripple Limit OFF ← ← √ K/C Ripple Value OFF ← ← √ K/C Fail Sign Ripple Limit 1 ← ← √ K/C ON ← ← √ K/C BW Test OFF ←
Key Operation Factory-shipped Setting Default Value Save/ Recall *RST Invert Image ON ← ← √ Backup Available Means of Defining a Setting K/C Multiport Test Set Setup Test Set 1 E5091_9 ← ← √ K/C Control OFF ← ← √ K/C Property OFF ← ← √ K/C A ← ← √ K/C Select Test Set Port 1 Port 2 T1 ← ← √ K/C Port 3 R1+ ← ← √ K/C Port 4 R1- ← ← √ K/C *1 Port 5 X1 ← ← √ K/C Port 6*1 Y1 ← ← √ K/C Port 7*1 Z1 ← ← √ K/C LOW ← ← √ K/C Control Lines Li
List of Default Values List of Default Values, Save/Recall Settings, and Backup Settings Key Operation Factory-shipped Setting Default Value Save/ Recall Backup *RST Available Means of Defining a Setting Misc Setup Beeper Beep Complete ON ← ← √ K/C Beep Warning ON ← ← √ K/C Talker/Listener Address 17 Non-changing ← √ K Power Meter Address 13 Non-changing ← √ K/C GPIB Setup Signal Generator Address 19 ← ← √ K/C 100.
Key Operation Factory-shipped Setting Default Value Save/ Recall Backup *RST Available Means of Defining a Setting Misc Setup Key Lock Front Panel & Keyboard Lock OFF ← ← K/C Touch Screen & Mouse Lock OFF ← ← K/C Color Setup Normal Data Trace 1 Red:5 Green:5 Blue:0 ← ← √ K/C Data Trace 2 Red:0 Green:5 Blue:5 ← ← √ K/C Data Trace 3 Red:5 Green:0 Blue:5 ← ← √ K/C Data Trace 4 Red:0 Green:5 Blue:0 ← ← √ K/C Data Trace 5 Red:5 Green:4 Blue:0 ← ← √ K/C Data Trace 6
List of Default Values List of Default Values, Save/Recall Settings, and Backup Settings Key Operation Factory-shipped Setting Default Value Save/ Recall *RST Backlight ON ← ← Backup Available Means of Defining a Setting K/C Service Menu System Correction ON ← ← √ K/C Avoid Spurious ON ← ← √ K/C High Temperature OFF Non-changing ← √ K/C Init Src Ctrl ON ← ← √ K/C Init Src Port Port 1 ← ← √ K/C Security Level None ← ← √ K/C *1.
D. Softkey Functions D Softkey Functions This appendix explains the functions of softkeys and hardkeys supplied on the Agilent E5070B/E5071B.
Softkey Functions E5070B/E5071B Menu (Top Menu) E5070B/E5071B Menu (Top Menu) Key Operation Double-click on each softkey menu title Function Displays the top menu of each menu item below. A preset operation will not cancel the menu display. Measurement Same as Format Same as . See “Format Menu” on page 698. Scale Same as . See “Scale Menu” on page 719. Display Same as . See “Display Menu” on page 694. Average Same as . See “Average Menu” on page 666. Calibration Same as .
Softkey Functions Analysis Menu Analysis Menu Key Operation Function SCPI Command Displays softkeys for performing analysis functions. Fixture Simulator Fixture Simulator Enables or disables the fixture simulator function. When using one or more :CALC{1-16}:FSIM:STAT of the six functions provided with the fixture simulator, the fixture simulator function must be enabled using this key along with all desired functions.
Softkey Functions Analysis Menu Key Operation Function SCPI Command (Continued) Fixture Simulator Topology*1 Port1 (bal)*2 2-3 Selects test ports 2 and 3 on the analyzer for connection with (balanced) port 1 on the DUT (ports a and b in Table 7-8 on page 296) respectively. :CALC{1-16}:FSIM:BAL:TOP:BBAL 2-4 Selects test ports 2 and 4 on the analyzer for connection with (balanced) port 1 on the DUT (ports a and b in Table 7-8 on page 296) respectively.
Softkey Functions Analysis Menu Key Operation Function SCPI Command (Continued) Fixture Simulator Topology*1 Port2 (bal) Selects test ports 4 and 1 on the analyzer for connection with (balanced) port 2 on the DUT (ports b and c for SE-Bal or ports c and d for Bal-Bal in Table 7-8 on page 296) respectively.
Softkey Functions Analysis Menu Key Operation Function SCPI Command (Continued) Fixture Simulator Topology*1 Port3 (bal)*2 4-3 Selects test ports 4 and 3 on the analyzer for connection with (balanced) port 3 on the DUT (ports c and d in Table 7-8 on page 296) respectively. Cancel Returns to the softkey display screen one level higher. :CALC{1-16}:FSIM:BAL:TOP:SSB Property :CALC{1-16}:FSIM:BAL:TOP:PROP:S Enables or disables the display of the balanced measurement topology property.
Softkey Functions Analysis Menu Key Operation Function SCPI Command (Continued) Fixture Simulator Port Matching Select Circuit ShuntL ShuntC Adds a shunt L - shunt C type matching circuit to the port selected in the Select Port menu. User Adds a user-defined matching circuit to the port selected in the Select Port :CALC{1-16}:FSIM:SEND:PMC: PORT{1-4} USER menu.
Softkey Functions Analysis Menu Key Operation Function SCPI Command (Continued) Fixture Simulator De-Embedding Displays softkeys for setting up the network de-embedding function. Network de-embedding is a function used to eliminate any user-defined network (with a reference impedance of 50 W) supplied in a 2-port Touchstone data file, from desired test ports to extend the calibration plane. De-Embedding Enables or disables the network de-embedding function.
Softkey Functions Analysis Menu Key Operation Function SCPI Command (Continued) Fixture Simulator Diff Matching*1 Sets the value of C for the shunt L - shunt C differential matching circuit to :CALC{1-16}:FSIM:BAL:DMC: BPOR{1-2}:PAR:C be added to the balanced port selected in the Select Bal Port menu. G Sets the value of G for the shunt L - shunt C differential matching circuit to :CALC{1-16}:FSIM:BAL:DMC: BPOR{1-2}:PAR:G be added to the balanced port selected in the Select Bal Port menu.
Softkey Functions Analysis Menu Key Operation Function SCPI Command (Continued) Fixture Simulator Cmn ZConversion*1 Displays softkeys for setting up the common impedance conversion function. When the reference impedance of two unbalanced ports is expressed as Z0, an automatic conversion sets the differential mode reference impedance of the converted balanced port to 2Z0 and the common mode reference impedance to Z0/2.
Softkey Functions Analysis Menu Key Operation Function SCPI Command (Continued) Fixture Simulator De-Embedding S4P Topology Ports Sets the ports of the analyzer connected to ports a and b in Figure 7-6 to test :CALC{1-16}:FSIM:EMB:TOP:A:PORT 2,4 ports 2 and 4, respectively. 3-1*1 Sets the ports of the analyzer connected to ports a and b in Figure 7-6 to test :CALC{1-16}:FSIM:EMB:TOP:A:PORT 3,1 ports 3 and 1, respectively.
Softkey Functions Analysis Menu Key Operation Function SCPI Command (Continued) Fixture Simulator De-Embedding S4P Topology Ports 3-2-4*2 Sets the ports of the analyzer connected to ports a, b, and c in Figure 7-6 to :CALC{1-16}:FSIM:EMB:TOP:B:PORT 3,2,4 test ports 3, 2, and 4, respectively. 3-4-1*2 Sets the ports of the analyzer connected to ports a, b, and c in Figure 7-6 to :CALC{1-16}:FSIM:EMB:TOP:B:PORT 3,4,1 test ports 3, 4, and 1, respectively.
Softkey Functions Analysis Menu Key Operation Function SCPI Command (Continued) Fixture Simulator De-Embedding S4P Topology Ports*2 Sets the ports of the analyzer connected to ports a, b, c and d in Figure 7-6 to test ports 3, 1, 2, and 4, respectively. :CALC{1-16}:FSIM:EMB:TOP:C:POR T 3,1,2,4 3-1-4-2 Sets the ports of the analyzer connected to ports a, b, c and d in Figure 7-6 to test ports 3, 1, 4, and 2, respectively.
Softkey Functions Analysis Menu Key Operation Function SCPI Command (Continued) Fixture Simulator De-Embedding S4P Topology Type (nwk2) De-Embed Select de-embedding. Cancel Returns to the softkey display screen in one level higher. User File (nwk2) Opens the dialog box to read in a 4 port touchstone data file that the user prepared for the network 2. You can embed/de-embed the network read in here. Return Returns to the softkey display screen one level higher.
Softkey Functions Analysis Menu Key Operation Function SCPI Command (Continued) Transform Window Displays softkeys for selecting a type of the window. Selects the maximum type. :CALC{1-16}:TRAN:TIME KBES 13 Selects the normal type. :CALC{1-16}:TRAN:TIME KBES 6 Minimum Selects the minimum type. :CALC{1-16}:TRAN:TIME KBES 0 User Displays softkeys for setting up the user window shape. Impulse Width Sets the impulse width value.
Softkey Functions Analysis Menu Key Operation Function SCPI Command (Continued) Limit Test Limit Line Offset Marker -> Amplitude Offset Marker to limit offset. Centers the limit lines around the current marker position by using the limit amplitude offset function. Return Returns to the softkey display screen one level higher. Fail Sign Turns on/off the display of the limit test fail sign. Return Returns to the softkey display screen one level higher.
Softkey Functions Analysis Menu *1. Only with Options 313, 314, 413, and 414. *2. Only with Options 413 and 414. D.
Softkey Functions Average Menu Average Menu Key Operation Function SCPI Command Displays softkeys for setting averaging options. Averaging Restart Resets the counter and restarts from “1”. :SENS{1-16}:AVER:CLE Avg Factor Sets the averaging factor. The averaging factor must be defined as an integer from 1 to 999. :SENS{1-16}:AVER:COUN Averaging Enables or disables averaging execution. :SENS{1-16}:AVER Ave Trigger Enables or disables averaging trigger.
Softkey Functions Calibration Menu Calibration Menu Key Operation Function SCPI Command Displays softkeys for setting and executing calibrations. Correction Enables or disables error correction. Calibrate Displays softkeys for selecting calibration options. Select Port Displays softkeys for selecting options for response calibration using the OPEN standard. Displays softkeys for selecting a test port. 1 Selects test port 1. :SENS{1-16}:CORR:COLL:METH:OPEN 1 2 Selects test port 2.
Softkey Functions Calibration Menu Key Operation Function SCPI Command (Continued) Calibrate Response (Thru) Select Ports Displays softkeys for selecting options for response calibrations using the THRU standard. Displays softkeys for selecting test ports. 2-1 (S21) Selects the transmission test (measurement of S21) for test port 1→2. :SENS{1-16}:CORR:COLL:METH:THRU 2,1 3-1 (S31)*1 Selects the transmission test (measurement of S31) for test port 1→3.
Softkey Functions Calibration Menu Key Operation Function SCPI Command (Continued) Calibrate Enhanced Response Select Port :SENS{1-16}:CORR:COLL:METH:ERES 3,1 Selects the transmission test (measurement of S41) for test port 1→4. :SENS{1-16}:CORR:COLL:METH:ERES 4,1 1-2 (S12 S22) Selects the transmission test (measurement of S12) for test port :SENS{1-16}:CORR:COLL:METH:ERES 1,2 Selects the transmission test (measurement of S32) for test port 2→3.
Softkey Functions Calibration Menu Key Operation Function SCPI Command (Continued) Calibrate Displays softkeys for executing 1-port calibrations. 1-Port Cal Displays softkeys for selecting a test port. Select Port 1 Selects test port 1. :SENS{1-16}:CORR:COLL:METH:SOLT1 1 2 Selects test port 2. :SENS{1-16}:CORR:COLL:METH:SOLT1 2 *1 3 Selects test port 3. :SENS{1-16}:CORR:COLL:METH:SOLT1 3 4*2 Selects test port 4.
Softkey Functions Calibration Menu Key Operation Function SCPI Command (Continued) Calibrate 2-Port Cal Transmission Displays softkeys for executing transmission calibrations. Symbols x and y refer to the two test ports selected in the Select Ports menu (x and y are always 1 and 2, respectively, for models with Option 213 or 214). Port x-y Thru Executes a THRU calibration on test ports x and y. Isolation (Optional) Returns to the softkey display screen one level higher.
Softkey Functions Calibration Menu Key Operation Function SCPI Command (Continued) Calibrate 3-Port Cal*1 Transmission Port y-z Thru Executes a THRU calibration on test ports y and z. Return Isolation (Optional) :SENS{1-16}:CORR:COLL:THRU y,z :SENS{1-16}:CORR:COLL:THRU z,y Returns to the softkey display screen one level higher. Displays softkeys for executing isolation calibrations.
Softkey Functions Calibration Menu Key Operation Function SCPI Command (Continued) Calibrate 4-Port Cal*2 Transmission :SENS{1-16}:CORR:COLL:THRU 2,3 :SENS{1-16}:CORR:COLL:THRU 3,2 Port 2-4 Thru Executes a THRU calibration between test ports 2 and 4. :SENS{1-16}:CORR:COLL:THRU 2,4 :SENS{1-16}:CORR:COLL:THRU 4,2 Port 3-4 Thru Executes a THRU calibration between test ports 3 and 4.
Softkey Functions Calibration Menu Key Operation Function SCPI Command (Continued) Calibrate 2-Port TRL Cal Displays softkeys for executing THRU/LINE calibrations. Symbols x and y refer to the two test ports selected in the Select Ports menu (x and y are always 1 and 2, respectively, for models with Option 213 or 214). Thru/Line x-y Thru/Line Executes a THRU/LINE calibration on test ports x and y.
Softkey Functions Calibration Menu Key Operation Function SCPI Command (Continued) Calibrate 3-Port TRL Cal*1 Reflect Displays softkeys for executing reflection calibrations. Symbols x, y and z refer to the three test ports selected in the Select Ports menu (x, y, and z are always 1, 2, and 3, respectively, for models with Option 313 or 314). Executes a OPEN/SHORT reflection calibration on test port x.
Softkey Functions Calibration Menu Key Operation Function SCPI Command (Continued) Calibrate 4-Port TRL Cal*2 Reflect Displays softkeys for executing reflection calibrations. Port 1 Reflect Executes a OPEN/SHORT reflection calibration on test port 1. :SENS{1-16}:CORR:COLL:TRLR 1 Port 2 Reflect Executes a OPEN/SHORT reflection calibration on test port 2. :SENS{1-16}:CORR:COLL:TRLR 2 Port 3 Reflect Executes a OPEN/SHORT reflection calibration on test port 3.
Softkey Functions Calibration Menu Key Operation Function SCPI Command (Continued) Calibrate Adapter Removal Select Port Displays softkeys for executing adapter removal/insertion. Displays softkeys for selecting a test port. Selects test port 1. :SENS{1-16}:CORR:COLL:METH:ADAP:REM 1 2 Selects test port 2. :SENS{1-16}:CORR:COLL:METH:ADAP:REM 2 3*1 Selects test port 3. :SENS{1-16}:CORR:COLL:METH:ADAP:REM 3 4*2 Selects test port 4.
Softkey Functions Calibration Menu Key Operation Function SCPI Command (Continued) Displays softkeys for executing ECal (Electronic Calibrations). ECal :CALC{1-16}:SEL:FORM GDEL Displays softkeys for executing 1-port ECal 1-Port ECal Port 1 Executes a 1-port ECal on test port 1. :SENS{1-16}:CORR:COLL:ECAL:SOLT1 1 Port 2 Executes a 1-port ECal on test port 2. :SENS{1-16}:CORR:COLL:ECAL:SOLT1 2 Port 3*1 Executes a 1-port ECal on test port 3.
Softkey Functions Calibration Menu Key Operation Function SCPI Command (Continued) ECal Displays softkeys for executing an ENHANCED RESPONSE ECal. Enhanced Response Executes a THRU ECal for test port 1→2. :SENS{1-16}:CORR:COLL:ECAL:ERES 2,1 3-1 (S31 S11)*1 Executes a THRU ECal for test port 1→3. :SENS{1-16}:CORR:COLL:ECAL:ERES 3,1 4-1 (S41 S11)*2 Executes a THRU ECal for test port 1→4. :SENS{1-16}:CORR:COLL:ECAL:ERES 4,1 1-2 (S12 S22) Executes a THRU ECal for test port 2→1.
Softkey Functions Calibration Menu Key Operation Function SCPI Command (Continued) ECal Orientation Port 2 Selects port 2. Port B Selects port B as port 2. :SENS{1-16}:CORR:COLL:ECAL:PATH 2,2 Port C Selects port C as port 2. :SENS{1-16}:CORR:COLL:ECAL:PATH 2,3 Port D Selects port D as port 2. :SENS{1-16}:CORR:COLL:ECAL:PATH 2,4 Return Returns to the softkey display screen one level higher. Port 3 Selects port 3. Port A Selects port A as port 3.
Softkey Functions Calibration Menu Key Operation Function SCPI Command (Continued) Modify Cal Kit Define STDs 1. (Std Name) Displays softkeys for changing the definition of the calibration kit CALC{1-16}:SEL:FORM SCOM selected in the Cal Kit menu. Displays softkeys for defining the standard for a calibration kit. The label (Std Name) on each softkey represents the name of each standard. As a default setting, undefined standards are tagged with a No Name label. Up to 21 standards may be defined.
Softkey Functions Calibration Menu Key Operation Function SCPI Command (Continued) Modify Cal Kit Define STDs 1. (Std Name) Offset Z0 Sets the impedance Z0 between the measurement plane and the standard being defined. Normally, this value is set to the characteristic impedance of the system. Offset Loss :SENS{1-16}:CORR:COLL:CKIT:STAN1:LOSS Sets the offset loss for the standard. The offset loss is an energy loss due to skin effect on the length of single coaxial cable.
Softkey Functions Calibration Menu Key Operation Function SCPI Command (Continued) Modify Cal Kit Specify CLSs Displays softkeys for selecting standards for the OPEN standard class. Open Displays softkeys for selecting standards for the OPEN standard class that applies to Set All Does not include 1. (Std Name) in the OPEN standard class. :SENS{1-16}:CORR:COLL:CKIT:ORD:THRU x,y,0 1. (Std Name) Includes 1. (Std Name) in the OPEN standard class. :SENS{1-16}:CORR:COLL:CKIT:ORD:OPEN x,1 2.
Softkey Functions Calibration Menu Key Operation Function SCPI Command (Continued) Modify Cal Kit Specify CLSs Displays softkeys for setting the THRU standard class. Thru Displays softkeys for selecting standard for the THRU standard class that applies to all test ports. Set All None Does not include 1. (Std Name) in the THRU standard class. :SENS{1-16}:CORR:COLL:CKIT:ORD:THRU x,y,0 1. (Std Name) Includes 2. (Std Name) in the THRU standard class. :SENS{1-16}:CORR:COLL:CKIT:ORD:THRU x,y,1 . .
Softkey Functions Calibration Menu Key Operation Function SCPI Command (Continued) Modify Cal Kit Label Kit Allows the user to label the calibration kit. Restore Cal Kit Displays softkeys for initializing the definition of calibration kit. Restores the definition of the calibration kit selected by Cal Kit to factory default settings. Cancel Returns to the softkey display screen one level higher. :SENS{1-16}:CORR:COLL:CKIT:RES Export Cal Kit...
Softkey Functions Calibration Menu Key Operation Function SCPI Command (Continued) Port Extensions Auto Port Extension Adjust Mismatch Enables/disables the function to use the DC loss value for the :SENS{1-16}:CORR:EXT:AUTO:DCOF Include Enables/disables the loss value calculation. :SENS{1-16}:CORR:EXT:AUTO:LOSS Return Returns to the softkey display screen one level higher. calculation. Extensions Enables or disables the port extension function.
Softkey Functions Calibration Menu Key Operation Function SCPI Command (Continued) Power Calibration Select Port Displays the softkey to execute the power calibration. Displays the softkey to select the test port you want to calibrate. Selects test port 1. None 2 Selects test port 2. None 3 Selects test port 3. None 4 Selects test port 4. None Cancel Returns to the softkey display screen one level higher. Correction Toggles on/off the power level error correction.
Softkey Functions Calibration Menu Key Operation Function SCPI Command (Continued) Power Calibration Sensor B Settings Displays the softkey to execute the power calibration. Displays the softkey to set the calibration coefficient for the power sensor connected to channel B. Ref Cal Factor Sets the value of the reference calibration coefficient. :SOUR:POW:PORT:CORR:COLL:BSEN:RCF Delete Deletes the cursor line on the calibration coefficient table.
Softkey Functions Calibration Menu Key Operation Function SCPI Command (Continued) Mixer/Converter Calibration Displays the softkey to execute a scalar-mixer calibration. Scalar Cal (Manual) Displays the softkey to execute a scalar-mixer calibration by using the calibration kits. Select Ports Displays the softkey to select test ports and calibration methods you want to use. Select 1-port calibration for the test port 1.
Softkey Functions Calibration Menu Key Operation Function SCPI Command (Continued) Mixer/Converter Calibration Scalar Cal (Manual) Transmission Displays the softkey to execute THRU standard measurements. Thru x-y @ Freq x [Thru] Executes a THRU calibration on test ports x and y by using the frequency that is set in test port x. The x and y change by the selected test port and calibration method.
Softkey Functions Calibration Menu Key Operation Function SCPI Command (Continued) Mixer/Converter Calibration Scalar Cal (ECal) Select Ports Select a scalar-mixer calibration for test port 3→4 and 4→3 (both directions). Cancel Returns to the softkey display in one level higher. Power Meter None Displays the softkey to execute a power calibration. Use Sensor Selects the channel of the power sensor you want to use for the measurement of power calibration data.
Softkey Functions Display Menu Display Menu Key Operation Function SCPI Command Displays softkeys for setting up display options. Allocate Channels Displays softkeys for setting the number of channels to be displayed and the channel window arrangement. The execution of measurements does not depend on the display status of each channel (measurements can be performed on channels that are not displayed).
Softkey Functions Display Menu Key Operation Function SCPI Command (Continued) Allocate Channels :DISP:SPL D12_34_56_78 Displays channel windows by splitting the screen vertically into four equal parts and then bisecting each window horizontally, with channels 1, 3, 5, and 7 displayed in left windows from top to bottom, and channels 2, 4, 6, and 8 displayed in right windows from top to bottom. D.
Softkey Functions Display Menu Key Operation Function SCPI Command (Continued) Allocate Traces :DISP:WIND{1-16}:SPL D12_33 Displays graphs by bisecting the screen vertically and then bisecting the upper window horizontally with graphs 1, 2, and 3 displayed, respectively, in the upper-left corner, upper-right corner, and bottom.
Softkey Functions Display Menu Key Operation Function SCPI Command (Continued) Display Selects both data trace and memory trace for on-screen display. By storing data obtained under certain conditions, it is possible to compare them to new measurement results (data trace) obtained under different conditions. To store data for a memory trace, go back to the previous menu and press Data Æ Mem.
Softkey Functions Format Menu Format Menu Key Operation Function SCPI Command Displays softkeys for setting up data formats. Log Mag Displays traces in a rectangular display format with log magnitude (dB) on :CALC{1-16}:SEL:FORM MLOG the Y-axis and frequencies on the X-axis (log magnitude format). Phase Displays traces in a rectangular display format with phase (-180× to +180×) :CALC{1-16}:SEL:FORM PHAS on the Y-axis and frequencies on the X-axis (phase format).
Softkey Functions Macro Setup Menu Macro Setup Menu Key Operation Function SCPI Command Displays the macro setup menu. Starts the VBA editor. A keyboard and mouse are necessary to use this editor. None New Project Creates a new VBA project. None Load Project Opens a dialog box for loading a saved VBA project. :MMEM:LOAD:PROG Load & Run Displays programs (VBA projects) stored under D:\VBA as a list of softkeys.
Softkey Functions Macro Setup Menu Key Operation Function SCPI Command (Continued) User Menu Button 10 Executes the procedure assigned to Button 10. Softkey label can be modified using a command. Return Returns to the softkey display screen one level higher. Preset User Menu Returns the softkey labels of user menu to the initial settings. Return Returns to the “E5070B/E5071B Menu (Top Menu)” on page 650.
Softkey Functions Marker Menu Marker Menu Key Operation Function SCPI Command Activates marker 1 and displays an input dialog box for setting the stimulus value for marker 1. Also displays softkeys for setting and moving each marker. Enables marker 1 if it is disabled. Also activates marker 1 and displays an input dialog box for setting the stimulus value. :CALC{1-16}:MARK1 :CALC{1-16}:MARK1:ACT :CALC{1-16}:MARK1:X Marker 2 Enables marker 2 if it is disabled.
Softkey Functions Marker Menu Key Operation Function SCPI Command (Continued) Ref Marker Mode :CALC{1-16}:MARK{1-10}:REF Enables or disables the reference marker mode. When enabled, stimulus values and response values of markers 1 to 9 will be displayed using values relative to the reference marker. When disabled, the reference marker will not appear on the screen. Return Returns to the “E5070B/E5071B Menu (Top Menu)” on page 650.
Softkey Functions Marker Function Menu Marker Function Menu Key Operation SCPI Command Function Displays softkeys for setting the sweep range using markers and other marker options. Sets the starting frequency to the stimulus value of the active marker on the :CALC{1-16}:MARK{1-10}:SET STAR active trace. Even if the reference marker is enabled and a relative stimulus value is displayed, the absolute value will be used.
Softkey Functions Marker Search Menu Marker Search Menu Key Operation Function SCPI Command Displays softkeys for performing searches using markers. Max Moves the active marker to a position on the trace that represents :CALC{1-16}:MARK{1-10}:FUNC:EXEC :CALC{1-16}:MARK{1-10}:FUNC:TYPE MAX the maximum response value. Min Moves the active marker to a position on the trace that represents :CALC{1-16}:MARK{1-10}:FUNC:EXEC :CALC{1-16}:MARK{1-10}:FUNC:TYPE MIN the minimum response value.
Softkey Functions Marker Search Menu Key Operation Function SCPI Command (Continued) Target Displays softkeys for setting and performing target searches. A target refers to a point on a trace that has a unique response value (target value) in a rectangular display format. A target search picks up points that have matching characteristics defined by Target Value and Target Transition.
Softkey Functions Marker Search Menu Key Operation Function SCPI Command (Continued) Bandwidth :CALC{1-16}:MARK:BWID Enables or disables bandwidth searching. When enabled, bandwidth parameters (Insertion loss, Low cutoff point, High cutoff point, Center frequency, Bandwidth and Q) will be displayed on the screen.If the Smith chart format or polar format is used, the main response value, among the two response values, will be used to perform searches (e.g., resistance in the Smith (R+jX) format).
Softkey Functions Measurement Menu Measurement Menu Key Operation Function SCPI Command Displays softkeys for setting measurement parameters. These softkeys will not be displayed unless either the balanced/unbalanced conversion function or the fixture simulator function is turned off (BalUn OFF or Fixture Simulator OFF) in the “Analysis Menu” on page 651 on the E5070B/E5071B. Selects parameter S11. :CALC{1-16}:PAR{1-16}:DEF S11 S21 Selects parameter S21.
Softkey Functions Measurement Menu *1.Only with Options 313, 314, 413, and 414. *2.Only with Options 413 and 414. *3.The numeric in parentheses indicates the stimulus port number.
Softkey Functions Measurement Menu (Balance Measurement, SE-Bal) Measurement Menu (Balance Measurement, SE-Bal) Key Operation Function SCPI Command Displays softkeys for setting measurement parameters (only for models with Option 313, 314, 413, or 414).
Softkey Functions Measurement Menu (Balanced Measurement, Bal-Bal) Measurement Menu (Balanced Measurement, Bal-Bal) Key Operation Function SCPI Command Displays softkeys for setting up measurement parameters (only for models with Option 412 or 414).
Softkey Functions Measurement Menu (Balanced Measurement, Bal-Bal) Key Operation Function SCPI Command (Continued) Selects parameter Scc22. Scc22 defines the way a common mode signal input to (balanced) port 2 on the DUT is reflected as a common mode signal. :CALC{1-16}:FSIM:BAL:PAR{1-16}: BBAL SCC22 Imbalance1 Selects parameter Imbalance1. :CALC{1-16}:FSIM:BAL:PAR{1-16}:BBAL IMB1 Imbalance2 Selects parameter Imbalance2.
Softkey Functions Measurement Menu (Balanced Measurement, SE-SE-Bal) Measurement Menu (Balanced Measurement, SE-SE-Bal) Key Operation Function SCPI Command Displays softkeys for setting measurement parameters (only for models with Option 413 or 414).
Softkey Functions Measurement Menu (Balanced Measurement, SE-SE-Bal) Key Operation Function SCPI Command (Continued) Selects parameter Scc33. Scc33 defines the way a common mode signal input to (balanced) port 3 on the DUT is reflected as a common mode signal. :CALC{1-16}:FSIM:BAL:PAR{1-16}:SSB SCC33 Imbalance1 Selects parameter Imbalance1. :CALC{1-16}:FSIM:BAL:PAR{1-16}:SSB IMB1 Imbalance2 Selects parameter Imbalance2.
Softkey Functions Preset Menu Preset Menu Key Operation Function SCPI Command Displays softkeys for restoring the preset conditions. OK Restores the preset conditions. Cancel Returns to the “E5070B/E5071B Menu (Top Menu)” on page 650.
Softkey Functions Save/Recall Menu Save/Recall Menu Key Operation Function SCPI Command Displays softkeys for saving and recalling data. Save State Displays softkeys for saving settings. Saves the current settings on the internal hard disk drive (D:) and names the saved :MMEM:STOR "State01.sta" file State01.sta. State02 Saves the current settings on the internal hard disk drive (D:) and names the saved :MMEM:STOR "State02.sta" file State02.sta.
Softkey Functions Save/Recall Menu Key Operation Function SCPI Command (Continued) Recall by File Name Displays softkeys that show the state file names for recalling settings. aaa Recalls from the internal hard disk drive (D:\State) the settings saved as the filename used in the softkey label. :MMEM:LOAD bbb Recalls from the internal hard disk drive (D:\State) the settings saved as the filename used in the softkey label.
Softkey Functions Save/Recall Menu Key Operation Function SCPI Command (Continued) Save SnP SnP Format Displays softkeys to save measurement data into Touchstone format file. Displays softkeys to select data format of a file in which data is saved. Selects data format for the active channel’s display format. If the display format does not correspond to data format of the Touchstone file, “real number imaginary number format “ is selected.
Softkey Functions Save/Recall Menu Key Operation Function SCPI Command SnP s4p 1-2-3-4 Cancel Displays softkeys to save data to a Touchstone file with s4p file type. Saves measurement data of test port 1, 2, 3, and 4 to a Touchstone file. :MMEM:STOR:SNP:TYPE:S4 P 1,2,3,4 :MMEM:STOR:SNP Returns to the “E5070B/E5071B Menu (Top Menu)” on page 650.
Softkey Functions Scale Menu Scale Menu Key Operation Function SCPI Command Displays softkeys for adjusting scales. Auto Scale Automatically adjusts scales for the active trace. :DISP:WIND{1-16}:TRAC{1-16}:Y:AU TO None Automatically adjusts scales for all traces within the active channel. :DISP:WIND{1-16}:TRAC{1-16}:Y:DIV Defines the number of divisions on the Y-axis of a rectangular display format. An even number from 4 to 30 must be used.
Softkey Functions Stimulus Menu Stimulus Menu Key Operation Function SCPI Command Sets the lowest frequency for sweeps. :SENS{1-16}:FREQ:STAR Also displays a menu (Stimulus Menu) for defining the sweep range. Start Sets the starting frequency for sweeps. :SENS{1-16}:FREQ:STAR Stop Sets the ending frequency for sweeps. :SENS{1-16}:FREQ:STOP Center Sets the center frequency of the sweep range. :SENS{1-16}:FREQ:CENT Span Sets the frequency span for sweeps.
Softkey Functions Sweep Setup Menu Sweep Setup Menu Key Operation Function SCPI Command Displays softkeys for setting up sweeps. Displays the menu to set the stimulus signal output. Power Sets the output power level of the internal signal source of the analyzer. Power :SOUR{1-16}:POW Displays softkeys for selecting the power range. -20 to 10 Sets the power range to -20 dBm to 10 dBm. :SOUR{1-16}:POW:ATT 0 -25 to 7 Sets the power range to -25 dBm to 7 dBm.
Softkey Functions Sweep Setup Menu Key Operation Function SCPI Command (Continued) Sweep Type Cancel Edit Segment Table Returns to the softkey display screen one level higher. Displays the segment sweep setup table as well as softkeys for editing the segment table. Freq Mode Alternates the setup mode for the sweep range between two methods: one using the starting and ending frequencies (Start/Stop), and the other using the center frequency and a frequency span (Center/Span).
Softkey Functions Sweep Setup Menu Key Operation Function SCPI Command (Continued) Frequency Offset External Source Set the offset for the base frequency*2. :SENS{1-16}:OFFS:LOC:OFFS Start Set the start frequency. :SENS{1-16}:OFFS:LOC:STAR Stop Set the stop frequency. :SENS{1-16}:OFFS:LOC:STOP Control Turns on/off of the external signal source control with GPIB. :SENS{1-16}:OFFS:LOC:CONT Power Set the power level. :SENS{1-16}:OFFS:LOC:POW Slope [xdB/GHz] Set the power slope.
Softkey Functions System Menu System Menu Key Operation Function SCPI Command Displays softkeys for performing limit tests and accessing control and management functions on the analyzer. Print Outputs the current screen to a printer. :HCOP Abort Printing Aborts printing. :HCOP:ABOR Printer Setup Opens a dialog box for setting up the printer. None Invert Image Inverts the colors of the screen display.
Softkey Functions System Menu Key Operation Multiport Test Set Setup Function SCPI Command Displays softkeys for setting the E5091A control. Displays the model selected for test set 1 (ID=1) Test Set 1 Displays softkeys for selecting a test port of the E5091A to which you want to connect port 3. Port 3 Selects test port R1+. :SENS{1-16}:MULT1:PORT3 R1 R2+ Selects test port R2+. :SENS{1-16}:MULT1:PORT3 R2 Selects test port R3+. Selects test port R2+. Selects test port R4+.
Softkey Functions System Menu Key Operation Function SCPI Command Displays softkeys for setting up the beeper function, GPIB, Network, internal clock, key lock function and color of display image. Misc Setup Displays softkeys for setting up the beeper function. Beeper Beep Complete Enables or disables beeps at the end of processes. When enabled, the user will be notified with a beep when a measurement has completed or settings have been saved.
Softkey Functions System Menu Key Operation Function SCPI Command (Continued) Misc Setup Displays softkeys for configuring network settings. Network Setup Telnet Server Enables or disables the telnet server function. SICL-LAN Server Enables or disables the SICL-LAN server function. None None Sets the address for controlling the analyzer from a controller via SICL-LAN. None Web Server Enables or disables the Web server function. None VNC Server Configuration...
Softkey Functions System Menu Key Operation Function SCPI Command (Continued) Misc Setup Color Setup Normal Mem Trace 1 Displays softkeys for setting up the color of the memory trace of Trace :DISP:COL{1-2}:TRAC1:MEM 1. The lower-level softkeys are the same as those for Data Trace 1. : : : Mem Trace 9 Displays softkeys for setting up the color of the memory trace of Trace :DISP:COL{1-2}:TRAC9:MEM 9. The lower-level softkeys are the same as those for Data Trace 1.
Softkey Functions System Menu Key Operation Function SCPI Command (Continued) Misc Setup Control Panel... Opens a control panel window. Return Returns to the softkey display screen one level higher. None Turns the backlight for the LCD screen on/off. :SYST:BACK Firmware Revision Displays the firmware revision information in a dialog box. *IDN? Service Menu Displays softkeys for maintenance services. Test Menu Displays softkeys for self diagnosis functions.
Softkey Functions System Menu Key Operation Function SCPI Command (Continued) Service Menu Displays softkeys for maintenance services. Channel/Trace Setup 16 Ch / 16 Tr 801 Sets the maximum number of channels to 16, the maximum number of None Points traces to 16, and the upper limit of the measurement point to 801. 24 Ch / 9 Tr 801 Points Sets the maximum number of channels to 24, the maximum number of None traces to 9, and the upper limit of the measurement point to 801.
Softkey Functions Trigger Menu Trigger Menu Key Operation Function SCPI Command Displays following softkeys for setting the trigger. Once the trigger mode is set, measurements are executed according to the trigger mode even when the channel is no longer on display due to reduction of the number of channels to be displayed from the “Display Menu” on page 694. Sets the active channel trigger mode to “hold”. A trigger sent from the trigger source to that channel will not prompt a sweep.
Softkey Functions Trigger Menu 730 Appendix D
Softkey Functions Trigger Menu D.
Softkey Functions Trigger Menu 732 Appendix D
E. General Principles of Operation E General Principles of Operation This chapter explains the general principles of operation for the Agilent E5070B/E5071B.
General Principles of Operation System Description System Description A network analyzer supplies a sweep signal to a DUT, measures its transmission and reflection, and displays the results as ratios against the input signal from the signal source. The E5070B/E5071B network analyzer consists of the circuit modules shown in Figure E-1. Figure E-1 System Diagram for the E5070B/E5071B Network Analyzer Synthesized Source The synthesized source generates an RF sweep signal in the following frequency range.
General Principles of Operation System Description Source Switcher The source switcher is used to switch test ports to which the RF signal is supplied from the source. Signal Separator The signal separator consists of directivity couplers that detect input and output signals at the test ports. On a test port to which a signal is output, the output signal and the reflection from the DUT are detected as the reference signal (R) and the test signal (T), respectively.
General Principles of Operation Data Processing Data Processing The internal data processing flowchart for the E5070B/E5071B is shown in Figure E-2. Figure E-2 Data Processing Flowchart ADC The ADCs (analog-to-digital converters) convert analog signals fed to the receiver and converted into IF signals (R1, R2, ⋅⋅⋅, Rn and T1, T2, ⋅⋅⋅, Tn) into digital signals. One ADC is available for each signal and the conversion takes place simultaneously.
General Principles of Operation Data Processing IF Range Correction Input signals that went through ranging at the receiver are reverted (corrected) to previous values before the ranging. Ratio Calculation The ratio between two signals is determined by performing divisions on complex numbers. In the case of absolute measurements (Option 008), the ratio of complex number can not be calculated.
General Principles of Operation Data Processing Corrected Data Array Unlike the raw data array, this array stores the results obtained after error corrections, port extensions, or the fixture simulator functions are applied. The user is allowed to read/write data from/to the corrected data array. Corrected Memory Array By pressing - Data → Mem, the contents of the corrected data array will be copied to this array. The user is allowed to read/write data from/to the corrected memory array.
General Principles of Operation Data Processing Display The results obtained after data processing are displayed on the screen as traces. E.
General Principles of Operation Data Processing 740 Appendix E
F. Replacing the 8753ES with the E5070B/E5071B F Replacing the 8753ES with the E5070B/E5071B This chapter describes the information necessary to replace Agilent 8753ES with the Agilent E5070B/E5071B.
Replacing the 8753ES with the E5070B/E5071B Important Functional Differences Important Functional Differences This section describes the key differences between the Agilent 8753ES and Agilent E5070B/E5071B. Channel and Trace Concepts In the 8753ES, channels 1 and 2 are independent from each other and have auxiliary channels, channels 3 and 4, respectively. Channels 3 and 4 can be displayed as additions to channels 1 and 2, respectively.
Replacing the 8753ES with the E5070B/E5071B Important Functional Differences Table F-2 Test Port Output Power Levels Function 8753ES E5070B/E5071B Output power levels −85 dBm to 10 dBm (std.) −15 dBm to 0 dBm (Options 213, 313, and 413) −85 dBm to 8 dBm (Options 014 and 075) Output power ranges −15 dBm to 10 dBm (std.
Replacing the 8753ES with the E5070B/E5071B Important Functional Differences Calibration The types of calibration kits supported by the 8753ES and E5070B/E5071B are shown in Table F-3. Table F-3 Supported Calibration Kits Type of calibration kit 8753ES E5070B/E5071B 7 mm 85031B 85031B,85050C/D 3.5 mm 85033C/D/E 85033D/E, 85052D N type 50 Ω: 85032B/E/F 75 Ω: 85036B/E 50 Ω: 85032B/F 75 Ω: 85036B/E 2.4 mm 85056/D N/A TRL 3.
Replacing the 8753ES with the E5070B/E5071B Important Functional Differences Trigger System The trigger system detects the signal for starting a measurement (trigger) and controls decisions on whether to measure or not measure. On the 8753ES, the trigger state is available for the pair of a main channel and an auxiliary channel (two pairs: channels 1 and 3 and channels 2 and 4). For each pair, three states are available: Hold, Waiting for Trigger, and Measurement.
Replacing the 8753ES with the E5070B/E5071B Important Functional Differences Data Flow The data flow in the 8753ES is shown in Figure F-1 while the flow in the E5070B/E5071B is shown in Figure F-2. As described in “Reading/Writing Data” on page 747, the types of data you can read/write using the 8753ES differ from those you can read/write using the E5070B/E5071B.
Replacing the 8753ES with the E5070B/E5071B Important Functional Differences Figure F-2 E5070B/E5071B Data Flow Reading/Writing Data Types of data that can be handled by the 8753ES and E5070B/E5071B are listed in Table 3-4.
Replacing the 8753ES with the E5070B/E5071B Important Functional Differences Table F-4 Reading/Writing Data Function 8753ES Reading data Pre-raw data (in Take4 mode) Calibration coefficient array (after interpolation) E5070B/E5071B Power meter calibration coefficient array (after interpolation) Entry area display All lists in list format Screen Display and Marker Functions The 8753ES allows up to four channels to be displayed on the screen. Up to five markers can be displayed on each channel.
Replacing the 8753ES with the E5070B/E5071B Important Functional Differences along with the minimum/maximum measurement value for each segment as retrieved items in test results. On the other hand, the E5070B/E5071B supports pass/fail results, measurement value, and upper/lower limit values for each measurement point as well as pass/fail results of the active trace for each channel. Analytical Functions Although the 8753ES does not support the fixture simulator function, the E5070B/E5071B does.
Replacing the 8753ES with the E5070B/E5071B Important Functional Differences Save/Recall For storing data, the 8753ES is provided with an internal register, internal disk drive (floppy disk), and external disk drive (connected through the GPIB). In contrast, the E5070B/E5071B provides an internal hard disk drive, and an internal disk drive (floppy disk).
Replacing the 8753ES with the E5070B/E5071B Comparing Functions Comparing Functions Table F-5 Functions of the 8753ES vs. the E5070B/E5071B Function 8753ES E5070B/E5071B Measurement Reset Can be executed by using the front panel and the GPIB command. Can be executed by using the front panel, the GPIB command, and telnet.
Replacing the 8753ES with the E5070B/E5071B Comparing Functions Table F-5 Functions of the 8753ES vs. the E5070B/E5071B Function 8753ES Measurement Sweep (cont'd.
Replacing the 8753ES with the E5070B/E5071B Comparing Functions Table F-5 Functions of the 8753ES vs. the E5070B/E5071B Function 8753ES E5070B/E5071B Measurement Trigger source (cont'd.
Replacing the 8753ES with the E5070B/E5071B Comparing Functions Table F-5 Functions of the 8753ES vs. the E5070B/E5071B Function Calibration 8753ES Calibration kit Available calibration kits: 7 mm: 85031B 3.5 mm: 85033C/D/E N type (50 Ω): 85032B/E/F N type (75 Ω): 85036B/E 2.4 mm: 85056A/D TRL 3.5 mm: 85052C User defined calibration kit Calibration type E5070B/E5071B Not more than two ports 7 mm: 85031B, 85050C/D 3.
Replacing the 8753ES with the E5070B/E5071B Comparing Functions Table F-5 Functions of the 8753ES vs. the E5070B/E5071B Function Calibration (cont'd.
Replacing the 8753ES with the E5070B/E5071B Comparing Functions Table F-5 Functions of the 8753ES vs.
Replacing the 8753ES with the E5070B/E5071B Comparing Functions Table F-5 Functions of the 8753ES vs. the E5070B/E5071B Function Markers (cont'd.
Replacing the 8753ES with the E5070B/E5071B Comparing Functions Table F-5 Functions of the 8753ES vs.
Replacing the 8753ES with the E5070B/E5071B Comparing Functions Table F-5 Functions of the 8753ES vs. the E5070B/E5071B Function Macros System 8753ES E5070B/E5071B Creating macros Uses a test sequence. Uses VBA. Automatic execution Will execute if “AUTO” is given as the name of the sequence 6. Only auto loading is possible. Self-test Can be executed by using the front panel and SCPI commands. Can be executed by using the front panel.
Replacing the 8753ES with the E5070B/E5071B Comparing Functions 760 Appendix F
Index Symbols .bmp Saving the screen image, 368 .csv Saving/Recalling files, 362 CSV file Saving trace data, 362 .
Index Confidence check, 179 Configuring ripple limit, 387 Connector Repeatability Errors, 94 Connectors Maintenance, 584 Conversion, 277 Conversion Loss Frequency-offset measurement, 325 Conversion Loss Measurement, 325, 329, 330 Corrected Data Array Data Processing, 738 Corrected Memory Array Data Processing, 738 Corrected System Performance specification, 539 CSV file Limit table, 381 D D*M(Trace status), 48 D+M(Trace status), 48 D/M(Trace status), 48 Data Entry Bar, 40 Data Format Data Processing, 738 s
Index external trigger delay time, 247 External Trigger Input Connector, 55 Index G Gat(Trace status), 48 Gating, 347 General Information specification, 551 GPIB setting, 428 GPIB address Power meter, 319, 323 GPIB address setting External signal source, 319 GP-IB Cables Accessories, 578 GPIB Connector, 55 Graph layout, 65 Graticule Labels turning off the display, 88 Group Delay Data Processing, 738 H Handler I/O Port, 55 Harmonic distortion, 328 Harmonic distortion measurement, 330 Harmonics measurement
Index K Keyboard Locking, 448 Keyboard Port, 57 L Layout of channel windows, 62 LCD Screen, 31 LCD Screen Backlight Turning off, 450 Limit Line Defining, 378 Limit test Concept, 376 Displaying judgement result, 377 Limit line ON/OFF, 382 Limit Line Offset, 383 Amplitude Offset, 383 Marker Amp. Ofs.
Index Definitions, 538 Number of channels, 62 Number of Points settings, 78 O Offset Data Processing, 738 Open dialog box, 359 OPEN/SHORT Response Calibration, 107 procedure, 107 Operational Manual, 7 Option, 451 Order Base Display, 419 Overview of Frequency Offset Measurement, 314 Index Q Quick Start Guide, 7 R Random Errors, 94 Ratio Calculation Data Processing, 737 Raw Data Array Data Processing, 737 Rear Panel, 54 Rear Panel Information specification, 552 Recalling a file Compatibility, 354 Receiver
Index symbols, 4 safety summary, 3 Sample Program, 8 Save Limit Table, 389 Save as dialog box, 358 Save/Recall Menu Softkey Functions, 715 SaveToTouchstone.vba, 363 SaveTouchstone data file Trace data SaveToTouchstone.
Index Systematic Errors, 95 Index U Unbalanced and Balanced Bandpass Filter Measuring Examples, 513 unbalanced port function, 285 Uncorrected System Performance specification, 544 Unknown Thru Calibration, 130, 153 Update Off, 416 USB (USBTMC) Interface Port, 58 USB ID, 451 USB Mass Storage Device, 447 USB Port, 39, 57 user characterized ECal, 171 V Vector Mixer Calibration, 223 Vector-Mixer Calibration, 223 Balanced mixer characterization, 230 Measured mixer, 224 Mixer characterization, 225 Overview, 22
Index W Warning Message, 630 Web Server, 443 Window labeling, 89 Window Display maximizing, 88 Window Displays settings, 88 X X-Axis Frequency data, 323 Y Yr(Trace status), 48 Yt(Trace status), 48 Z Z0, 69 Zr(Trace status), 48 Zt(Trace status), 48 768 Index
REGIONAL SALES AND SUPPORT OFFICES For more information about Agilent Technologies test and measurement products, applications, services, and for a current sales office listing, visit our web site: http://www.agilent.com/find/tmdir. You can also contact one of the following centers and ask for a test and measurement sales representative.