User`s guide
1- 80 Agilent EasyEXPERT User’s Guide Vol. 1, Edition 1
Main GUI
Configuration
Select this radio button and specify the compensation data file to perform the
capacitance compensation for the path from the B2200A/B2201A input
ports to the end of positioner/manipulator/probe card. See Agilent B1500A
User’s Guide for creating the compensation data file. The data after
compensation is displayed and saved.
• E5250A (option)
The following setup is available for the E5250A hardware profile.
• User Compensation Data File
Select this radio button and specify the compensation data file to perform the
capacitance compensation for the path from the E5250A input ports to the
end of positioner/manipulator/probe card. See Agilent B1500A User’s Guide
for creating the compensation data file. The data after compensation is
displayed and saved.
Clear Clears the User Compensation Data File field.
Triaxial Cable Enter the length of the 16494A triaxial cable
connected between the E5250A outputs to the
connector plate.
User Triaxial Cable Enter the length of the triaxial cable between
the connector plate and the DUT. Enter 0 if
the triaxial cable is not used.
User Coaxial Cable Enter the length of the coaxial cable between
the connector plate and the DUT. Enter 0 if
the coaxial cable is not used.