1- 78 Agilent EasyEXPERT User’s Guide Vol. 1, Edition 1
Main GUI
Configuration
Status Shows one of the following status.
Not Detected N1268A has not been detected yet.
Communication Established Normal status. N1268A can be used.
Communication Lost Once the N1268A was detected. However the
communication has been broken now.
Diagnosis Effective when the Enable Ultra High Voltage Unit check box is checked.
The Start button starts the diagnosis for the N1268A. Dialog box will be opened to
navigate you the cable connection during the diagnosis. The Status shows the
execution result.
Apply button This button applies the setting.
SMU Output Setting Limits
Only for the B1505A. This screen is used to set the upper limit of the SMU output
value and compliance value.
Voltage Setting
Limit
Sets the upper limit of the voltage output value and compliance value.
Effective
value: 200 V to 10 kV, in 100 V step.
Current Setting
Limit
Sets the upper limit of the current output value and compliance value.
Effective
value: 1 A to 1500 A, in 1 A step.
Apply This button applies the setting.
Event Log
The event log tab screen displays the event log.
Date/Time Displays date and time of the event.
Message Event message
ID Process ID of the event
Process Process name
User User name. Windows login name.