User`s guide

1- 76 Agilent EasyEXPERT Users Guide Vol. 1, Edition 1
Main GUI
Configuration
Current Control
SMU
Specifies the module (MCSMU or HCSMU) connected to the I
Control port of the N1266A.
Self Test Effective only when the Enable High Voltage Medium Current Unit check box is
checked.
The Start button starts the self-test for the N1266A. Measurement terminals must be
opened during the self-test. The Status shows the execution result.
Output Enable Series Resistor (100 kohm) for HVSMU
This check box is used to connect/disconnect the built-in series resistor (100 k) on
the HVSMU direct path. If this box is checked, the series resistor is connected.
Default Specifies the input-to-output connection status (HVSMU,
HVMCU, or blank) in the normal state. The N1266A makes the
path to the output port from the input port specified by the
Default field. If this field is a blank, the High output port is
open.
When a measurement is performed, the N1266A performs
automatic switching in every test. The N1266A makes the path
to the output port from the measurement resource used for the
test.
The connection status is indicated by the Status indicator
mounted on the N1266A. See
Table 1-5.
Auto Detection This check box is used to enable/disable the N1266A automatic detection. If this
box is checked, the automatic detection is performed when EasyEXPERT is
launched.
Status Shows one of the following status.
Not Detected N1266A has not been detected yet.
Communication Established Normal status. N1266A can be used.
Communication Lost Once the N1266A was detected. However the
communication has been broken now.
Apply button This button applies the setting.