Agilent EasyEXPERT User’s Guide Vol. 1, Edition 1 1- 75
Main GUI
Configuration
Status Shows one of the following status.
Not Detected N1265A has not been detected yet.
Communication Established Normal status. N1265A can be used.
Communication Lost Once the N1265A was detected. However the
communication has been broken now.
Diagnosis Effective when the Enable Ultra High Current Unit check box, the Enable Selector
check box, or/and the Enable Gate Control check box is checked.
The Start button starts the diagnosis for the N1265A. Dialog box will be opened to
navigate you the cable connection during the diagnosis. The Status shows the
execution result.
Apply button This button applies the setting.
NOTE
Before starting diagnosis
If the N1266A expander is connected between the HVSMU and the N1265A,
disconnect the N1266A. And connect the cable from the HVSMU Force terminal to
the N1265A Selector Input HVSMU terminal.
HVSMU Current Expander
Only for the B1505A. This tab screen is used to set the N1266A HVSMU current
expander. The N1266A can be controlled by the B1505A via the Digital I/O
connector.
Enable High
Voltage Medium
Current Unit
This check box is used to enable/disable the high voltage medium current unit
(HVMCU). If this box is checked, HVMCU can be used.
Input/Control Specifies the HVSMU and the modules used for controlling the HVMCU. For the
port which no module is connected, blank the entry field.
HVSMU Specifies the HVSMU connected to the HVSMU input port of
the N1266A.
Voltage Control
SMU
Specifies the module (MCSMU or HCSMU) connected to the V
Control port of the N1266A. The SMU channel set to this field
is assigned to the HVMCU and is used to indicate the HVMCU.