Agilent EasyEXPERT User’s Guide Vol. 1, Edition 1 1- 73
Main GUI
Configuration
HVSMU Specifies the HVSMU connected to the HVSMU port of the
N1265A directly or via the N1266A current expander.
SMU Specifies the module (MPSMU or HPSMU) connected to the
SMU port of the N1265A.
For the port which no module is connected, blank the entry field.
Output Enable Series Resistor (100 kohm) for HVSMU
This check box is used to connect/disconnect the built-in series resistor (100 k) on
the HVSMU direct path. If this box is checked, the series resistor is connected.
Default Specifies the selector input-to-output connection status (UHCU,
HVSMU, SMU, GNDU, or blank) in the normal state. The
selector makes the path to the output port from the input port
specified by the Default field. If this field is blank, the output
port is open.
When a measurement is performed, the selector performs
automatic switching in every test. The selector makes the path
to the output port from the measurement resource used for the
test. The connection status is indicated by the Status indicator
mounted on the N1265A. See
Table 1-4.
Enable Gate
Control
This check box is used to enable/disable the Gate path of the N1265A. If this box is
checked, the Gate path can be used.
Gate Control SMU Specifies the module (MCSMU or HCSMU) connected to the
Gate port of the N1265A. For the port which no module is
connected, blank the entry field. Built-in series resistor can be
connected to the Gate path. To connect the series resistor, set
the Resistor field.
Resistor Specifies the resistance of the built-in series resistors for the
Gate path. If the series resistor is not used, select 0 ohm.