1- 72 Agilent EasyEXPERT User’s Guide Vol. 1, Edition 1
Main GUI
Configuration
Slave HCSMU Specifies the HCSMU module used as the slave channel of the dual HCSMU
operation. The module must be connected to the Slave port of the 16493S-020/021
dual HCSMU adapter.
Apply This button applies the dual HCSMU operating condition. To apply the setup
changes successfully, the B1505A must be in the idle state and the Standby output
must be OFF.
UHC Expander / Fixture
Only for the B1505A. This tab screen is used to set the N1265A ultra high current
expander/fixture. The N1265A can be controlled by the B1505A via the Digital I/O
connector.
Enable Ultra High
Current Unit
This check box is used to enable/disable the ultra high current unit (UHCU). If this
box is checked, UHCU can be used.
Control Specifies the modules used for controlling the UHCU. For the port which no module
is connected, blank the entry field.
Voltage Control
SMU
Specifies the module (MCSMU or HCSMU) connected to the V
Control port of the N1265A. The SMU channel set to this field
is assigned to the UHCU and is used to indicate the UHCU.
Current Control
SMU
Specifies the module (MCSMU or HCSMU) connected to the I
Control port of the N1265A.
Self Test Effective only when the Enable Ultra High Current Unit check box is checked.
The Start button starts the self-test for the N1265A. Measurement terminals must be
opened during the self-test. The Status shows the execution result.
Self Calibration Effective only when the Enable Ultra High Current Unit check box is checked.
The Start button starts the self-calibration for the N1265A. Measurement terminals
must be opened during the self-test. The Status shows the execution result.
Enable Selector This check box is used to enable/disable the selector of the N1265A. If this box is
checked, the selector can be used.
Input UHCU Displays the SMU channel assigned to the UHCU.