User`s guide

Agilent EasyEXPERT User’s Guide Vol. 1, Edition 1 1- 71
Main GUI
Configuration
Table 1-3 Module Selector Status Indicator and Input-to-Output Connection Path
CAUTION If dual HCSMU (DHCSMU) is used with the module selector, the maximum current
must be ± 30
A to prevent the module selector from performance degradation and
failure.
Dual HCSMU Combination
Only for the B1505A installed with two HCSMU modules. This screen is used to set
the dual HCSMU operation.
Enable This check box is used to enable/disable the dual HCSMU operation. If this box is
checked, the dual HCSMU operation is enabled and two HCSMU modules can be a
SMU which supports ±
40 A (pulse), ± 2 A (DC).
Master HCSMU Specifies the HCSMU module used as the master channel of the dual HCSMU
operation. The module must be connected to the Master port of the 16493S-020/021
dual HCSMU adapter.
Module used for
the test
Status
indicator
Input-to-output connection
a
of output terminals
a. In the normal state, the module selector makes the path to the output port from the input port
specified by the Default field. When a measurement is performed, the module selector performs
automatic switching in every test. The selector makes the path to the output port from the mod
-
ule used for the test.
High Sense High Force Low Sense Low Force
HPSMU or
MPSMU
HPSMU HP/MPSMU
Sense
HP/MPSMU
Force
HCSMU
Low Sense
+
GNDU
Force
+
GNDU
Sense
HCSMU
Low Force
HCSMU or
DHCSMU
HCSMU HCSMU
High Sense
HCSMU
High Force
HVSMU HVSMU HVSMU Force
Open
HVSMU and built-in
series resistor
HVSMU Force
+
series resistor
b
b. This condition is when all Input fields HVSMU, HCSMU, and HPSMU are blank.
Open Open Open