User`s guide
1- 70 Agilent EasyEXPERT User’s Guide Vol. 1, Edition 1
Main GUI
Configuration
Enable Module
Selector
This check box is used to enable/disable the module selector. If this box is checked,
the module selector can be used.
Input Specifies the modules connected to the module selector’s HVSMU, HCSMU, and
HPSMU input ports respectively. For the port which no module is connected, blank
the entry field.
Enable Series Resistor (100 kohm) for HVSMU
This check box is used to connect/disconnect the built-in series resistor (100 k) on
the HVSMU connection path. If this box is checked, the series resistor is connected.
Output Specifies the module selector connection status in the normal state. The module
selector makes the path to the output port from the input port specified by the
Default field.
When a measurement is performed, the module selector performs automatic
switching in every test. The selector makes the path to the output port from the
module used for the test. The connection status is indicated by the Status indicator
mounted on the module selector. See
Table 1-3 on page 1-71.
Status Shows one of the following status.
Not Detected Module selector has not been detected yet.
Communication Established Normal status. Module selector can be used.
Communication Lost Once the module selector was detected. However
the communication has been broken now.
Auto Detection This check box is used to enable/disable the module selector automatic detection. If
this box is checked, the automatic detection of the module selector is performed
when EasyEXPERT is launched.
Diagnosis The Start Diagnosis button is used to start the module selector diagnosis. And the
Status shows the diagnosis result.
Apply This button applies the setting.