User`s guide
Agilent EasyEXPERT User’s Guide Vol. 1, Edition 1 1- 69
Main GUI
Configuration
SMU/PG Selector
Not available for the B1505A. The SMU/PG Selector tab screen is used to specify
the input connection and the channel connection status of the Agilent 16440A
SMU/PGU selector (B1500A-A04). The selector can be controlled by the B1500A
via the Digital I/O connector and the Agilent 16445A selector adapter.
Two selectors can be used. The first selector must be connected to the Digital I/O
connector via the 16445A. And if the second selector is used, it must be connected
to the first selector’s Control Output connector. See
“SMU/PG Selector” on page
4-49.
Enable SMU/PG
Selector
To use the selector, connect the selector to the B1500A, and check this check box.
Selector CH1/CH2 CH1 and CH2 indicates the channels 1 and 2 of the first selector respectively.
Selector CH3/CH4 CH3 and CH4 indicates the channels 1 and 2 of the second selector respectively.
SMU Assignment Specifies the SMU connected to the SMU input of the selector channel CH1, CH2,
CH3, or CH4.
Input/Output Path Switching operation of the selector channel CH1, CH2, CH3, or CH4
The CH1 and CH3 also provides the PGU Open status which is set by using the
Advanced window of the Measurement Setup.
Module Selector
Only for the B1505A. This tab screen is used to specify the input connection and the
default input-to-output path of the module selector. The selector can be controlled
by the B1505A via the Digital I/O connector.
Value Input-to-output connection and description
Always SMU Selector always makes the path from Input SMU to Output.
Normally PGU
(AUX)
Selector performs automatic switching in every test. The
selector normally makes the path from Input PGU to Output
and makes the path from Input SMU to Output only for the test
which uses the SMU connected to the Input SMU port.