User`s guide

1- 68 Agilent EasyEXPERT Users Guide Vol. 1, Edition 1
Main GUI
Configuration
Enter the label used to specify the input port to the right entry field. The port name
and the connected module name can be the label. Or select one from the pull-down
menu. There must be no colon (:) in the label.
If the Kelvin check box is checked, the coupled even number port is grayed out.
If the CMU check box is checked, the SWM IN10 for the E5250A or the SWM
IN14 for the B2200A/B2201A is grayed out.
Kelvin Check this box if this port is used for the Kelvin connection. Connect the switching
matrix inputs to the SMU as follows.
for the E5250A (m: integer 1 to 3)
SMU INPUT 2m-1 to the SMU Force terminal
SMU INPUT 2m to the SMU Sense terminal
for B2200A/B2201A (m: integer 1 to 6)
Input 2m-1 to the SMU Force terminal
Input 2m to the SMU Sense terminal
CMU Check this check box if this port is connected to the CMU. Connect the switching
matrix inputs to the CMU as follows.
for the E5250A
AUX INPUT CV1 to the CMU Hpot/Hcur terminals
AUX INPUT CV2 to the CMU Lpot/Lcur terminals
for B2200A/B2201A
Input 13 to the CMU Hpot/Hcur terminals
Input 14 to the CMU Lpot/Lcur terminals
To obtain accurate capacitance measurement result, perform capacitance
compensation. See Agilent B1500A Users Guide for the capacitance compensation
when the switching matrix is used.
Also see “Extended Configuration” on page 1-79.
Extended
Configuration
This button opens the “Extended Configuration” on page 1-79 dialog box.