User`s guide

Agilent EasyEXPERT User’s Guide Vol. 1, Edition 1 1- 67
Main GUI
Configuration
ASU Serial
Number
Serial number of the usable ASU. This cell shows the serial number even if the ASU
is disconnected.
If an incorrect ASU is connected to the SMU, this cell displays an error message
instead of the serial number. Error message is also displayed if an ASU is connected
to the SMU which is not available for using ASU.
ASU I/O Path ASU switching operation, SMU or AUX
Switching Matrix
Not available for the B1505A. The Switching Matrix tab screen is used to confirm
the GPIB connection with Agilent B2200A/B2201A/E5250A switching matrix, and
defines the connections between the measurement modules and the switching matrix
input ports.
Switching Matrix
Type
Displays the type of the switching matrix. One of the following.
B2200A Femto Leakage Switch Mainframe
B2201A 14ch Low Leakage Switch Mainframe
E5250A Low Leakage Switch Mainframe
No Switching Matrix (if switching matrix is not connected)
GPIB Address Specifies the GPIB address of the switching matrix. 1 to 30.
Poll Confirms if the switching matrix with the specified GPIB Address is connected to
the B1500A.
Start Self Test This button starts the selftest of the switching matrix.
SWM IN SWM INn means the switching matrix input port n. where, n is an integer 1 to 10 for
the E5250A or an integer 1 to 14 for the B2200A/B2201A.
Value Input-to-output connection and description
SMU ASU always makes the path from the ASU Force input to the ASU
output.
AUX ASU performs automatic switching in every test. The ASU normally
makes the path from the AUX In input to the ASU output and makes the
path from the SMU input to the ASU output for the test using the SMU.