User`s guide
Agilent EasyEXPERT User’s Guide Vol. 1, Edition 1 1- 65
Main GUI
Configuration
Main Frame
Diagnosis
This area lists Item and Status.
Item Name of diagnosis item. To select, check the left check box.
Status Diagnosis status. Pass or fail.
Select All Selects all items for diagnosis. This checks the box for all items.
Unselect All Unselects all items. This unchecks the box for all items.
Start Diagnosis Starts diagnosis for the selected items. To perform diagnosis, specify the items to
perform and click this button. A dialog box appears. Follow the dialog box to
perform the diagnosis. Some items need your judgement, pass or fail. Also you will
need to connect an adapter, press keys, check LEDs, and so on.
Modules
The Modules tab screen is used to perform self-test of modules.
Slot Configuration This area lists Slot, Module Type, Name, and Status.
Slot Slot number where the module is installed. To select, check the
left check box.
Module Type Module model number
Name Module name. See Table 1-2.
Status Self-test status, Passed, Failed, Recovered, or ---- (self-test
has not been performed)
SCUU Specifies if the module is connected to the SCUU.
Notes Error information when the module self-test failed.
Select All Selects all modules for self-test. This checks the check box for all modules.
Unselect All Unselects all modules. This unchecks the box for all modules.
Start Self Test Starts self-test for the selected modules.
Recover Module This button is for service use only. Makes the failed modules respond again. This
button does not solve the problem that causes the Failed status.
Abort Aborts self-test.