1- 54 Agilent EasyEXPERT Users Guide Vol. 1, Edition 1
Main GUI
Repeat Measurement Setup
Radius Specifies the radius value of the temperature range in deg C.
Range: 0.5 to 50. Resolution: 0.5.
Delay Specifies the delay time from the temperature condition
satisfaction to starting of the action specified by the Action
field, in seconds. Range: 0 to 3600, Resolution: 1.
Action Selects an action executed if the measured temperature is in the
specified range, from the following alternatives:
TRIG: Does the single test execution after the
specified delay time, and then goes to the
next operation. This is useful to do the
single test execution in the specified
temperature range.
NOP: No operation (single test execution is
skipped). Waits for the delay time, and goes
to the next operation. This is useful to wait
in the specified temperature range.
END: Sets the status to True, and exits the repeat
measurement after the delay time. This is
useful to exit the repeat measurement in the
specified temperature range.
Add Adds a new entry for the thermo-trigger condition and action
above the selected entry.
Delete Deletes the selected entry.
Up/Down Moves the selected entry upward or downward.
OK Applies the setup changes and closes this dialog box.
Cancel Cancels the setup changes and closes this dialog box.
Apply Applies the setup changes.