User`s guide
Agilent EasyEXPERT User’s Guide Vol. 1, Edition 1 1- 53
Main GUI
Repeat Measurement Setup
Thermo-trigger List
This dialog box is opened by clicking the N1265A Thermo-trigger List button, and
is used to set the Thermo-trigger function which is defined by listing multiple sets of
the temperature condition and action executed when the condition is satisfied. A
setup in a row of the list is used for one loop sequence of the repeat measurement.
The top row is for the first sequence of the repeat measurement, the second row is
for the second sequence, and so on. If the repeat measurement is repeated over the
number of rows set in the list, the list is used from the top row again. This dialog box
can have a maximum of 1001 rows.
And also, in this dialog box, you can import any list from an external file and export
the list to an external file, in the CSV or XMLSS form.
Figure 1-6 Thermo-trigger List dialog box
N (index) Select an entry that you wish to edit during the editing. During
the repeat measurement execution, this button indicates the
current-waiting entry.
Temp Selects the temperature used in the condition from the
following alternatives:
T1: Temperature value measured by the thermometer channel 1.
T2: Temperature value measured by the thermometer channel 2.
T1-T2: Difference between T1 and T2.
Center Specifies the center value of the temperature range in deg C.
Range: -50 to 300. Resolution: 0.5.