User`s guide

1- 46 Agilent EasyEXPERT Users Guide Vol. 1, Edition 1
Main GUI
Tracer Test
The waveform sampling is started from the beginning of the pulse period within the
target observing step. The observing period on the waveforms is eventually decided
by setting the sampling delay (Delay) and sampling duration (Duration).
Figure 1-2
shows a sampling setup example.
NOTE If the observing period does not fully cover the measurement time, the measurement
data at the sweep step will be "NaN".
Figure 1-2 Sampling Setup Example
VAR1 Step: The ordinal number (N) of the sweep step on the primary sweep
source output. N-th step is specified. For the dual polarity
sweep, the ordinal number of the positive sweep step is valid.
For the interlacing sweep, every N-th step on repeated sweeps is
specified.
To specify the last step, enter LAST. Changing the number of
sweep step has no impact on the LAST setting.
For the interlacing sweep, every last step on repeated sweeps is
specified.
VAR2 Step: The ordinal number (M) of the sweep step on the secondary
sweep source output. M-th step is specified.
To specify each step, enter EACH.
VAR1
VAR2
VAR1 Step = 2
VAR2 Step = EACH
Delay = 0 s
Duration = pulse_period
1st
2nd
3rd
1st
2nd
3rd
Pulse
period
Oscilloscope View monitors
waveforms in these timeframes.