Main GUI
Tracer Test
Hold Time Hold time (in seconds) of the primary sweep source output just
before starting a sweep. 0 to 655.35. Setting resolution: 10 ms.
In the hold time, the primary sweep source forces the sweep
start level for staircase sweep output, or the pulse base level for
pulsed sweep output.
At the first step of the primary sweep, the longer one is applied
from the VAR1 and VAR2 Hold Time settings.
The following parameters are also displayed for the full mode if the IPULSE or
VPULSE is set to the VAR2 channel.
Pulse Base Pulse base value
Pulse Delay Pulse delay time. Set 0 for HR/MP/HPSMU.
Pulse Width Pulse width. Must be the same value for all of HR/MP/HPSMU.
CONST This area defines the constant output source.
Unit Measurement resource for the constant output. If multiple
CONST channels are defined, this field is used to specify the
channel for setup.
Source Output value.
Compliance Compliance value. Enter the current limit value for the voltage
source, or the voltage limit value for the current source.
Measurement is performed within this value.
V Compliance Only for UHCU, UHVU, or HVMCU. Limit value for the
voltage sweep output. The output voltage of the measurement
resource is automatically measured, and the measured voltage is
compared with this limit value. And then, if it exceeds the limit,
the measurement status is set to the compliance status, and the
sweep measurement stops. Voltage range and setting resolution
depend on the measurement resource.
The following parameters are also displayed for the full mode if the IPULSE or
VPULSE is set to the CONST channel. Click the CONST title bar to switch the
mode. The Pulse Peak entry field appears instead of the Source entry field.
Pulse Peak Pulse peak value.
Pulse Base Pulse base value
Pulse Delay Pulse delay time. Set 0 for HR/MP/HPSMU.
Pulse Width Pulse width. Must be the same value for all of HR/MP/HPSMU.