User`s guide

1- 34 Agilent EasyEXPERT Users Guide Vol. 1, Edition 1
Main GUI
Tracer Test
Hold Time Hold time (in seconds) of the primary sweep source output just
before starting a sweep. 0 to 655.35. Setting resolution: 10 ms.
In the hold time, the primary sweep source forces the sweep
start level for staircase sweep output, or the pulse base level for
pulsed sweep output.
At the first step of the primary sweep, the longer one is applied
from the VAR1 and VAR2 Hold Time settings.
The following parameters are also displayed for the full mode if the IPULSE or
VPULSE is set to the VAR1 channel.
Pulse Base Pulse base value
Pulse Delay Pulse delay time. Set 0 for HR/MP/HPSMU.
Pulse Width Pulse width. Must be the same value for all of HR/MP/HPSMU.
VAR2 This area defines the secondary sweep source output. The VAR2 channel performs
the linear single sweep operation. All of the following parameters are displayed for
the full mode. Only Start, Stop, Step, and NOS are effective for the short mode.
Click the VAR2 title bar to switch the mode.
Start Sweep start value
Stop Sweep stop value
Step Incremental step value of the sweep output. The value is
automatically calculated by using the start, stop, and NOS
values.
NOS Number of sweep steps. 1 to 101.
Compliance Compliance value. Enter the current limit value for the voltage
source, or the voltage limit value for the current source.
Measurement is performed within this value.
Pwr Comp. Power compliance value (in W), or OFF
V Compliance Only for UHCU, UHVU, or HVMCU. Limit value for the
voltage sweep output. The output voltage of the measurement
resource is automatically measured, and the measured voltage is
compared with this limit value. And then, if it exceeds the limit,
the measurement status is set to the compliance status, and the
sweep measurement stops. Voltage range and setting resolution
depend on the measurement resource.