User`s guide
Agilent EasyEXPERT User’s Guide Vol. 1, Edition 1 1- 33
Main GUI
Tracer Test
VAR1 This area defines the primary sweep source output. All of the following parameters
are displayed for the full mode. Only Mode, Stop, NOS, and Compliance are
displayed for the short mode. Click the VAR1 title bar to switch the mode.
Mode Sweep mode. The single mode sweeps from start to stop, and
the double mode sweeps from start to stop to start.
LIN-SGL Linear sweep, single mode
LOG-SGL Logarithmic sweep, single mode
LIN-DBL Linear sweep, double mode
LOG-DBL Logarithmic sweep, double mode
Start Sweep start value
Stop Sweep stop value
Step Incremental step value of the sweep output. The value is
automatically calculated by using the start, stop, and NOS
values.
NOS Number of sweep steps N. 1 to 1001.
Interlacing Number of the interlace sweep M. 1 to 100.
Sweep is repeated M times so that the final sweep step is equal
to Step/M. Total number of steps will be
N+(N-1)(M-1).
Compliance Compliance value. Enter the current limit value for the voltage
source, or the voltage limit value for the current source.
Measurement is performed within this value.
Pwr Comp. Power compliance value (in W), or OFF
V Compliance Only for UHCU, UHVU, or HVMCU. Limit value for the
voltage sweep output. The output voltage of the measurement
resource is automatically measured, and the measured voltage is
compared with this limit value. And then, if it exceeds the limit,
the measurement status is set to the compliance status, and the
sweep measurement stops. Voltage range and setting resolution
depend on the measurement resource.
Dual Polarity Dual polarity sweep ON or OFF. If this parameter is ON, the
VAR1 channel performs positive sweep from start to stop and
negative sweep from -start to -stop in this order. For the pulsed
negative sweep, the pulse base value is -1
(Pulse Base).