User`s guide
4- 72 Agilent EasyEXPERT User’s Guide Vol. 1, Edition 1
Function Details
Interlock Function
Interlock Function
The interlock function is designed to prevent electrical shock when a user touches
the measurement terminals. If the Interlock terminal is open, maximum output is
limited to 42 V.
To perform high voltage measurement more than 42 V, connect the Interlock
terminal to an interlock circuit of test fixture or connector plate. The interlock circuit
must have one LED and two mechanical switches placed near the shielding box
opening doors. And they must be connected together. For details on how to install
the interlock circuit, see Agilent B1500A or B1505A User’s Guide.
The interlock function works as shown below.
• When the interlock circuit is open, maximum output is limited to 42 V.
• When the interlock circuit is close, source channel can force its maximum output
value.
• When the interlock circuit is opened in the high voltage condition more than
42 V, output voltage is set to 0 V immediately.
CAUTION If the actual output value is not set to 0 V within 120 msec, the B1500 automatically
performs the initialization. Then perform self-test. Any module may be defective.
WARNING Dangerous voltage of up to the maximum voltage of SMUs may be present at
force, guard, and sense terminals if the Interlock terminal is closed.