SMU Measurement Time
SMU Measurement Time
Measurement time depends on integration time, measurement range, and other
measurement conditions, and can be expressed by the following formula:
Measurement time = Integration time + Overhead time
Integration time is the time required for measurement, and does not include such
factors as range changing or data compensation, which would be the overhead time.
Integration Time
Integration time is the time required to get measurement data. For accurate and
reliable measurement, integration time should be increased. Integration time can be
set for each type of A/D converter (ADC) used by the measurement channel.
• High-speed ADC
Installed in each measurement module. For high speed measurement. Sets the
number of averaging samples for getting one measurement data by setting the
Mode and the Factor value. Only PLC and TIME are available for SMU pulse.
AUTO Number of averaging samples = Factor reference
Factor = 1 to 1023
where reference is the number of averaging samples
automatically set by the B1500; this cannot be changed.
MANUAL Specify the number of averaging samples (Factor).
Factor = 1 to 1023
PLC Number of averaging samples = Factor 128
Factor = 1 to 100
The B1500 gets 128 samples in a power line cycle. Hence, the
Factor value is equal to the number of power line cycles.
TIME Specify the actual measurement time (Factor).
Factor = 2 s to 20 ms, 2 s step
For HR/MP/HPSMU, it performs the operation of MANUAL.
Then the number of averaging samples is automatically set to
the following value.