User`s guide

4- 64 Agilent EasyEXPERT Users Guide Vol. 1, Edition 1
Function Details
SMU Pulse
Pulse width must be the same for all of the HR/MP/HPSMU pulse output
channels. If a different value is entered, the longest value is set.
Pulse period = AUTO sets the longest one of the followings automatically.
Maximum period given by the pulse width and the duty ratio
Pulse period = 5 ms, for pulse delay + width 3 ms
Pulse period = pulse delay + width + 2 ms, for 3 ms < pulse delay + width 100
ms
Pulse period = pulse delay + width + 10 ms, for 100 ms < pulse delay + width
Measurement delay time = AUTO sets the optimum value automatically so that the
measurement is completed when the transition from peak to base is started by the
pulse output channel which starts the transition at first.