User`s guide
4- 62 Agilent EasyEXPERT User’s Guide Vol. 1, Edition 1
Function Details
SMU Pulse
SMU Pulse
SMU can apply voltage or current pulse. This section describes the pulse setup
parameters. The parameters are different by the number of pulse output channels as
shown in
Figure 4-32.
To Set Pulse Output
The following points must be noted when setting the pulse output.
• Cannot use the 1 nA output range and lower ranges.
• Cannot use multiple pulse output channels for the tests other than the Multi
Channel I/V Sweep.
• Cannot use multiple measurement channels for the tests other than the Multi
Channel I/V Sweep.
• The measurement channel uses the compliance range. The channel cannot
perform the auto ranging operation or the limited auto ranging operation.
Figure 4-32 SMU Pulse Setup Parameters
Voltage or current
Time
Hold time
Pulse period
Measurement time
Peak value
Base value
Previous value
Pulse width
Voltage or current
Hold time
Pulse period
Measurement delay time Measurement time
Peak value
Base value
Previous value
Pulse width
Pulse delay time
2. Using multiple pulse output channels
1. Using one pulse output channel