User`s guide
4- 60 Agilent EasyEXPERT User’s Guide Vol. 1, Edition 1
Function Details
SMU Compliance
Power Compliance
In addition to V compliance or I compliance, you can set power compliance for the
staircase sweep source. Power compliance is not available for HVSMU.
If you specify power compliance, SMU changes the current or voltage compliance
value every sweep step. The value is the lower value of either the specified
compliance value or the compliance value given by the following formula. See
Figure 4-30.
Current compliance = Specified power compliance value / Step voltage
Voltage compliance = Specified power compliance value / Step current
Figure 4-30 Power Compliance Output Area
Available values The following values are available for the power compliance:
MPSMU/HRSMU 1 mW to 2 W
HPSMU 1 mW to 20 W
HCSMU 1 mW to 40 W (DC), 1 mW to 400 W (pulse)
DHCSMU 1 mW to 80 W (DC), 1 mW to 800 W (pulse)
MCSMU 1 mW to 3 W (DC), 1 mW to 30 W (pulse)
UHCU 1 mW to 7.5 kW (pulse, 500 A range), 1 mW to 22.5 kW
(pulse, 1500 A range)
UHVU 1 mW to 100 W (DC), 1 mW to 200 W (pulse)
HVMCU 1 mW to 600 W (pulse, 2200 V range), 1 mW to 900 W
(pulse, 1500 V range)