User`s guide

Agilent EasyEXPERT User’s Guide Vol. 1, Edition 1 4- 59
Function Details
SMU Compliance
SMU Compliance
Compliance is the output limiter to prevent damage to the test device from
overcurrent, overvoltage, or overpower. Voltage compliance is for the current output
channels, and current compliance is for the voltage output channels.
When a channel reaches compliance, the channel acts as a constant voltage source or
a constant current source. The channel keeps the output value when reaching
compliance.
Compliance can be set with the same resolution and accuracy as output current or
output voltage. For the values, see Agilent B1500A or B1505A Users Guide.
Polarity and Output Area
Polarity of current compliance
The B1500 automatically sets current compliance for both positive and negative
polarity. For opposite polarity as the output voltage, the compliance value is
increased by an amount that is 2.5 % to 12 % of the range value in the lowest
range that covers the specified compliance value. See
Figure 4-29.
Polarity of voltage compliance
The B1500 automatically sets voltage compliance to the same polarity as the
output current. There is no compliance for opposite polarity.
Figure 4-29 Relation of Compliance and Output
2XWSXW9ROWDJH
&XUUHQW
_&RPSOLDQFH_
_&RPSOLDQFH_
9! 9
2XWSXW&XUUHQW
9ROWDJH
_&RPSOLDQFH_
_&RPSOLDQFH_
,! ,