Agilent EasyEXPERT User’s Guide Vol. 1, Edition 1 4- 41
Function Details
Bias Hold Function
Bias Hold Function
This function is used to keep source output after measurement. Source modules
apply the specified bias between measurements in a quick test or application test
that defines some classic test setups, or a repeat measurement. And the source
modules change the output value and the unused modules are disconnected when the
next measurement is started.
Output channels
and output value
If the bias hold function is OFF, the channel stops the output immediately after
measurement.
If the function is ON, the channel continues the output as shown below.
I/V Sweep measurement:.
VAR1 and VAR1’ output channels: Sweep start or stop value
VAR2 output channel: Sweep stop value
CONST output channels: Output value
SMU pulse output channels: Base value
C-V Sweep measurement:
Bias sweep output channel: Sweep start or stop value
CONST output channels: Output value
I/V-t Sampling measurement:
Constant source output channels: Base or source value
SPGU output:
SPGU keeps the output as shown in Table 4-6.
Period of bias hold From the end of a measurement to the end of a single measurement
From the end of a measurement to the start of the next measurement during a
quick test, application test, or repeat measurement
Reset of source
output
The source output is reset after a quick test, application test, repeat measurement, or
single measurement. The SMU is set to 0 V output and 0.1 mA compliance. The
MFCMU is set to 0 V DC bias and 0 V oscillator level.