Agilent EasyEXPERT User’s Guide Vol. 1, Edition 1 4- 37
Function Details
Sweep Abort Function
Sweep Abort Function
This function is used to stop sweep measurement automatically when one of the
following conditions is detected. This function is useful to reduce sweep time and to
prevent damage to the device during measurement.
Effective measurement modes:
I/V Sweep measurement
Multi Channel I/V Sweep measurement
I/V List Sweep measurement
C-V Sweep measurement
Stop conditions:
The output reaches voltage compliance or current compliance
A measurement value exceeds the specified measurement range
An SMU oscillates
An error is caused in the MFCMU
This automatic sweep abort function is enabled by using the Sweep [xxxx] status
field on the Measurement Setup screen of the Classic Test definition.
Sweep [CONTINUE AT ANY] status: Sets the abort function OFF. Not
available if the power compliance has been set.
Sweep [STOP AT ANY ABNORMAL] status: Sets the abort function ON.
Sweep [SKIP AT ANY ABNORMAL] status: Sets the abort function ON.
However if a stop condition is detected during a subordinate sweep
measurement, this function stops the VAR1 sweep once, changes the VAR2 step
value, and continues the sweep measurement. This mode is not available for
Desktop EasyEXPERT and for I/V List Sweep and C-V Sweep measurements.
When abort occurs
After measurement is aborted, the source forces the following value.
Sweep source: Start value
Pulsed source or Pulsed sweep source: Pulse base value