User`s guide

4- 16 Agilent EasyEXPERT Users Guide Vol. 1, Edition 1
Function Details
I/V-t Sampling Measurement
Interval 0.0001 + 0.00002 (number of measurement
channels-1)
No of Samples Number of samples. Integer. 1 to the following value.
For linear: 100001 / (number of measurement channels)
For logarithmic: 1 + (number of data for 11 decades)
Total Sampling
Time
This field just displays the total sampling time that is the
time from the measurement start time for the first point to
the end of sampling measurement. It does not include the
hold time.
Total Sampling Time = Interval No of Samples
Output
Sequence
Source output sequence. SIMULTANEOUS or
SEQUENTIAL. See
“Source Output Sequence and Time
Origin” on page 4-18.
Hold Time Time since the Source value output until the first sampling
point, in seconds. 0 to 655.35 s, resolution 0.01 s.
The following values are also available for Interval < 0.002
s. |Hold Time| will be the time since the sampling start until
the Source value output.
-0.09 to -0.0001 s, resolution 0.0001 s.
Base Hold
Time
Hold time of the Base value output until the Source value
output, in seconds. 0 to 655.35 s, resolution 0.01 s.
3. Specify the following parameters for the SMU in the Constants area on the
Measurement Setup screen.
Base, Source Base or Source value, in V or A.
Base value is available for all output channels for the
SIMULTANEOUS mode and the last source channel for the
SEQUENTIAL mode.
Compliance Compliance value, in A or V.