User`s guide
4- 12 Agilent EasyEXPERT User’s Guide Vol. 1, Edition 1
Function Details
Multi Channel I/V Sweep Measurement
Multi Channel I/V Sweep Measurement
Multi Channel I/V Sweep measurement is similar to I/V Sweep measurement.
Changes from I/V Sweep are listed below.
• All SMU can be set to the VAR1 sweep source (up to 10 channels)
• All SMU can be set to the pulse source (up to 10 channels)
• There is no restriction for the VAR1 output mode, voltage or current
• VAR1’ sweep mode is not available
The timing of the sweep output and measurements are shown in Figure 4-8. The DC
output channels start to output in the order in which they are listed on the Channel
Setup screen. If multiple measurement channels are used, the channels perform the
measurements in order. See
“Multiple Measurement Channels” on page 4-69.
If multiple pulse output channels are used, the DC output channels start to output in
the order in which they are listed on the Channel Setup screen, and then the pulse
output channels start to output simultaneously. And the all measurement channels
perform measurements simultaneously.
Figure 4-8 Multi Channel I/V Sweep Measurements using High-Resolution A/D Converter
Start measurement
Voltage or current
Stop value
Start value
0
Time
:Measurement
Hold time
Delay time
Stop value
Start value
0
Voltage or current
Stop value
Start value
0
Voltage or current
Primary sweep source
Second sweep source
Tenth sweep source
Delay time