4- 10 Agilent EasyEXPERT User’s Guide Vol. 1, Edition 1
Function Details
I/V Sweep Measurement
Period SMU forces the next pulse after specified pulse period.
Width Time from when SMU output starts to change from base
value to time when SMU starts to return from peak value.
Measurements are made while the peak value is output.
Base The base output value of the SMU pulse.
Pulse peak value Pulse peak value depends on the Function setting of the pulse source SMU. See
Table 4-1.
Table 4-1 Pulse Peak and Function
Measurement
channel
Only one SMU can be used for the measurement channel.
The measurement channel always uses the compliance range which covers the
setting compliance value. The Range button in the Measurement Setup is not
available.
Function Pulse peak
CONST Source value in the Constants area
VAR1 Calculated from the Start, Stop, Step values of the VAR1.
VAR2 Calculated from the Start, Stop, No of Step values of the
VA R 2.
VAR1’ Calculated from the Start, Stop, Step values of the VAR1
and the Offset and Ratio values of the VAR1’.