User`s guide
3- 14 Agilent EasyEXPERT User’s Guide Vol. 1, Edition 1
Application Test Definition
Test Output
Test Output
This setup screen is used to set the test result output (test result display and record)
that is the output of the test definition.
The test output setup is optional. If you have the following requirements, define the
analysis parameters and set the display parameters.
• To send the test result data to the subsequent tests in an application test
• To make the program branching depends on the test result
• To make the calculation using the test data in the Test Contents
• To display/record the test result of this application test
To read the analysis parameters defined in the application test used in a test
definition, use the
“External Variable Setup (page 3-23)” dialog box.
Define Own Output
Parameters
Check this check box to set the test result outputs of the test definition. While this
box is checked, the X-Y Graph, List Display, and Parameters areas are available.
The test result output will be obtained every execution of the test definition.
Allocate Data
Display for each
test
This field sets the Multi Display function to Enable, Disable, or Comply with above
level.
If Enable is selected, the test result data of the same test setup name will be
displayed on the exclusive Data Display window and the test result data of the
different test setup name will be displayed on the new Data Display window.
If Disable is selected, the test result data will be always displayed on the singular
Data Display window even if it is the test result data of the different test setup name.
Select Comply with above level for complying with the Multi Display ON/OFF
status or the setting of the application test which uses this application test internally.
Define Analysis
Parameters...
This button opens the window. This “Analysis Parameter Definition (page 3-15)”
window is used to define the vector data and the scalar data used in the test result
output. The defined variable is called as the analysis parameter.
X-Y Graph,
List
Display,
Parameters
The areas are used to set the test result output (test result display and record) that is
the output of the test definition. See
“Display Setup” on page 2-46.