User`s guide
Agilent EasyEXPERT User’s Guide Vol. 1, Edition 1 3- 5
Application Test Definition
Test Specification
Test Specification
This setup screen is used to define the test definition information and the input
parameters for the test definition.
Test Information The test information is defined in this area.
Category Displays the category of the test definition.
The button opens a dialog box used to set the category. It must
be 1 to 30 characters. Multiple categories can be set to a test
definition.
Test Name Specifies the test definition name. It must be 1 to 30 characters.
Icon Specifies a bitmap file used for the test definition icon listed in
the Library area of the main screen.
Browse...: Opens a dialog box used to select a bitmap file.
Cancel: Returns to the default setting.
Description Enters the description of the test definition.
Device Parameters
Definition
The device parameters are defined in this area. Type of the device parameter is
Numeric.
Name Device parameter name
Default Default value of the parameter
Description Description of the parameter
Add Adds a row for the device parameter.
Delete Deletes the selected device parameter. To select the parameter,
use the left radio button.
Up Moves the selected device parameter upward.
Down Moves the selected device parameter downward.