User`s guide
Agilent EasyEXPERT User’s Guide Vol. 1, Edition 1 2- 55
Classic Test Definition
Range Setup Window
Range Setup Window
This window is opened by clicking the Range button on the C-V Sweep
Measurement Setup screen, and is used to set the measurement ranging operation.
CMU Measurement
Range
This area sets the CMU measurement ranging operation.
Unit CMU used for the measurement
Mode Ranging mode, AUTO (auto ranging) or FIXED (fixed range).
If Mode=AUTO, the MFCMU automatically searches for and
measures at the range that provides the best resolution.
If
Mode=FIXED, the MFCMU measures at the range specified
by the Range field.
Range Measurement range. This field is effective when Mode=FIXED.
Available measurement ranges depend on the measurement
frequency setting. See
Table 2-3.
Table 2-3 Available Measurement Ranges
SMU Measurement
Range
This area sets the SMU measurement ranging operation.
For the measurement ranging function, see “SMU Ranging Mode” on page 4-55.
Unit SMU used for the measurement
Name Variable name for the measurement data
Mode Ranging mode, AUTO (auto ranging), LIMITED (limited auto
ranging), or FIXED (fixed range)
Range Range value. For auto ranging, ignore this field. For limited
auto ranging, the measurement channel never uses ranges less
than the specified range.
Frequency Available measurement ranges
200 kHz 50 , 100 , 300 , 1 k, 3 k, 10 k, 30 k, 100 k, 300 k
2 MHz 50 , 100 , 300 , 1 k, 3 k, 10 k, 30 k
5 MHz 50 , 100 , 300 , 1 k, 3 k