User`s guide

Agilent EasyEXPERT User’s Guide Vol. 1, Edition 1 2- 53
Classic Test Definition
Advanced Setup Window
Advanced Setup Window
This window is opened by clicking the Advanced button on the I/V Sweep / Multi
Channel I/V Sweep / I/V List Sweep / I/V-t Sampling Measurement Setup screen,
and is used to set the following functions of the measurement channel.
Channel Settings This area sets the series resistor and the filter.
Unit SMU used for the measurement
VName Variable name of voltage output or measurement data
IName Variable name of current output or measurement data
Series R Series resistor NONE or 1MOHM. See “SMU Series Resistor”
on page 4-71.
SMU series resistor is not available for MCSMU, HCSMU, and
HVSMU.
SMU Filter SMU filter ON or OFF. See “SMU Filter” on page 4-70.
V Compliance Only for UHCU, UHVU, or HVMCU. Limit value for the
voltage sweep output. The output voltage of the measurement
resource is automatically measured, and the measured voltage is
compared with this limit value. And then, if it exceeds the limit,
the measurement status is set to the compliance status, and the
sweep measurement stops according to the setting in the Sweep
[xxxx] status field on the Measurement Setup screen. Voltage
range and setting resolution depend on the measurement
resource.
Wait Time Control This area sets the source output wait time and the measurement wait time. See “Wait
Time” on page 4-68. The Output Wait sets the source output wait time. The
Measurement Wait sets the measurement wait time.
Factor Factor value. 0 to 10, 0.1 resolution. This is the N value shown
in
“Wait Time” on page 4-68.