User`s guide
Agilent EasyEXPERT User’s Guide Vol. 1, Edition 1 2- 35
Classic Test Definition
Direct Control
CH3 CH3 status, DEFAULT or PGU OPEN.
CH1 represents channel 1 on the first selector and CH3 represents channel 1 on the
second selector. See
“SMU/PG Selector” on page 4-49 for details on the selector.
Primary Setup
Primary setup is the initial setting set for when the Direct Control test is started. This
setting is different from the initial setting set by the device clear or the *RST
command. The differences are listed below.
• Data output format
FMT 13, 0 is set.
• Program memory
Setup is not changed. The last used setting is retained.
• Internal variables
Setup is not changed. The last used setting is retained.
• Measurement condition setup commands
Setup is not changed for the following commands. The last used setting is
retained.
MM
WV, WI, WSV, WSI, WNX
PV, PI, PWV, PWI
BDV
QSV, QSO
LGI, LGV, LSI, LSV, LSSI, LSSV
BGI, BGV, BSI, BSV, BSSI, BSSV
FC, WDCV
PDCV, PWDCV
WFC
WA CV
MDCV