User`s guide

Agilent EasyEXPERT User’s Guide Vol. 1, Edition 1 2- 29
Classic Test Definition
C-V Sweep
Up Moves the selected frequency upward.
Down Moves the selected frequency downward.
Integration Time This area sets the A/D converter of MFCMU.
Mode A/D converter operation mode, AUTO or PLC
Factor Factor for the initial value. See “Mode, Factor” on page 4-20.
Timing This area defines the hold time and the delay time. See “To Specify Source Output
Time” on page 4-67.
Hold Hold time. This is the time from beginning of channel output to
beginning of delay time. 0 to 655.35 s, 10 ms resolution.
Delay Hold time. This is the time from end of hold time to beginning
of measurement. The MFCMU starts measurement as soon as
the wait time elapses. 0 to 655.35 s, 0.1 ms resolution.
Constant This area defines the constant voltage/current source.
Unit SMU used for the constant source
V Name Variable name for constant voltage output or measurement data
I Name Variable name for constant current output or measurement data
Mode Output mode, V (voltage), I (current), or COMMON (ground).
Source Output value
Compliance Compliance value. Enter the current limit value for the voltage
source, or the voltage limit value for the current source.
Sweep status This field specifies the sweep abort condition. See “Sweep Abort Function” on page
4-37.
Range This button opens a window used to set the measurement ranging operation. See
“Range Setup Window” on page 2-55.
ADC/Integ This button opens a window used to set the A/D converter of the SMU measurement
channels. See
“ADC and Integration Time Setup Window” on page 2-50.
Advanced This button opens a window used to set the other functions of the measurement
channel. See
“Advanced Setup Window for C-V Sweep” on page 2-57.