User`s guide

Agilent EasyEXPERT User’s Guide Vol. 1, Edition 1 2- 27
Classic Test Definition
C-V Sweep
Channel Setup
This setup screen is used to define the channel setups and the variables for the
channel output data, time stamp data, and index data.
Channel Definition This area defines the channel setups and the variables for the channel output data.
NOTE UHCU and HVMCU cannot be used.
Unit Measurement resource used for the source output or
measurement
V Name Variable name of voltage output or measurement data
I Name Variable name of current output or measurement data
Mode SMU output mode, V (voltage), I (current), or COMMON
(ground)
Function Output function, VAR1 (primary sweep) or CONST (constant)
Add Adds a row for the channel setup.
Delete Deletes the channel setup. To select the channel setup, use the
left radio button.
Up Moves the selected channel setup upward.
Down Moves the selected channel setup downward.
If multiple output channels are defined, the channels start output sequentially. The
order is top to bottom of the channels defined in this screen. However the last output
channel is always the MFCMU. The output stops in the opposite order.
Miscellaneous
Variables
This area defines the variables for the time stamp data and the index data.
Time Stamp Name Variable name of the time stamp. Time stamp data gives the
time when step measurement is started.
Index Name Variable name of the data index. Integer above 0 is returned.