User`s guide
2- 16 Agilent EasyEXPERT User’s Guide Vol. 1, Edition 1
Classic Test Definition
I/V List Sweep
VAR2 This area defines the secondary sweep source output.
Unit SMU used for the VAR2 output
Name Variable name for the VAR2 output data
Source Click the grid button to open the Define vector data dialog box.
See
“Define vector data” on page 2-18.
Compliance Compliance value. Enter the current limit value for the voltage
source, or the voltage limit value for the current source.
Pwr Comp Power compliance value (in W), or OFF
Timing This area defines the hold time and the delay time. See “To Specify Source Output
Time” on page 4-67.
Hold Hold time. This is the time from the beginning of channel
output to the beginning of delay time. 0 to 655.35 s, 10 ms
resolution.
Delay Delay time. This is the time between end of step changing and
the beginning of measurement. For the first step, this is the time
between the end of hold time and the beginning of
measurement. The measurement channels start measurement as
soon as the wait time elapses. 0 to 65.535 s, 0.1 ms
resolution.Click the grid button to open the Define vector data
dialog box. See
“Define vector data” on page 2-18.
SMU Pulse This area defines the SMU pulse setup parameters. See “SMU Pulse” on page 4-62.
Unit SMU used for the pulse output
Period Pulse period. Cannot set this value. SMU applies the pulse after
the delay time immediately.
Width Pulse width
Base Pulse base value