Classic Test Definition
Multi Channel I/V Sweep
Name Variable name for the VAR2 output data
Start Sweep start value
Stop Sweep stop value. The value is automatically calculated from
the start, step, and number of step values.
Step Incremental step value for the sweep output
No of Step Number of sweep steps
Compliance Compliance value. Enter the current limit value for the voltage
source, or the voltage limit value for the current source.
Pwr Comp Power compliance value (in W), or OFF
Timing This area defines the hold time and the delay time. See “To Specify Source Output
Time” on page 4-67.
Hold Hold time. This is the time between the beginning of channel
output and the beginning of delay time. 0 to 655.35 s, 10 ms
resolution.
Delay Delay time. This is the time between the end of the hold time
and the beginning of measurement. The measurement channels
start measurement as soon as the wait time elapses. 0 to 65.535
s, 0.1 ms resolution.
Sweep status This field specifies the sweep abort condition. See “Sweep Abort Function” on page
4-37.
Range This button opens a window for setting measurement ranging operation. See “SMU
Range Setup Window” on page 2-48.
ADC/Integ This button opens a window for setting the A/D converter of the measurement
channels. See
“ADC and Integration Time Setup Window” on page 2-50.
Pulse This button opens a window for setting the SMU pulse output. See “Pulse Setup
Window” on page 2-51.
Advanced This button opens a window for setting other functions of the measurement channel.
See
“Advanced Setup Window” on page 2-53.
SPGU Pulse /
ALWG Setup
This button opens a window used to set the SPGU output. See “SPGU Pulse Setup
Window” on page 2-63 or “SPGU ALWG Setup Window” on page 2-67.