User`s guide
Agilent EasyEXPERT User’s Guide Vol. 1, Edition 1 2- 11
Classic Test Definition
Multi Channel I/V Sweep
Measurement Setup
This setup screen is used to set the source channel outputs, the measurement ranging
mode, the sweep abort condition, the measurement channel A/D converter, the
series resistor, the filter, and the source output/measurement wait time.
VAR1 This area defines the primary sweep source output.
Direction Sweep direction, single (start to stop) or double (start to stop to
start)
Linear/Log Sweep output, LINEAR, LOG10, LOG25, or LOG50 where the
number following LOG indicates the number of sweep points in
one decade.
No of Step Number of sweep steps
Unit SMU used for the VAR1 output
Name Variable name for the VAR1 output data
Start Sweep start value
Stop Sweep stop value
Compliance Compliance value. Enter the current limit value for the voltage
source, or the voltage limit value for the current source.
Pwr Comp Power compliance value (in W), or OFF
Constants This area defines the constant output source.
Unit SMU used for the constant source
V Name Variable name for the constant voltage output or measurement
data
I Name Variable name for the constant current output or measurement
data
Mode Output mode, V (voltage) or I (current)
Source Output value
Compliance Compliance value. Enter the current limit value for the voltage
source, or the voltage limit value for the current source.
VAR2 This area defines the secondary sweep source output.
Unit SMU used for the VAR2 output